Hello i tested contacts in machine in put where we connect ecg cable for electrodes they are working but i cant find what the problem . But when i use for patients they dont give any signal . Do you hava anything in your mind what might be problem ? And thanks for valuable information
Please sir, my Contec ECG show time out with red bar and can’t print ECG On patient.I already delete patient data if memory full ..but it didn’t disappear and can’t take ECG more What shall I do .
Just Printing single channel print in 1 line, 3 channel prints 3 leads at a time, and like wise 12 channel prints 12 lines together so its easy to compare the leads in the same line
The difference between 3-channel and 12-channel ECG (Electrocardiogram) machines lies in the number of leads and the amount of information detected, enhanced and recorded, on real time. the design of the three channel ECG, the machine activates 3 leads at the time of detecting and recording. A 12-channel ECG has 10 electrodes that are placed on the patient's limbs and chest to record the electrical activity of the heart from 12 different perspectives . The 12 leads include standard limb leads (I, II, III), augmented limb leads (aVR, aVL, aVF), and precordial chest leads (V1 through V6).
Earthing cableshould be connected to a proper earthing provided by an electrician because you will not get a proper grounding through your medical bed because Bed is having PVC or a plastic shoes in each leg of the bed so you will not get the grounding ultimately you have to connect this ground lead to the proper grounding or earthing provided by electric board
No this machine works on thermal print , it does not have toner, its chemical coated paper where printer applies heat to print on the paper. . please revers the paper or try to use good quality papers.
Hello i tested contacts in machine in put where we connect ecg cable for electrodes they are working but i cant find what the problem . But when i use for patients they dont give any signal . Do you hava anything in your mind what might be problem ?
And thanks for valuable information
Clear cut concept , helpful
Thank you for watching
Informative video Thank you
Glad it was helpful!
Printing problem is there. Only few lines are being printed, not the entire waveform. What to do ?
Please sir, my Contec ECG show time out with red bar and can’t print ECG On patient.I already delete patient data if memory full ..but it didn’t disappear and can’t take ECG more
What shall I do .
Whats the practical diffrence between 3 channel vs 6 channel vs 12 channel
Just Printing single channel print in 1 line, 3 channel prints 3 leads at a time, and like wise 12 channel prints 12 lines together so its easy to compare the leads in the same line
The difference between 3-channel and 12-channel ECG (Electrocardiogram) machines lies in the number of leads and the amount of information detected, enhanced and recorded, on real time. the design of the three channel ECG, the machine activates 3 leads at the time of detecting and recording. A 12-channel ECG has 10 electrodes that are placed on the patient's limbs and chest to record the electrical activity of the heart from 12 different perspectives . The 12 leads include standard limb leads (I, II, III), augmented limb leads (aVR, aVL, aVF), and precordial chest leads (V1 through V6).
Where must earthing cable be put? Is it on the metal part of bed? How to get grounding?
Earthing cableshould be connected to a proper earthing provided by an electrician because you will not get a proper grounding through your medical bed because Bed is having PVC or a plastic shoes in each leg of the bed so you will not get the grounding ultimately you have to connect this ground lead to the proper grounding or earthing provided by electric board
Shouldn't the neutral lead be connected to right leg @@oticameditronix5508
Can we connect with a metal piece that is connected to the ground??@@oticameditronix5508
What's is normal setting that's are using in normal time
Make settings to default. It will be work for regular use
Thank you is good information ❤
Name of the software to connect to the pc?
Our ecg mechine is not switch on .....even after connecting to switch board ... can u suggest any tips to on the machine
Replace the battery or fuse or cable. And see if anything. Otherwise need to send it to service centre
Hello, can i use my own code to monitor the data in my software (if SDK exist)
No
This is only giving rhythm leads but i want to take full leads in auto system in which function should i go??
You have to change the print an sample settings from the menu.
Printer prints but without ink on the paper. Is there a toner?
No this machine works on thermal print , it does not have toner, its chemical coated paper where printer applies heat to print on the paper. . please revers the paper or try to use good quality papers.
Esme automatic reading hota h ?
Yes.
do you have pc software installer?
yes. please contact Oticamart.com
How to connect it with computer by usb cable and for what?
You can connect it with USB CABLE. BUT FOR interface you need to buy software from the company
@@oticameditronix5508 software kitna me aayega
ECG standarization k liye height 1mv set nh horhi h 0.5aa rha h 1 mv kaisa set kare
Settings ne jake. Sample settings me ja Ke change karna he
My machine showing time out so what I do ?
Check the ECG CABLE.
My device shows pending ECG.wat to do.
Please help
Show it to instrument technician
What is reason for pr depression in report
Please check you have connected Green Grounding cable to proper earthing.
Which is best mode
CMS 300 G
What is the rythm lead for 3 channel lead
you can chose from the setting its lead II
How to connect it to computer?
Also wondering.?????
Good information
So nice of you
Why the machine could not turn on? Even not charging the batteries...
Please visit nearby Medical Equipment dealer or from where you have purchased
Sir E C G timeout information
What about the filters button?
Computer me connect kaise hoga ?
Apki software purchase karna hoga
I want catridge
Thank you sir
I need this book....please help
Visit Oticamart.com
How to set the date & time
Go to settings and you can change date and time from there
Cost please
35000 inr
how to get long lead II
Go to settings. Then. Sample setting and select. Continues lead. And u will get it
I cannot find lead 11
How to change it its only 3,6,leads
Interpretation and HR etc nahi aa raha
Contact your supplier
Touch not working
Please contact your machine provider. U need to change that touch panel
Thanku
Welcome
Hi everybody
Hello
Hou macha cash
I want
Explai SEN setting abit more
yes. please contact Oticamart.com for explainations
Sir ECG timeout information please sir
Go to settings do all settings in default mode. And retry. If issue still persist you should contact service provider
phone no.
8530346465
My ecg machine doesnt show conclusion..so how can i set it for conclusion
Go to settings. Then go in sample settings and then start interpretation