Thank you for the video! Is it a requirement to vent and unload the autoclave the moment it is done? Is it less effective to let it cool for an hour without venting and then taking it out?
You can let it sit or vent it immediately. Either is perfectly fine. It really just depends on when you need the instruments inside. Venting will make the pressure between the chamber and room equivalent and also help in the cooling process. If you're going to let it sit I would suggest popping the door so the temperature can reduce gradually. Just watch out for the steam.
If it is at the beginning of a cycle it could either be tou need to wait longer or you didn't fill the water chamber with enough water. This autoclave takes a lot of water! This same action at the end of the cycle equalizes the sterilization chamber and ambient pressure. It also releases steam out the water chamber top vent so never put your hand there because you'll get burned.
Thank you for the video! You have a knack for it, very enjoyable voice.
Very easy to understand! Thank you.
Thank you for your support!
Thank you for the video! Is it a requirement to vent and unload the autoclave the moment it is done? Is it less effective to let it cool for an hour without venting and then taking it out?
You can let it sit or vent it immediately. Either is perfectly fine. It really just depends on when you need the instruments inside. Venting will make the pressure between the chamber and room equivalent and also help in the cooling process. If you're going to let it sit I would suggest popping the door so the temperature can reduce gradually. Just watch out for the steam.
Release steam , door pops, let sit 15 min to dry.
Thank you for your help!! 😊
Thank you 😊
Thank you for your support!
Gracias! Thanks!
Why is it when press on the fill/vent, water does not fill the chamber. Is something wrong? Ty
If it is at the beginning of a cycle it could either be tou need to wait longer or you didn't fill the water chamber with enough water. This autoclave takes a lot of water! This same action at the end of the cycle equalizes the sterilization chamber and ambient pressure. It also releases steam out the water chamber top vent so never put your hand there because you'll get burned.
Do you have spare parts fot M7.? The cover for the steam chamber breaks off
Try the manufacturer's website.
@@MedicalLabLadyGill Thank you
@@gliceriofuellos2306 you're welcome! I hope you find what you need.
Formerly slow
Thank you 🙏🏿
Thank you 😊