Weighing Balance Calibration (

Поделиться
HTML-код
  • Опубликовано: 28 окт 2024

Комментарии • 15

  • @ajitjadhav607
    @ajitjadhav607 11 месяцев назад +3

    Hello sir, hope you are doing well.
    Let us know, how we can find out the smallest net weight & is there any formula given in USP general chapter?

    • @PharmaScholars
      @PharmaScholars  11 месяцев назад +1

      Yes dear, there is a formula , I have mentioned in this video. Check it out @ 7:38

    • @ajitjadhav607
      @ajitjadhav607 11 месяцев назад

      @@PharmaScholars Hello madam, hope you are doing well.
      Let me know that in repeatability there is a statement that if SD is less than 0.41d, then we should use 0.41 x least count of balance for calculations. So in which unit we have to take value of d. In g or mg ?

    • @PharmaScholars
      @PharmaScholars  11 месяцев назад +1

      It depends upon the least count of respected weighing balance. Your least count is the minimum value display on your weighing balance. Suppose it is 0.1mg, then unit will be mg, & if it is 0.1g then unit will be g.

    • @ajitjadhav607
      @ajitjadhav607 11 месяцев назад

      Thnk u for your reply madam, but we convert both mg to g and g to mg. My question for calculation which unit we should use ?

    • @PharmaScholars
      @PharmaScholars  11 месяцев назад

      U can use any unit as per your requirement, but calculation will be easy if you use same unit as mentioned as d value.

  • @Cmpunk022
    @Cmpunk022 3 месяца назад +2

    Thank you❤

  • @rajendrakumawat3601
    @rajendrakumawat3601 2 месяца назад

    if u wiil explain with practically in excel its much better for understand

  • @ganeshkilli8265
    @ganeshkilli8265 7 месяцев назад +1

    Sir,
    Why 0.41 using and where it omes from

    • @PharmaScholars
      @PharmaScholars  7 месяцев назад +1

      This is a standard factor given by USP.