Video: How To - Setting up and Calibrating the HI98115 GroLine pH Tester

Поделиться
HTML-код
  • Опубликовано: 3 дек 2024

Комментарии •

  • @phongvo8300
    @phongvo8300 3 года назад

    Very good and funny videos bring a great sense of entertainment!

  • @LookHowSmartYouAreGetting
    @LookHowSmartYouAreGetting 3 года назад

    REVISED INSTRUCTION: starting at 00:45 in the video, TO CALIBRATE: #1 While cap is on, press black button *ONCE & RELEASE* to turn on unit. (Holding down button from OFF state for 6 secs does not Produce the CAL option).. Once on, THEN #2 press & hold down button (approx 6 secs) only until you see "CAL" then quickly release to remain in that mode. You should then be able to proceed with the other instruction to calibrate and take cap off before dipping.. (Be careful with glass tip & always replace tip with a couple drops of 004 (or storage solution also avail for purchase) in cap to keep from drying out). Note, the "Sto" took a little bit to appear, so don't get rushed into thinking that step failed.. "wait for iiiit" lol if all else fails call the Tech support guy, he's awesome and will resolve any issues.. Hopefully they'll update their insert & video to be much more literal with each step (especially STEP 1).. CUZ IF YOU DON'T START OFF ON THE RIGHT FOOT YOU'LL NEVER GET THERE

  • @darthmalak190
    @darthmalak190 3 года назад

    0:45 for setup -- 1:18 for calibration

  • @ernestolopez2044
    @ernestolopez2044 3 года назад

    Can i use the quick calibration solution with this product?

  • @jmannUSMC
    @jmannUSMC 5 лет назад

    Can't seem to find this products manual anywhere. Searching the manuals section on the mobile website just returns me back to the product page, where I don't see anything. Really wishing I hadn't lost the paper copy 😕

    • @hanna_hydro2393
      @hanna_hydro2393  5 лет назад

      Hi Joseph. Contact sales@hannainst.com and they can send you a manual.

  • @mosthated8190
    @mosthated8190 5 лет назад +1

    When doing a 2 point calibration , it says "ERR" when reading the second calibration of 4.01 ?

    • @LookHowSmartYouAreGetting
      @LookHowSmartYouAreGetting 3 года назад +1

      I was to Err is to be expected and simply proceed.. don't be alarmed..

    • @dylanbosinger3970
      @dylanbosinger3970 2 года назад

      Is that correct mine was reading error as well too like he said I kept it in there anyways and proceeded to the next session of 4.01 then to the STO symbol

  • @ashiqroopchand7826
    @ashiqroopchand7826 5 лет назад

    Why does my hanna not give a stable reading? ? ?

    • @hanna_hydro2393
      @hanna_hydro2393  5 лет назад +1

      Hi Ashiq. Which Hanna product do you use? Typically the main reason you won't get a stable reading is the electrode is not properly hydrated. It's important that you rinse with DI Water after use, use a cleaning solution, and store it in storage solution.

  • @battlestargalactica3593
    @battlestargalactica3593 5 лет назад

    Wish it came with the storage chemical.

    • @ashtondavis790
      @ashtondavis790 5 лет назад

      use the ph 4

    • @TheLoukalTv
      @TheLoukalTv 4 года назад

      @@ashtondavis790 what about the 7 I use both

    • @ashtondavis790
      @ashtondavis790 4 года назад

      @@TheLoukalTv you want to use ph 4 so calibrations dont get thrown off

    • @LookHowSmartYouAreGetting
      @LookHowSmartYouAreGetting 3 года назад

      Customer Support sd use a couple drops of #4 in your black cap if you haven't purchased 'storage solution'.