HOW TO CREATE HPLC SEQUENCE FILE ? | EMPOWER 3 SOFTWARE
HTML-код
- Опубликовано: 23 май 2024
- HOW TO CREATE HPLC SEQUENCE FILE ? | EMPOWER 3 SOFTWARE | HINDI
your quires;
this video based on HPLC sequence file creation.
in this video explained how to create HPLC sequence of by empower 3 software.
in this video also explained about important point about HPLC sequence like how to save properly with tracibility.
HPLC
HPLC sequence file
sequence file
sequence
file
HPLC method file
empower 3 software
vial no
HPLC vial
number of injection
replicate
standard replicate
HPLC troubleshooting
HPLC interview questions and answers
QC Department interview questions and answers
interview
question
answer
_____________________________________________
#hplc #sequence #troubleshooting #interviewofpharmaindustry #industry
#prof.karanajaygupta #EnglishExcel2208 #ASAlok #axionlabs #QualityHUBIndia #pharmaway #baayo #bhavyamangla #mjdchemistry #shomusbiology #zendesk
_____________________________________________
follow on Instagram;
/ safdarkhankayani
like us on Facebook;
safdar.khank...
subscribe for more videos;
/ @voiceofkayani5419
linkdin;
www.linkedin.com/help/linkedi...
email I'd;
chemist.kayani143@gmail.com
Best n informative video , sir
Thnk you , sir
Hlo .....I am having a qus that there is +_ 15 days for calibration of HPLC ...in case of delay calibration how to handle it .....best practices in industry
1. Where should we enter it in annual calibration plan eg . If we need to do it on 30 March but due to some reason we performed on 7th April then should we write in April month or March month in annual plan
2. The next calibration date ...will it include delay time...
Dear,
Delay time calibration like +/- 15 day is not the best practices.,..if calibration is delayed, cannot be used instrument during due date calibration
Suppose due was on 10 March but but didn't perform and you performed calibration on 20 match. It's ok .. but instrument cannot be used during 10 to 19 march for any analysis.. and for next calibration date will be count from 20 march on document..
That is çlear justification....
If any quries or not clear, then I will discuss in video for next
Thanks a lot .....
Pls create video on validation method parameter calculation like accuracy (80, 90, 110 % recovery)
@@voiceofkayani5419sir thanks for your kind & quick reply .....sir please discuss GCHS parameter & their impact on analysis like hold time. , ramp, final hold time , thermostat time , transfer line temp , pressurize time , withdrawal time
Sir, result and report kaise check krte hai
Kis ki?
@@voiceofkayani5419 waters alliance hplc pr sample run krne ke baad