Equilibration Time in Autoclave

Поделиться
HTML-код
  • Опубликовано: 11 дек 2024

Комментарии • 20

  • @AshishMishra-om6mk
    @AshishMishra-om6mk 4 месяца назад +1

    jai ho

  • @sumitkumarsharma3914
    @sumitkumarsharma3914 7 месяцев назад +1

    Very informative vedio
    Thank you so much sir

  • @nirmalsartana4061
    @nirmalsartana4061 7 месяцев назад +1

    Nice information sir

  • @drx.trimbak3126
    @drx.trimbak3126 6 месяцев назад +1

    Sir please make video on dwell time

  • @manojpatel6518
    @manojpatel6518 7 месяцев назад +1

    Video on Manufacturing and storage tank validation .

  • @saivijay1580
    @saivijay1580 7 месяцев назад +1

    Sir thank you very much for your valuable knowledge sharing. Always a pleasure to have your video. Request you to please make a video on "if an injection product is failed to comply with in-process specification which is pertaining to bulk sample, in tests such as pH and Assay and if Finished product specification of same batch is well within acceptance criteria, why can't we release batch without solid identification of root cause for bulk failure as ultimately finished product is sent for administration to pateint.

    • @PHARMAVEN
      @PHARMAVEN  7 месяцев назад

      You are right but in practical sense, the batch was failing in process, if there is no laboratory error or if we cannot have another sample to test once lab error is confirmed, then this batch cannot be released

  • @pravinsingh8170
    @pravinsingh8170 23 дня назад +2

    Autoclave me hot point and cold point kya hota hai

    • @PHARMAVEN
      @PHARMAVEN  21 день назад

      Hot spot and cold spot identified based on temperature penetration and distribution studies by placing thermocouples

  • @subratkumardash6465
    @subratkumardash6465 3 месяца назад +1

    Sir can you make a video about the drying process and why it important

    • @PHARMAVEN
      @PHARMAVEN  3 месяца назад

      Drying of load in autoclave?

  • @saivijay1580
    @saivijay1580 7 месяцев назад +1

    Hi Sir!! Please make a detailed video on difference between collapsed cake vs improper cake vs melt back defects which are blatant defects identified in Lyophilised visual inspection. I would be a great help if expected explanation from your side is along with pictorials.

    • @PHARMAVEN
      @PHARMAVEN  7 месяцев назад

      Ok I will make

  • @NareshChauhan-t7o
    @NareshChauhan-t7o 5 дней назад

    Sir autoclave sencer or thermocouple sencer ke temperature me kitna difference hona chahiye

  • @nikhilverma1560
    @nikhilverma1560 4 месяца назад

    Sir please tell me the temperature distribution acceptance criteria of probe to probe and within probe

  • @prasad131286
    @prasad131286 3 месяца назад

    Lag time difference between previous validation cycle and present cycle . Value is higher than the previous validation . What is the answer

  • @manojpanchal2737
    @manojpanchal2737 6 месяцев назад +1

    Sir,
    Chamber temperature or drain temperature

    • @PHARMAVEN
      @PHARMAVEN  6 месяцев назад

      For Equilibration time, it is chamber temperature

  • @ravindrajoshi3374
    @ravindrajoshi3374 7 месяцев назад +1

    Then sir, what is the difference between 'lag time' and 'equilibrium time'.
    Does equilibrium time related to empty load only,
    Does it mean difference between reference temperature (i.e.inbuilt reference temperature sensor of autoclave) and data logger sensor
    Plz correct me if I am wrong, plz guide me.

    • @PHARMAVEN
      @PHARMAVEN  7 месяцев назад +1

      Equilibration time is for loaded chamber
      Book on validation by Agalloco has given below explanation about lag time with reference to Dry heat sterilization
      The heating lag is defined
      as the difference between the time required for the product to reach the minimum required temperature and the time
      required for the sterilizer air to reach the minimum required temperature. A heating lag will be magnified during the maximum load of product (i.e., maximum density or mass).