Common issue of hypervisor mindray CMS try to discharge the patient on the monitor and restart the monitor and admit patient again.. if still idle restart the CMS..
Pacetech Vital max 4000 monitor? There is no reset in the menu, try to remove the Nvram in the main board and wait for 5 seconds and return it back again.
@@roysantik Yes vital max 4000 monitor I have problems one message display on screen (made in USA) and I can't to access the parameters ,how I can to repair
Big...fan sir keep up the good work ✌️
Saludo po sir pinoy pala kau..
I have a pdj 3000 patient monitor which remains stuck at the boot menu please help me thank you
If stuck in the logo you need to update the software or reinstall the software again.
@@roysantik Can you send me software ans update instruction
Sir cmc monitor showing idle sector 9 monitors are connected but 3 are showing idle sector what to do
Common issue of hypervisor mindray CMS try to discharge the patient on the monitor and restart the monitor and admit patient again.. if still idle restart the CMS..
Sir, Error 28 is a Peak voltage reading problem, can you check the mosfet transistors.
sir what is the software upgrdaing tool? which software we need to update software?
I replaced cpu board and monitor turned onn but when I shut it down it does not get started what to do now plze tell
Try to remove the cmos battery in the cpu board in 10sec and return it back again..
HELLO, MY FRIEND, THE SAME DEVICE OPERATES LATELY, WHAT CAN I DO? CAN YOU SHARE THE INSTALLATION SOFTWARE?
do you have a cpu board but no Software was install? I have an installation software, how i can send it to you?
Sir please how can I get the software
Sir if you have access in Mindray portal you can download the software...
Tellme what to do sir
How can repair display whiteness problem
whiteness problem of the patient monitor mostly defective display board. change display board.
How can reset Vitalmax 4000
Pacetech Vital max 4000 monitor? There is no reset in the menu, try to remove the Nvram in the main board and wait for 5 seconds and return it back again.
@@roysantik
Yes vital max 4000 monitor I have problems one message display on screen (made in USA) and I can't to access the parameters ,how I can to repair
Mindray pm9000E spo2 comm stop
Try to check the flex cable harness or the spo2 module board.
@@roysantik modul board spo2 has been replaced with a new one