Foot Pedal Rebuild Training #1

Поделиться
HTML-код
  • Опубликовано: 29 авг 2024
  • In this video, we demonstrate a foot pedal repair using a DCI foot pedal rebuild kit.
    These videos are for entertainment purposes only. Do not attempt to perform any service without proper training, service literature or qualified service technicians. The Dental Equipment Repair Channel and Bright Idea Dental Repair assume no responsibility for any damages or harm that may result from working on hazardous equipment. USE AT YOUR OWN RISK.

Комментарии • 39

  • @abrilforero1978
    @abrilforero1978 Год назад +4

    SUPER HAPPY, I CAN SMILE AGAIN AFTER HAVING CHECKED AND REPAIRED MY PEDAL, GREAT VIDEO!

  • @user-gr6kg6zx6x
    @user-gr6kg6zx6x 9 месяцев назад +2

    thank you for making it so simple

  • @nguyenthanhnha1234
    @nguyenthanhnha1234 Год назад +1

    I'm a technician from Vietnam who specializes in repairing Japanese, American, German, Chinese.... looking forward to one day exchanging experiences to share with you.

  • @adelpayghani4216
    @adelpayghani4216 Месяц назад +1

    very good

  • @HTMWorkshop
    @HTMWorkshop 2 года назад

    Great video! Glad I found it

  • @nathanielgalimba4470
    @nathanielgalimba4470 2 года назад +1

    thank you !

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  2 года назад +1

      You are welcome! Where are you from?

    • @nathanielgalimba4470
      @nathanielgalimba4470 2 года назад

      @@DentalEquipmentRepairChannel I am from the Philippines!!... thank you for sharing all the techniques and know-hows!! Keep safe!

  • @ratandmonkey2982
    @ratandmonkey2982 Год назад +2

    nice video. My rebuild didn't fix the problem. Still leaking air from the poppet valve. Guess I was just unlucky. Had to buy a new pedal.

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  Год назад

      Sometimes that's the way it goes.

    • @ratandmonkey2982
      @ratandmonkey2982 Год назад

      brand new pedal doesn't work either ! Think regulator might be under-pressured. Working on it.

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  Год назад

      @@ratandmonkey2982 interesting. What pressure is coming into unit? Can you fully describe problem? There should be full pressure (80 psi) on one large tube going to pedal and the other should blast air out when the pedal is pushed down. What make/model of delivery unit do you have?

    • @ratandmonkey2982
      @ratandmonkey2982 Год назад +1

      @@DentalEquipmentRepairChannel I wish I had seen your reply days ago. Air from the drive air line to the foot pedal (air from the control head to the pedal) was moving in reverse direction and leaking out the foot pedal valve when the foot pedal was not touched at all. It was not a leak of air coming from the air regulator. I believe I fixed it. Through a lot of diagnostic work, I seemed to trace the problem to the master control block. It was an ASI "advanced endodontics system: delivery unit coupled with an A-Dec chair. I replaced all the "pucks" in the control block and the leak stopped. None of the pucks appeared damaged, but they were pretty well indented through use. Hopefully that keeps things fixed for a while. What a nightmare.

  • @tazbirulalam
    @tazbirulalam 4 месяца назад

  • @YounisAlhussini
    @YounisAlhussini Год назад +1

    Very excellent

  • @abrilforero1978
    @abrilforero1978 Год назад +1

    QUESTION, and how do you know the pedal number in order to buy the kit?, my says IPX1, but nothing else, thanks

  • @ilyashoshana3950
    @ilyashoshana3950 11 месяцев назад

    Hi I help to my friend dentist to relocate
    And connect the chairs
    Water and air connected but system loosing
    Air after pressing pedal please advise
    Thanks
    Ilya

  • @irmaotero6176
    @irmaotero6176 Год назад

    I have a situation. I can't find a video where anyone goes further down in the toggle repair. Passed the spring and toggle. Mine broke and more parts came out. There an extra pin and a smaller tiny spring as well. Don't know how that goes in first. Please help.

  • @shihabahmed2620
    @shihabahmed2620 Год назад

    Why the water in high speed handpeice doesn’t stop immediately ? Is the cause in foot control?

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  Год назад

      There could be residual pressure in the signal line from the foot pedal. The pressure does need to bleed off in order for the water to stop.

  • @user-bf6gb1zy2b
    @user-bf6gb1zy2b 2 года назад +1

    Cause of water entering the pedal of the dental unit

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  2 года назад

      I'm sorry I missed this one. That's a hard issue. You would have to be getting water bypassing at the master block somehow. Perhaps there is moisture in the main air line from the compressor. No water should be going to the foot pedal, only the 1/4" air line.

    • @fransiscoscaramanga674
      @fransiscoscaramanga674 2 года назад

      if the compressor is not compromised, no moisture in the tank.... the water must be coming from h/p block, especially if it's an adec unit.... replace the side clear gaskets..

  • @user-bf6gb1zy2b
    @user-bf6gb1zy2b 2 года назад +1

    Please tell, if you know, Farban

  • @abdoimrane1611
    @abdoimrane1611 Год назад

    Hello, can you show me how you disassebled the handpiece nsk s-max m65

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  Год назад

      Hi! I don't have this handpiece to teach with. Call these guys... premiumhandpieceparts.com/?creative=648025185970&device=m&placement&gad=1&gclid=EAIaIQobChMIqZiYzJbc_wIVldjICh23MAv5EAAYASAAEgKU5_D_BwE

  • @user-bf6gb1zy2b
    @user-bf6gb1zy2b 2 года назад

    Master, what is the solution to this problem?

  • @user-bf6gb1zy2b
    @user-bf6gb1zy2b 2 года назад

    Cause of water entering the pedal of the dental unit ؟؟

    • @DentalEquipmentRepairChannel
      @DentalEquipmentRepairChannel  2 года назад

      The one incoming hose on this pedal is from drive air (1/4") and the two outgoing hoses are handpiece (1/4") and chip air (1/8"). The water would have to be coming from the air inlet (water in air lines) or you have something connected wrong. It could also be a leak in the control block in the delivery unit head. Maybe a bad gasket? Strange things can happen when the gaskets go bad.
      You can check for water in air lines by blowing the air water syringe AIR onto a mirror. It should be dry. If you have moisture in air lines, could be something else with compressor or bad water master valve allowing water to go into your air lines. .
      This is a tough problem that might take a tech onsite to fix.

    • @user-bf6gb1zy2b
      @user-bf6gb1zy2b 2 года назад

      @@DentalEquipmentRepairChannel Thank you for your explanation, dear brother

    • @user-bf6gb1zy2b
      @user-bf6gb1zy2b 2 года назад

      Master, what is the solution to this problem?