  • @julianbaccaro7733
    @julianbaccaro7733 5 лет назад

    which is the acceptance criteria for slope and offset? does it change if another electrode is used? for example, 1230B

    • @ADWAHungary
      @ADWAHungary  2 года назад

      Dear Julian,
      The accepted offset is +- 30 mV in 7pH, the accepted slope is between 80 to 110%, if you consider the theoretical slope 59.16 mV/pH@ 25 Celcius degree.

  • @vinaydasari4455
    @vinaydasari4455 10 лет назад +2

    What is the slope range of adwa ad1020 ph meter?

    • @ADWAHungary
      @ADWAHungary  10 лет назад

      Please see:
      www.adwainstruments.com/site/component/content/article/56