which is the acceptance criteria for slope and offset? does it change if another electrode is used? for example, 1230B
Dear Julian,The accepted offset is +- 30 mV in 7pH, the accepted slope is between 80 to 110%, if you consider the theoretical slope 59.16 mV/pH@ 25 Celcius degree.
What is the slope range of adwa ad1020 ph meter?
Please see:www.adwainstruments.com/site/component/content/article/56
which is the acceptance criteria for slope and offset? does it change if another electrode is used? for example, 1230B
Dear Julian,
The accepted offset is +- 30 mV in 7pH, the accepted slope is between 80 to 110%, if you consider the theoretical slope 59.16 mV/pH@ 25 Celcius degree.