Biometry-2 | Measurement of Axial length | Smart Optometry

Поделиться
HTML-код
  • Опубликовано: 11 сен 2024
  • AXIAL LENGTH: Axial length is the distance between anterior and posterior pole of eyeball
     At birth 17 mm
     In adult 24 mm
     In myopia more than 24 mm
     In hyperopia less than 24 mm
    METHODS: There are two methods
    A. Ultrasonic
    B. Optical
    ULTRASONIC METHOD:
    i. It contain a 10MHz solid probe with inbuilt fixation
    ii. It has to be aligned with optical axis of the eye for accurate measurement
    iii. This can be done in two technique :
    A. Immersion technique
    B. Contact technique
    1. IMMERSION TECHNIQUE:
    --------Patient lie in supine position looking a fixation at the ceiling
    --------A scleral shell is applied to the eye filled with 1% OR 2% Methylcellulose
    --------Probe is Immersed in the solution keeping 5-10 mm away from the cornea
    --------Probe is gently moved intil properly aligned with optical axis
    --------Echo spike will displayed in the monitor.
    ECHO SPIKES ARE:-
    --- IS = INITIAL SPIKE PRODUCED BY PROBE
    --- C1 = ANTERIOR CORNEAL SPIKE
    --- C2 = POSTERIOR CORNEAL SPIKE
    --- L1 = ANTERIOR LENS SPIKE
    --- L2 = POSTERIOR LENS SPIKE
    --- R = RETINAL SPIKE
    --- S = SCLERAL SPIKE
    --- O = ORBITAL SPIKE
    2. CONTACT TECHNIQUE:
    i. Alternative to immersion technique doesn’t use scleral shell
    ii. Probe directly contact with cornea
    iii. These can be done in two methods:
    1. Hand held
    2. Attaching with slit lamp or applanation tonometer
    PROCEDURE:
    ------ Patients seated after applying local anesthetic drop
    ------ Asked to fixed a target straight with non-testing eye
    ------ Probe is properly aligned along with visual axis & brought forward to touch the cornea
    ------ Five spike will be displayed on the screen.
    ECHO SPIKES ARE:
    --- C = POSTERIOR COrNEAL SPIKE
    --- A = ANTERIOR LENS SPIKE
    --- P = POSTERIOR LENS SPIKE
    --- R = RETINAL SPIKE
    --- S = SCLERAL SPIKE
    B.. OPTICAL TECHNIQUE:
    i. Use partial coherence laser
    ii. Biometer measure the time required for infrared light to travel to retina
    iii. This technique doesn’t have contact with cornea, so corneal compression
    artifacts are eliminated.
    FOR DETAILS NOTE;
    VISIT OUR WEBSITE: www.smartopto....
    --------------------------------------------------------------------------------------------------------------------
    Binocular Subjective Refraction:
    Part-1: • Binocular Subjective r...
    Part-2: • Septum Method - Turvil...
    Part -3: • Polarization Method - ...
    Part -4: • Fogging Method - Binoc...
    Part-5: • Steps of Binocular Sub...
    Biometry:
    Part-1: • Introduction Of Biomet...
    Paet-2: • Biometry-2 | Measureme...
    Part-3 • Biometry-3 | Measureme...
    Optical Coherence Topography:
    Full: • Optical Coherence Tomo...
    Part-1: • Optical Coherence Tomo...
    Part-2: • Optical Coherence Tomo...
    Dry Eye:
    Part-1: • Dry Eye - Introduction...
    Part -2: • Dry Eye - Classificati...
    Part -3: • Dry Eye - Clinical Fea...
    Part -4: • Dry Eye - Treatment (w...
    Part -5: • Tear Film Break-up Tim...
    Part- 6: • Schirmer test - Dry Ey...
    Full: • Dry Eye - Full Tutorial
    Myopia
    Part-1: • Myopia - Introduction ...
    Part-2: / dvrkegdkxq
    Part-3: • Myopia - Clinical Pict...
    Part -4: • Myopia - Ocular examin...
    Part -5: • Myopia - Management or...

Комментарии • 25