#03 Rigol DM858 (E) DCV failure - Part 3

Поделиться
HTML-код
  • Опубликовано: 23 авг 2024
  • Rigol DCV failure in 01.00.00.19 and .20 version (at least).
    Will occur in DCV Auto mode. Scenario is as follows:
    1) free floating leads connected to DMM, while DMM setting in DCV (AUTO)
    - this result in normal random small voltage in mV range (DMM switch in 1V or 100mV automatically)
    2) Switch on Power supply with 33 V DC
    3) Connect DMM leads directly on 33V DC outlets
    - DMM remains in mV/V scale while consuming ~ 25 mA (Wrong situation)
    - It stays in this faulty situation even if the Voltage is changing to lower (often 28V) and (38.. and sometimes even 60V). It looks likes there is a temperature component in the game. I have got different faulty ranges - sometimes is narrows it from 28-60 to 28-38 Volts.

Комментарии • 30

  • @Recovery_Drone
    @Recovery_Drone 4 месяца назад +2

    Uma pena. Iria comprar um essa semana, vou aguardar novas atualizações :(

  • @M-P-X
    @M-P-X 4 месяца назад

    Vette bug zeg, dit verwacht je niet van Rigol, ik blijf dit volgen.

  • @frankkudelsko1561
    @frankkudelsko1561 4 месяца назад +1

    I ordered the DM858 10 days ago from Conrad (France). Received the product not before the end of April. I will update to v00.01.00.00.20 dated 2024_03_28. We'll see if I find the same problem. This bug is totally unacceptable from a manufacturer like Rigol.

    • @marcelmostert3081
      @marcelmostert3081  4 месяца назад

      I am very interested to see if the problem is still there or not on your future DM858. Version 1.00.00.20 didn’t provide a solution for this bug (I retested it). I am afraid that it might be hardware related. It looks like the protection mechanism is kicking in before the Autorange can scale up. I would assume that the protection mechanism is not relying on Software. But we will see how Rigol will solve this problem.

    • @frankkudelsko1561
      @frankkudelsko1561 4 месяца назад

      ​@@marcelmostert3081 I received my DM858 much sooner than expected. I did the last firmware update 00.01.00.00.20 and I get exactly the same bugs. In short, we're waiting for the next update, which will correct the problems. Bad point for Rigol. 😒

    • @marcelmostert3081
      @marcelmostert3081  4 месяца назад +1

      @@frankkudelsko1561 Thanks for checking. Good to know that you have the same bugs - so irrelevant if it is a 858 or 858E model. Meanwhile I received the info from my supplier that they have found the right person of Rigol to take care about the DCV-failure topic. But no update received yet.

    • @frankkudelsko1561
      @frankkudelsko1561 4 месяца назад +1

      @@marcelmostert3081 Good news if your supplier has found a contact at Rigol. Now it's going to take some time for Rigol's engineers to provide a fix for the various bugs described in your videos.

    • @ilex2288
      @ilex2288 4 месяца назад

      Hola, mi pregunta general... El mío falló cuando hice medición de 380v. En función auto rango no me funciona, pero si lo cambio a manual si me funciona, tampoco me funcionó ya la función de Ω , diodo, capacitor ..... Me pide el vendedor que se lo regrese pero desde México a China lo veo difícil

  • @FloppyDos
    @FloppyDos 4 месяца назад +1

    Yep same bug, tested on a lineaire power supply and a smps. Latest firmware. DM 585

    • @FloppyDos
      @FloppyDos 3 месяца назад +1

      Extra info...if I try a couple of times, it start to work. Not good though

    • @marcelmostert3081
      @marcelmostert3081  2 месяца назад

      I noticed that when the device is longer switched-on the problem - or the voltage range in error [28V--50V] - becomes smaller) -> Some temperature effect?

  • @juliorrsantos
    @juliorrsantos 16 дней назад

    Any news about this problem? I've just received my DM858E and it's with the same bug.

    • @marcelmostert3081
      @marcelmostert3081  15 дней назад

      Hi, did you receive a HW1.0.0 version?

    • @juliorrsantos
      @juliorrsantos 15 дней назад

      @@marcelmostert3081 it's a "Hardware version 1" not even 1.0.0

    • @juliorrsantos
      @juliorrsantos 15 дней назад

      I would be nice if they publish what we could do to reproduce the fix by ourselves

    • @marcelmostert3081
      @marcelmostert3081  7 дней назад

      yes, communication can be better. Via SCPI command SYSTEM:CFGVER 1 the problem looks solved. Although I have noticed it is not 100%. I didn't made a remark in my latest HW buglist video - but I am able to revive the same bug in a special scenario. But in the more normal cases it is more or less "solved"

  • @jacirp.r1277
    @jacirp.r1277 4 месяца назад

    talvez tenha problemas com cabo de força, pois no video eu acho que deveria inverter os cabos do own com rigol.

    • @marcelmostert3081
      @marcelmostert3081  4 месяца назад

      I have changed the cables - not showed on the video - this gave the same wrong result. From another user who has also the DM858 (not E version) does have the same problem. So unfortunately, it is really a bug. Question now is: Is it hardware or software.

  • @ilex2288
    @ilex2288 4 месяца назад +1

    Tengo el DM858E y me falló cuando hize la medición de 380v en fuente PFC. Solo funciona en función manual (en auto no funciona) y las funciones de Ω, diodo capacitor etc.. no funcionan😢

    • @marcelmostert3081
      @marcelmostert3081  Месяц назад +1

      Good evening, I think I found the root cause of your corrupted DM858E. It appears that the impedance is very low for a short duration (and strangely enough how higher the DC, how lower the impedance). I measured 1500 ohm at 56-66Volt. But with 380 DC, I don't know, but I guess it destroys your multimeter. This situation of low duration is only for around 150-220ms and occurs when the auto-range has to scale up from

    • @ilex2288
      @ilex2288 Месяц назад

      @@marcelmostert3081 muchas gracias, me dieron garantía y hubo cambio de equipo, ahora con el nuevo la función DC la uso manual, no automática.

    • @marcelmostert3081
      @marcelmostert3081  Месяц назад +1

      @@ilex2288 Also thanks. Indeed, not using the Auto range function is the best option. Have a nice evening.

  • @ssalient
    @ssalient 2 месяца назад

    This is not great from a brand like this.
    What I also find "interesting" is that these newer models (over the DM3058[E]) the counts are lower. This seems to be a trend with Rigol? the DM3055 model had 480.000 counts, the later 3058 240.000 (or at least aound 200.000 counts) and now this latest model only 120.000. This fact alone is a big turn off for me already.

    • @marcelmostert3081
      @marcelmostert3081  2 месяца назад

      Yes, Marketing calls it 5 1/2 digit :) But it is only 20% over 5 digits. It is the bare minimum now. Wondering if Rigol 6 digits (6 1/2) will come out in this form factor - they have first to do a lot of HW reworks.

  • @BJcanal270
    @BJcanal270 4 месяца назад

    My works perfect. 🤔

  • @wilsmit7322
    @wilsmit7322 4 месяца назад +1

    Reinstall firmware if possible ....

    • @marcelmostert3081
      @marcelmostert3081  4 месяца назад +4

      I have upgraded to the latest release (.20) and the bug is still there.