If I’m not using my nebulizer frequently, is it still necessary to sterilize once a week or just after use? Can I leave accessories in boiling water to clean?
Hi Jasmine, It is recommended to clean and dry the kit only before and after use. Frequent cleaning may reduce service life of the parts. The mouthpiece, medication tank, vaporizer head, top cover are made of polypropylene (PP) and can left in boiling water for cleaning. NEVER clean mask in boiling water as it is made of polyethylene+Polyolefin elastomers and will deform (get spoiled). Please note the PVC air tube and air filter are never washable.
You got a like 👍 on the video. Thanks for saving time on reading long user manual ❤, we got a good start understanding this device. Once again Thanks a lot Bicycle Colours!
Bought this yesterday ..: So I’m putting 1.25 mg levolin in the nebuliser but it only vaporises for 3 minutes … is that okay or there is some issue in the machine? The doctor mentioned it needs to be taken for atleast 10 minutes but the medicine doesn’t fume after 3 minutes and finishes like little bit remains at the end but does not vaporise
Hi, I would suggest to let doctor know about the type of device you are using and ask for recommendation on amount of medication to be used in the device. As I know each brand and type of device has it's own specification.
Juat bought it last night and had to use it eith the medication prescribed by the doctor, the INHALATION TOP as it has two outlets one which is either connected to the face mask or the mouthpiece....after i turn on the steam from the liquid half of it goes away from the top outlet and the other half of the steam comes through the mask or mouthpiece connected....is it normal for the steam to exit from the top open section of the inhalation top or should it just give steam in the mouthpiece or mask area outlet
Hi. I was also facing the same problem then I checked the level of medication that I filled. I found out that I was filling it way above the prescribed level that was leading to vapours being coming out through the top portion. Check the medication level that you are filling in it. It has a mark of maximum level above which the vapour will come out through the top portion.
Hello sir, i hv project on asthma and i m performing experiment on animals so i m going to use this type of nebulizer..... So i can put chemical solutions which i want and vopour can come our or no???? Please suggest something
Hi Pooja, The device will not dispense hot air. The device pulls in room air (at room temperature). If you are operating in an air conditioned room then that could be one reasons for cold air. Please check the tubing for moisture. This may also sometimes result cold air than generally expected.
Hi, If you feel there is much medicine being coming out then the best way to check is to use device with new tank and mask kit. You should be able to get the kit separate as a spare item. Sometimes the fault happens either with medication tank or vaporizer head.
Did your issue get resolved with the steam leaving the top hole because I am facing the same and I started using it since yesterday first time and need to use it twice for more days so can you please guide your observations or any changes you did for the steam to just come from the mask or mouthpiece attached
Hi, I have purchased a nebuliser. I used put a 0.5 mg respule inside the container, which is extremely less quantity. But the mist is coming from the sideways (holes) of the facemask as if its leaking. Please guide.
Hi Sonal, It's recommended to slowly breath in and out through nose. But if the nose is blocked then it will be difficult to breath. Please contact your physician who can evaluate your situation more closely and recommend most appropriate treatment.
Hi Aayushi, The aerosol should only come out of mouth piece or mask. Can you pls check if nothing is loosely connected? If all connections are good and still this problem exists then good to replace attachments.
Hi Looks like the baffle on vaporizer head is broken. The baffle is a small piece of plastic, like tongue. This cannot be fixed. Can you please try using device with new tank and mask kit.
Hi Zaira, Try connecting tube alone and check if device is blowing air. If this is working well then there is some fault with either medication tank or vaporizer head.
If I’m not using my nebulizer frequently, is it still necessary to sterilize once a week or just after use? Can I leave accessories in boiling water to clean?
Hi Jasmine, It is recommended to clean and dry the kit only before and after use. Frequent cleaning may reduce service life of the parts.
The mouthpiece, medication tank, vaporizer head, top cover are made of polypropylene (PP) and can left in boiling water for cleaning. NEVER clean mask in boiling water as it is made of polyethylene+Polyolefin elastomers and will deform (get spoiled).
Please note the PVC air tube and air filter are never washable.
😊
You got a like 👍 on the video. Thanks for saving time on reading long user manual ❤, we got a good start understanding this device. Once again Thanks a lot Bicycle Colours!
Hi Nehal n Kaushik, Glad you liked it!
Perfect video... worth spending time watching, Thanks a lot 🙂
Hi Charu Latha, Glad you liked it! 😊
Bought this yesterday ..: So I’m putting 1.25 mg levolin in the nebuliser but it only vaporises for 3 minutes … is that okay or there is some issue in the machine? The doctor mentioned it needs to be taken for atleast 10 minutes but the medicine doesn’t fume after 3 minutes and finishes like little bit remains at the end but does not vaporise
Hi, I would suggest to let doctor know about the type of device you are using and ask for recommendation on amount of medication to be used in the device. As I know each brand and type of device has it's own specification.
Thank you so much
Hi Sayandas, Glad the video was useful!
Juat bought it last night and had to use it eith the medication prescribed by the doctor, the INHALATION TOP as it has two outlets one which is either connected to the face mask or the mouthpiece....after i turn on the steam from the liquid half of it goes away from the top outlet and the other half of the steam comes through the mask or mouthpiece connected....is it normal for the steam to exit from the top open section of the inhalation top or should it just give steam in the mouthpiece or mask area outlet
Hi, The mist out of sideway holes is usual. These holes are to prevent rebreathing of exhaled gases.
Thank you for the response ,is the mist coming out of the top section is that usual as well
Hi, The mist should only be out from mask and it should not appear from 'Inhalation Top' or 'Medication Tank'.
Hi. I was also facing the same problem then I checked the level of medication that I filled. I found out that I was filling it way above the prescribed level that was leading to vapours being coming out through the top portion. Check the medication level that you are filling in it. It has a mark of maximum level above which the vapour will come out through the top portion.
Does the vaporisor heads rotates when the device is turned on?
Hi Prasad, No, the vaporiser head doesn't rotate.
Sir can i change the medicine chamber which is available in market mask kits.. Is compitable
Hi Souma, I would recommend to purchase Omron brand chamber and mask for Omron nebulizer.
@@bicyclecolours8988 sir this is for my friend... When the medicine chamber got damaged we can replace with other
Hi Souma, You can try... but before buying make sure the tube can be connected between medicine chamber and machine.
@@bicyclecolours8988 thank u sir
Cheers Souma! :-)
Keep me posted if you managed to get this sorted for your friend.
Hello sir, i hv project on asthma and i m performing experiment on animals so i m going to use this type of nebulizer..... So i can put chemical solutions which i want and vopour can come our or no???? Please suggest something
Please sir give me reply for my message????
Please sir give me reply for my message????
Hi Snehal, Yes, this instrument can be used to produce 'MIST' but you may need to search something else if you are looking to generate 'VAPOUR'.
Hi Snehal, If you are looking for a device which can only generate vapour then Vaporizers could be a good choice.
I am getting cold air from the nebuliser - is it normal ? Generally you need hot air right
Hi Pooja, The device will not dispense hot air. The device pulls in room air (at room temperature). If you are operating in an air conditioned room then that could be one reasons for cold air. Please check the tubing for moisture. This may also sometimes result cold air than generally expected.
Thank u
Hi, You are welcome :-)
My nebulizer is not vapourising and clear what is fault i purchased 2 days before i was making a trial . Can we replace or it can be solved
Hi Manzoor, I would recommend get the instrument replaced under warranty. It would be extra cost if you try to fix the device.
Too much medicine is going out through the top hole is it ok or mistake?
Hi, If you feel there is much medicine being coming out then the best way to check is to use device with new tank and mask kit. You should be able to get the kit separate as a spare item. Sometimes the fault happens either with medication tank or vaporizer head.
@@bicyclecolours8988Sir please tell which mask kit is compaitable for this
Facing the same issue
Did your issue get resolved with the steam leaving the top hole because I am facing the same and I started using it since yesterday first time and need to use it twice for more days so can you please guide your observations or any changes you did for the steam to just come from the mask or mouthpiece attached
Hi, I have purchased a nebuliser. I used put a 0.5 mg respule inside the container, which is extremely less quantity. But the mist is coming from the sideways (holes) of the facemask as if its leaking. Please guide.
Hi, The mist out of sideway holes is usual. These holes are to prevent rebreathing of exhaled gases.
Breath as usual but slightly deeper so that medicine reaches your lungs. Excess air goes out, it is normal
It takes only a minute to end the repsule. Is it normal?
Nebulization lete samey nose se Lena hai aur mouth se saas chorna hai kya??
Hi Sonal, It's recommended to slowly breath in and out through nose. But if the nose is blocked then it will be difficult to breath. Please contact your physician who can evaluate your situation more closely and recommend most appropriate treatment.
Inhalation tank se bhi steam bahar ari hai is this normal?? I have purchased new machine today
Mouth pipe se nebuliser karte time inhalation tank top se bhi ari hai steam
Hi Aayushi, The aerosol should only come out of mouth piece or mask. Can you pls check if nothing is loosely connected? If all connections are good and still this problem exists then good to replace attachments.
can i buy the nebulizer head from somewhere?
Hi, I would like to know how much quantity of medication in ml we can put for 6 months old?
Hi Anand, Please contact doctor to receive advice on medication type and quantity.
Hello mine has cold mist , what do I do?
Hey, the mist is always cold. The device doesn't produce warm mist.
I want to buy best nebulizer for my baby,i have 2 option Medtech Handyneb Smart and Omron NE C106... Who is best ?
Hi Mahesh, Comparing the specifications both brands seems to be equivalent.
The best is to go with recommendation from consulting doctor.
Last mai ye khali ho jana chahiye khya ,ya fir medicine thodi bahot bachti hai
Hi, Please have doctor advice on quantity of medication to be used in nebulizer.
Thank you.
Steam comes from the top of the tank and not from the mask
Solution??
Hi Looks like the baffle on vaporizer head is broken. The baffle is a small piece of plastic, like tongue. This cannot be fixed. Can you please try using device with new tank and mask kit.
put less medicine in the container
When should i use air filter
Hi, The air filter should be always installed when using the device. Pls make sure the air filter is clean and not damp.
Paani dalta h kya isme steam ke liye
Sorry, this device is not meant for steam inhalation.
Sir nebulizer suddenly not working sir.1 year aaguthu sir
Hi Aswinth, If the device is still under warranty kindly contact Omron customer support. Hope the issue gets sorted soon.
मेरी नेबुलीज़र मशीन काम नहीं कर रही हैं, जो मैंने थोड़े समय पहले ली थी। क्या कमी या खराबी हो गई हैं,कृपया बताएं?
Hi Srikant, Can I know what type of issue you are experiencing with the device?
The best is to get the device replaced if it is under warranty.
@@bicyclecolours8988 yes , under warranty है
Replace का क्या process होगा।
Hi Srikant, You need to call the customer support for your country given on your package/warranty card and they will guide you through the process.
Where to get filter
Hi, The model code for NE-C106 air filter is “NEB-AFR-30AP”. Hope you can use this code to search and order on online / in store.
தமிழில் வீடியோ போடவும்
Dear Rajasekarn, Yes, sure... will do soon.
Thank you!
El mío no avienta humo 😢
Hi Zaira, Try connecting tube alone and check if device is blowing air. If this is working well then there is some fault with either medication tank or vaporizer head.
Isme dawa. Kaunse dalte
Hi, The medicine to be the one prescribed by doctor.