Ion Chambers and Reference Dosimetry. By: Thomas Milan

Поделиться
HTML-код
  • Опубликовано: 9 фев 2025

Комментарии • 5

  • @medicalphysicsuwa
    @medicalphysicsuwa  3 года назад +1

    Erratum:
    On slide 7 a graphite calorimeter is pictured, but incorrectly labelled as a graphite cavity chamber. At ARPANSA:
    · the graphite calorimeter (pictured) is an absorbed dose to water standard for Co-60/MV
    · the graphite cavity chamber (not pictured) is an air kerma in air standard for Co-60

  • @rimelius
    @rimelius 2 года назад

    Thank you for this really useful video

  • @new.me197
    @new.me197 Год назад

    I like it

  • @melaniaistrati6778
    @melaniaistrati6778 3 года назад

    Hi, can you tell me where did you got the formula for the quality factor Kq = a (TPR-20.10) + b (TPR-20.10) + c and the values a, b, c? Maybe a link or something? Thank you very much!

    • @medicalphysicsuwa
      @medicalphysicsuwa  3 года назад +1

      This formula is provided by the calibration lab for the chamber calibrated for a centre. It is chamber specific, so the a, b, c for our camber would not be the same as your chamber.