  • @robtemple3445
    @robtemple3445 2 года назад +5

    I just unboxed my new TC1 tester today. Charged it until the LED turned green. Dead! No joy.
    Searched RUclips for TC1 tester repair, & found your day saving video.
    Thank You! Worked like a charm!

  • @paulusvanos1826
    @paulusvanos1826 4 года назад +5

    You surely made my day ! Thanks for the excellent tip!
    I drilled a 3 mm hole in the front side panel, at about the same position as the charging LED hole, but at the far right end side. Then I made the hole to a small horizontal rectangular slot, and glued a miniature slide switch to the inside of the side panel, as a small power on/off switch.

    • @mandovibrazrocha2295
      @mandovibrazrocha2295 9 месяцев назад

      Nice idea, batt life now is longer, we now have a reset, software never gets tired... it lasts for ever... so thanks a lot.

  • @j05eph8jin
    @j05eph8jin 2 года назад +1

    The tip has brought my dead TC1 back to life. Thank you!

  • @IamVlanOne
    @IamVlanOne 4 года назад +4

    You sir are a legend! Just did this and it worked!

  • @mrcarrepairman
    @mrcarrepairman Год назад +1

    I have same problem. Tester comes alive once in a awhile but then it's dead. The reset fix in this video works sometimes. Battery fully charged.

  • @njokuchiemezuo3109
    @njokuchiemezuo3109 2 года назад +1

    Worked. Though I have a TC-7h which comes with a connector so you need not to desolder anything. Just disconnect and connect after 10secs. Thanks.

  • @cajuncoinhunter
    @cajuncoinhunter Год назад +1

    Hey there Mon Ami ..... I am glad I stumbled across your video ... I've been having my tester for 3 years now and it acted up today, it wouldn't turn off and it was frozen in the no part or bad part mode... I knew it needed a reset , but how was the 64,000.00 $$ question .... I searched TC1 tester reset and nothing came up , so I reworded the search and voila you came up , Thanks for the tip on taking the battery out of the circuit , BTW mine had a plug on the end of it .... I joined your channel and rang the bell .....Thanks again..

  • @ilkkasorsa9778
    @ilkkasorsa9778 5 лет назад +3

    Thank you for the tip! My tester is otherwise just like your tester but has a connector for the battery on the pcb. So, one can't really say that chinese RD engineer hasn't done anything to mitigate this problem in later revisions.. If this problem will persist I guess I'll take a little bit similar step like chinese engineer and install a nc reed relay or mercury switch.

  • @rmark1083
    @rmark1083 2 года назад +1

    Did exactly as you said and it worked.
    Can't thank you enough!🍺

  • @tomcat501976
    @tomcat501976 2 года назад +2

    Hi, just seen this video. I have one like this but not in the case. It keeps showing bad or unknown part. I'm going to try this reset on it. Thanks great video.

    • @cajuncoinhunter
      @cajuncoinhunter Год назад

      Mine was doing the same as yours , it worked on mine ....

  • @Ni5ei
    @Ni5ei 4 года назад +4

    Or, since you have to open it up and use your soldering iron anyway, solder in a breaker switch which will act as a reset switch next time it happens.

  • @HobbyzoneUtara
    @HobbyzoneUtara 4 года назад +1

    Nice man. My TC1 back to normal.

  • @danedewaard8215
    @danedewaard8215 Год назад

    Great Video!!! VERY HELPFUL!!!!!!!!!!!

  • @lambdasun4520
    @lambdasun4520 8 месяцев назад

    I believe I have the same problem. Screen randomly starts flashing on and off and then it shuts down. I will try cutting a little hole on the side with a dremel and adding a switch.

  • @davidbolha
    @davidbolha 10 месяцев назад

    According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), posted by user "zappenduster" on 21st June 2018 the gadget consumes about 0,2 mA in Standby/Power-Off mode. 🤔😉🙄
    That means 150 mAh per month (0,2 mA x 24 h x 31 days). So a fully charged battery would be drained after three months of non-use. 🤔😕😞😟😳

  • @speedsterh
    @speedsterh 3 года назад +2

    My TC1 switches ON but only detects a 5V cell whatever is connected to it (and whatever the pins). I thought that maybe the ZIF socket had its pins oxydated and made the test fail, but I opened the casing, directly soldered some flying leads from the 1-2-3 points on the PCB to some grip-wires and the problem is the same ! If anyone found a fix for this, I'd definitely want to hear from you !

    • @enricoborsetti4833
      @enricoborsetti4833 2 года назад +1

      Mine switches on but the display is all lit up, completely white :-(

    • @davidbolha
      @davidbolha 10 месяцев назад

      According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), originally posted by user "zappenduster" on 21st June 2018, user "indman" replied (1st July 2018) that you should try:
      - either removing the component (marked by DZ1 on the PCB) 6V8A (which is a zener diode).
      - OR removing the 6-pin component V05 (on the PCB it is marked by DZ2). It's full model name is SRV05-4.
      If the problem remains, then you have damaged the ATMega324 processor. In that case you're going to have to get a replacement one and a new firmware is necessary. Meaning you'll have to reprogram it. 🤔😕😟
      But do read that outlined topic on that forum for more possible ideas on fixing it.

  • @lllCANDYMAANlll
    @lllCANDYMAANlll 2 года назад +1

    I might put a switch in mine when I get it

  • @aviationproA
    @aviationproA 2 года назад +2

    I got a cap discharged during testing it charges but won't turn on.
    Do u have an idea how to fix it??

    • @sham1234567891
      @sham1234567891 2 года назад +2

      Same problem. Can't fix it the way suggested.

    • @davidbolha
      @davidbolha 10 месяцев назад

      According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), originally posted by user "zappenduster" on 21st June 2018, user "indman" replied (1st July 2018) that you should try:
      - either removing the component (marked by DZ1 on the PCB) 6V8A (which is a zener diode).
      - OR removing the 6-pin component V05 (on the PCB it is marked by DZ2). It's full model name is SRV05-4.
      If the problem remains, then you have damaged the ATMega324 processor. In that case you're going to have to get a replacement one and a new firmware is necessary. Meaning you'll have to reprogram it. 🤔😕😟
      But do read that highlighted topic I outlined for more fixes.

    • @davidbolha
      @davidbolha 10 месяцев назад

      @@sham1234567891 See my post here under this comment. 😉😊

  • @chromium5988
    @chromium5988 Год назад +1

    My TC1 has a connector so I don't have to desolder the battery, but this fix doesn't help :( It was working fine one minute, then it said to charge the battery, then the screen stopped coming on altogether. I charge it and the light goes green after a few seconds, but I can no longer get anything on the screen :( I'd hardly used this, it was sitting on standby in case I ever needed it and then died lol

  • @adamcooper5108
    @adamcooper5108 2 года назад

    Thanks for the tip! Worked like a charm.

  • @TheRcEngineer
    @TheRcEngineer Год назад

    This really worked, thank you !

  • @ValgniTehki
    @ValgniTehki Месяц назад

    The problem with my one is its stuck in battery testing mode. Won't go back to capacitance testing for some reason. I have taken out the battery and put it back but still the same.

  • @klaastuik1437
    @klaastuik1437 3 года назад +1

    I have a TC1 AND did a test on a 9volt battery cell and now when i want to start it , i get short a white screen and shuts down.
    Is there somewhon now's how to repair this ?

    • @davidbolha
      @davidbolha 10 месяцев назад

      According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), originally posted by user "zappenduster" on 21st June 2018, user "indman" replied (1st July 2018) that you should try:
      - either removing the component (marked by DZ1 on the PCB) 6V8A (which is a zener diode).
      - OR removing the 6-pin component V05 (on the PCB it is marked by DZ2). It's full model name is SRV05-4.
      If the problem remains, then you have damaged the ATMega324 processor. In that case you're going to have to get a replacement one and a new firmware is necessary. Meaning you'll have to reprogram it. 🤔😕😟
      You shouldn't be measuring anything above 5 V.

  • @BradKwfc
    @BradKwfc Год назад +1

    Anyone have one that turns on but has a blank screen. Backlight comes on but nothing shows up.
    I disconnected and reconnected the battery.... Still get a blank screen

    • @davidbolha
      @davidbolha 10 месяцев назад +1

      According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), originally posted by user "zappenduster" on 21st June 2018, user "indman" replied (1st July 2018) that you should try:
      - either removing the component (marked by DZ1 on the PCB) 6V8A (which is a zener diode).
      - OR removing the 6-pin component V05 (on the PCB it is marked by DZ2). It's full model name is SRV05-4.
      If the problem remains, then you have damaged the ATMega324 processor. In that case you're going to have to get a replacement one and a new firmware is necessary. Meaning you'll have to reprogram it. 🤔😕😟
      Your problem might have happened from connecting a charged capacitor to the probe leads when you were trying to measure it.

  • @dwehiabufreih3161
    @dwehiabufreih3161 4 года назад +2

    i try to switch on the t7 i see 5.5v

  • @aboimad5178
    @aboimad5178 4 года назад +1

    doesnt work with me i have short maybe its on the v05 ic or some thing else this happend after i tested a capacitor many time many ways then oops it turned off and not turning on again

    • @Cashpots
      @Cashpots  4 года назад +3

      The capacitor charge has blown it I'm afraid. Can't win 'em all!

    • @leonardochile8735
      @leonardochile8735 3 года назад +1

      El mil se dañó por medir un capacitor cargado. Alguien sabe como arreglarlo?

  • @robinedminson6461
    @robinedminson6461 4 года назад +2

    Tried that. No dice, still dead!

    • @miketaylor803
      @miketaylor803 3 года назад +1

      have a look here... this chap found a shorted cap on the dc - dc converter for the zener tester
      www.eevblog.com/forum/testgear/tc1-lcr-meter-transistor-tester-fix/

    • @garygranato9164
      @garygranato9164 2 года назад +1

      +1

    • @davidbolha
      @davidbolha 10 месяцев назад

      @@miketaylor803 Gr8 topic indeed. I fixed mine yesterday by accidentally measuring a capacitor that had some residual charge in it. 😍😇🤗😏👌

  • @leohobbleohobb3781
    @leohobbleohobb3781 5 лет назад

    The tc1 brake it jst start to give flicering display etc.had 2 both just quit working as they should the 1 i got had a flickering display while testing. the second worked fine.have not used it for some time and recharged it ,no luck with selftest.it also suddenly got display flickering.it is bonkers

  • @dwehiabufreih3161
    @dwehiabufreih3161 4 года назад +1

    i have problem in t7

  • @Marcel-dit-Nono
    @Marcel-dit-Nono 19 дней назад

    Bonjour il faut mettre un bouton marche arrêt

  • @leonardochile8735
    @leonardochile8735 3 года назад

    Mi tester al medir capacitores me muestra una fuente. Esto paso por medir un capacitor y no descargarlo antes, por lo que se dañó y ya no mide capacitores. Alguien sabe cómo arreglarlo?

    • @davidbolha
      @davidbolha 10 месяцев назад

      According to a post on EEV Blog forums ( titled *"TC1 lcr-meter transistor-tester fix"* ), originally posted by user "zappenduster" on 21st June 2018, user "indman" replied (1st July 2018) that you should try:
      - either removing the component (marked by DZ1 on the PCB) 6V8A (which is a zener diode).
      - OR removing the 6-pin component V05 (on the PCB it is marked by DZ2). It's full model name is SRV05-4.
      If the problem remains, then you have damaged the ATMega324 processor. In that case you're going to have to get a replacement one and a new firmware is necessary. Meaning you'll have to reprogram it. 🤔😕😟
      Hope it helps.

  • @giuseppemascagna3091
    @giuseppemascagna3091 3 месяца назад

    Thx save me 15$