Rigel 288+ overview & testing to IEC 60601 & IEC 62353

Поделиться
HTML-код
  • Опубликовано: 25 июл 2024
  • In this short video, Michael Walton, Category Manager at Rigel Medical covers the main features of the Rigel 288+ Electrical Safety Analyzer and shows how it can test to both IEC 60601 and IEC 62353.
    If you would like to find out more about the Rigel 288+ Electrical Safety Analyzer, please visit www.rigelmedical.com/288+ or email sales@rigelmedical.com
  • НаукаНаука

Комментарии • 13

  • @ParminderSingh-re4xp
    @ParminderSingh-re4xp 6 месяцев назад

    Thanks, great presentation

  • @FranherKrusher
    @FranherKrusher 5 лет назад

    Good presentation! The analyzer is very friendly. Thank you!

  • @lucianovidus7854
    @lucianovidus7854 4 года назад

    Ottimo video, esposizione semplice e lineare, ottimo strumento.

    • @SeawardElectronic
      @SeawardElectronic  4 года назад

      Ciao Luciano, grazie. Puoi trovare altri contenuti educativi sul nostro sito Web all'indirizzo www.rigelmedical.com/webinars/

  • @czarekduparek6522
    @czarekduparek6522 5 лет назад +1

    Damn man, you rule!

  • @sintorajan9520
    @sintorajan9520 3 года назад

    Thank you 🙏

  • @technoeducation5314
    @technoeducation5314 2 года назад

    Currently we are using 2x Rigal electrical Safety

  • @robmartin8072
    @robmartin8072 4 года назад +1

    Merci.

  • @technoeducation5314
    @technoeducation5314 2 года назад

    Nice 👍

  • @amalgovind3461
    @amalgovind3461 4 года назад +1

    How did you performed the enclosure leakage test without connecting to any exposed conductive part?

    • @SeawardElectronic
      @SeawardElectronic  4 года назад +2

      Hi Amal, Michael has provided the following feedback directly from the IEC 60601-1 guidance:
      "If ME EQUIPMENT has an ENCLOSURE or a part of the ENCLOSURE made of insulating material, metal foil of maximum 20 cm x 10 cm is applied in intimate contact with the ENCLOSURE or relevant part of the ENCLOSURE.
      "The metal foil is shifted, if possible, to determine the highest value of the TOUCH CURRENT.
      The metal foil should not touch any metal parts of the ENCLOSURE that are possibly PROTECTIVELY EARTHED; however, metal parts of the ENCLOSURE that are not PROTECTIVELY EARTHED can be covered partly or totally by the metal foil.
      "Where it is intended to measure the TOUCH CURRENT in the SINGLE FAULT CONDITION of
      interruption of a PROTECTIVE EARTH CONDUCTOR, the metal foil is arranged to contact parts of the ENCLOSURE that are normally PROTECTIVELY EARTHED.
      "Where the surface of the ENCLOSURE contacted by the PATIENT or OPERATOR is larger than
      20 cm x 10 cm, the size of the foil is increased corresponding to the area of contact.
      "
      We hope this helps.
      Thank you.

    • @gustavofoss2280
      @gustavofoss2280 10 месяцев назад

      @@SeawardElectronic im studying the iec 62353 and 60601, where is the metal foil located in the presentation?