Case 164: Manual of PCI - Wire fracture

Поделиться
HTML-код
  • Опубликовано: 4 окт 2024
  • A patient with prior TAVR with a Sapien valve was referred for PCI due to angina. He had left main as well and mid and distal RCA heavily calcified lesions. We decided to first perform PCI of the RCA lesions. Right heart catheterization showed a right atrial pressure of 7 mmHg and pulmonary artery pressure of 41/15 mmHg, hence no hemodynamic support was used. We wired the RCA with a workhorse wire but a Turnpike Spiral and Turnpike LP microcatheter could not cross the mid RCA lesion. We removed the workhorse wire and were able to advance a Viper Flex tip wire to the right PDA. Multiple rounds of orbital atherectomy were performed. The Viper Flex tip wire became twisted distally and could not be withdrawn. We advanced the Turnpike catheter to the distal segment of the Viper wire and were then able to retrieve the wire but the tip fractured and remained in the vessel. IVUS did not show wire unraveling and comparison with a new Viper Flex tip wire showed that the fracture was at the distal 3 cm. We did multiple retrieval attempts using the “knuckle twister” technique, twirling wires and snares (Gooseneck and 2-4 mm EnSnare) without success. We abandoned wire retrieval efforts and predilated and stented the right coronary artery, trapping the wire fragment behind the stent struts with a nice final result. The left main intervention was staged for a later date.

Комментарии • 7

  • @shangz0216
    @shangz0216 11 месяцев назад

    Thanks for the educative case presentation.

  • @tom11298
    @tom11298 11 месяцев назад +3

    I understand it's easier to suggest things in hindsight, but wouldn't it be helpful to use a manual or mechanical thrombus aspiration device to retrieve/aspirate the lost part of the guidewire? Thanks

    • @manosbrilakis
      @manosbrilakis  11 месяцев назад +1

      I have not done it but sounds like a great idea!

  • @endosurgeon5275
    @endosurgeon5275 11 месяцев назад

    Would to see how to remove stent migrated from balloon inside the coronary artery. I have one this case and can give you for other peoples

  • @maisamtaherian986
    @maisamtaherian986 11 месяцев назад

    Thank you very much. Would you please explain what to do in case of wire unraveling which extended to ascending Aorta?

    • @ivankoshkin837
      @ivankoshkin837 11 месяцев назад +1

      It becomes an open surgery case. Unfortunately, had that in our clinic.

    • @manosbrilakis
      @manosbrilakis  11 месяцев назад +2

      Surgery