Swan - Swansensor Oxytrace G - Air Calibration

Поделиться
HTML-код
  • Опубликовано: 14 окт 2024
  • For technical questions, our support team will gladly help you at
    swan.ch/en/Def...
    Calibrating a Swansensor is that easy! Anyone can do it!
    Products:
    • AMI Oxytrace A-22.401.000: www.swan.ch/Cat...
    • AMI Oxytrace QED A-22.451.000: www.swan.ch/Cat...
    Safety Data Sheets for Filling solution 1ALK: www.swan.ch/Get...
    AMI Oxytrace & AMI Oxytrace QED Manual: www.swan.ch/Get...
    For technical question, our support team will gladly help you out at: support@swan.ch www.swan.ch/en/...
    Swan Analytische Instrumente AG
    Studbachstrasse 13
    CH-8340 Hinwil
    www.swan.ch
    LinkedIn: / swananalyticalinstruments

Комментарии • 3

  • @corey8496
    @corey8496 2 года назад +1

    Ok so if the saturation level doesn't ever get above 23% what could be the issue?
    Alky solution was changed and ran at air for 1 hour then tried to do an air calibration but would not take a cal because of the saturation was not high enough.
    Is saturation referring to the akly solution of the anode and cathode?

  • @ardanrizqul6329
    @ardanrizqul6329 Год назад

    How to fix error 031

    • @SwanAnalyticalInstruments
      @SwanAnalyticalInstruments  Год назад

      Hi Ardan,
      This E031 is caused by the electronic failure that handle the signal output on the mainboard. Usually a reboot (power off then power on) of the instrument helps to solve this error. If that still doesn’t help to solve this error, then the customer will need to exchange the mainboard.
      For further problem/ details, please ask the customer to write an email to support@swan.ch then we can keep track of the problem with the case number.