Writing a Custom Affinity Method Using Scouting for the AKTA Avant Chromatography System.

Поделиться
HTML-код
  • Опубликовано: 12 сен 2024
  • In this video we write a Unicorn Method from scratch, without using any of the preset phases. The method includes Scouting and is meant to be run on an AKTA Avant Chromatography system.
    It's a long video so get some popcorn and get ready to nerd out over chromatography. If you ever actually watch it leave a comment, I'd be grateful.
    How to setup and some example results can be found at: • Setting Up a Custom Af...
    One of these days I'll learn how to spell equilibration properly, but I don't think it matters for the purpose of this video to have it absolutely correct.
    Here is the text of the method:
    UNICORN 7.6.0
    Method: MabSelect method
    Text instructions
    Main method:
    0.00 Base: CV, Vc=7.854 {ml}, (Mabselect tricorn column)#Column type
    0.00 Phase: Method Settings
    0.00 Base: SameAsMain
    0.00 Alarm pre column pressure: Enabled, (0.75)#Pre column pressure limit {MPa}, 0.00 {MPa}
    0.00 Alarm delta column pressure: Enabled, 0.75 {MPa}, 0.00 {MPa}
    0.00 Wavelength: (280)#UV1 {nm}, Off {nm}, Off {nm}
    0.00 Noise reduction UV: (0.2)#UV averaging time {sec}
    0.00 Injection valve: Manual load
    0.00 Outlet valve: Out-Waste
    0.00 pH valve: (In-line)#pH cell, (In-line)#Flow restrictor
    0.00 Alarm air sensors: (Enabled)#Air sensor alarm on inlet valve A,
    (Enabled)#Air sensor alarm on inlet valve B, (Disabled)#Air sensor alarm
    on sample inlet valve
    0.00 Inlet A: (A1)#Wash Buffer
    0.00 Inlet B: (B1)#Elution Inlet
    0.00 Pump wash: (Off)#Wash A, (Off)#Wash B, Off, (Buffer)#Wash Sample
    0.00 Block: Mehtod Flowrate
    0.00 Base: SameAsMain
    0.00 System flow: (3.600)#Flow rate {ml/min}, (Pre column pressure)
    #Pressure control
    0.00 End_Block
    0.00 End_Block
    0.00 Phase: Column Equilibration
    0.00 Base: SameAsMain
    0.00 Block: Delay A
    0.00 Base: Volume, Any
    10.00 End_Block
    0.00 Column position: (Position 1)#Column position, Down flow
    (1.00)#Length of Equilibration End_Block
    0.00 Phase: Load Cell Harvest
    0.00 Base: Volume, Any
    0.00 System flow: 0.000 {ml/min}, Off
    0.00 Block: Delay B
    0.00 Base: Time, Any
    0.05 End_Block
    0.00 Injection valve: Direct inject
    0.00 Sample inlet: (S1)#Sample Inlet position
    0.00 Injection mark
    0.00 Sample flow: (3.600)#Sample flowrate {ml/min}, Off
    0.00 Outlet valve: (Out-Waste)#Cell Harvest Waste
    0.00 Hold until: Air sensor Sample inlet valve, Equals, Air, (500.00)
    #Cell Harvest Load volume {base}
    0.02 Block: Reset system
    0.00 Base: SameAsMain
    0.00 Sample flow: 0.000 {ml/min}, Off
    0.00 Block: Delay C
    0.00 Base: Time, Any
    0.05 End_Block
    0.00 Injection valve: Manual load
    0.00 Block: Mehtod Flowrate
    0.00 Base: SameAsMain
    0.00 System flow: (3.600)#Flow rate {ml/min}, (Pre column
    pressure)#Pressure control
    0.00 End_Block
    0.00 End_Block
    0.02 End_Block
    0.00 Phase: Wash Column
    0.00 Base: SameAsMain
    1.00 Outlet valve: Out-Waste
    4.00 End_Block
    0.00 Phase: Elution
    0.00 Base: SameAsMain
    0.00 Gradient: 100.0 {%B}, 0.00 {base}
    0.00 Block: Delay D
    0.00 Base: SameAsMain
    0.10 End_Block
    0.00 Auto zero UV
    0.00 Block: Delay E
    0.00 Base: Time, Any
    0.05 End_Block
    0.00 Peak fractionation parameters: Level, 0.150 {min}, (100.000)#Start
    Peak Frac {mAU}, 100.000 {mAU/min}, (75.000)#End Peak Frac {mAU}, 75.000
    {mAU/min}
    0.00 Peak fractionation: Volume, 15 ml tubes, 12.0 {ml}, Next tube
    3.00 End_Block
    0.00 Phase: Column Cleaning
    0.00 Base: SameAsMain
    0.00 Gradient: 0.0 {%B}, 0.00 {base}
    0.00 Stop peak fractionation
    0.00 Block: (CIP)#Skip or CIP
    0.00 Base: Time, Any
    0.00 System flow: 2.500 {ml/min}, Off
    0.00 Inlet A: (A2)#CIP solution
    30.00 End_Block0.00 End_Block
    0.00 Phase: Column Reequilibration
    0.00 Base: SameAsMain
    0.00 Inlet A: A1
    1.00 Block: Mehtod Flowrate
    0.00 Base: SameAsMain
    0.00 System flow: (3.600)#Flow rate {ml/min}, (Pre column pressure)#Pressure control
    0.00 End_Block
    (5.00)#Length of reequlibration End_Block
    (Unused)
    0.00 Block: Skip
    0.00 Base: SameAsMain
    0.00 End_Block
    A note on delay blocks (if you've come this far): I tested this method and it doesn't look like delay blocks B to E are necessary. These used to be necessary (Unicorn version 3 & 5 needed them), I think the authors of the Unicorn software added some features in either 6 or & 7 that make it less necessary. I decided to leave them in however since I still believe it will increase fidelity and demonstrates problem solving. Take it for what you will.