    Each calibration kit comes with calibration solutions which come in 5 single-use, glass vials and a detailed calibration certificate. Standard calibration points for the AW probe are 0, 10, 35, 80 and 95% RH.
    Step one: Collect all of the as found calibration points and save each test point.
    Step two: Apply the calibration points if desired.
    It's important to note that the instrument and standards are at the same ambient temperature (avoid direct sunlight and ventilation vents) so the calibration is completed with the utmost accuracy.
    Fast forward to 3:07 to see how to perform a calibration with the HP23-AW psychrometer.