Hi, Yes, the Densito can be used in battery production, for example, to measure the density of electrolytes. The instrument also allows temperature compensation of the measured values to your temperature of interest.
Thank you for your message, please find information on compatibility under www.mt.com/DERE-norms or ask a local representative. Contact details can be found at www.mt.com under "Contact Us"
As mentioned in the reference manual: 1 Press the Menu key. 2 Navigate to Methods > Method settings > Calculation > Result limits. 3 To deactivate Result limits, navigate to No. 4 To activate Result limits, navigate to Yes. 5 Press the OK key. The Lower limit and Upper limit are displayed. 6 Set the Lower limit and Upper limit. For more information, please contact your local representative: www.mt.com/us/en/home/site_content/contact_us.html "
Hi, If you use the built-in sample pump, viscosities up to 100 mPa*s are possible. If you use a syringe to load the instrument, viscosities up to 1000 mPa*s can be measured. For higher viscosities, we recommend the Easy D40.
is it good to use in industrial batteries density & temp ?
Hi,
Yes, the Densito can be used in battery production, for example, to measure the density of electrolytes. The instrument also allows temperature compensation of the measured values to your temperature of interest.
Is it safely compatible with petrolium and disel industries
Thank you for your message, please find information on compatibility under www.mt.com/DERE-norms or ask a local representative. Contact details can be found at www.mt.com under "Contact Us"
How u access quality control menu ??
As mentioned in the reference manual:
1 Press the Menu key.
2 Navigate to Methods > Method settings > Calculation > Result limits.
3 To deactivate Result limits, navigate to No.
4 To activate Result limits, navigate to Yes.
5 Press the OK key. The Lower limit and Upper limit are displayed.
6 Set the Lower limit and Upper limit. For more information, please contact your local representative: www.mt.com/us/en/home/site_content/contact_us.html "
Can we check viscous/ treacly liquid?
The sample pump works for samples with viscosities up to 100 mPa*s. Measurement of samples with a syringe are possible up to 1000 mPa*s
Is it appropriate for high viscosity?
Hi,
If you use the built-in sample pump, viscosities up to 100 mPa*s are possible. If you use a syringe to load the instrument, viscosities up to 1000 mPa*s can be measured. For higher viscosities, we recommend the Easy D40.