Pipette Calibration Demonstration complete

Поделиться
HTML-код
  • Опубликовано: 15 ноя 2024

Комментарии • 9

  • @KasimierLP
    @KasimierLP 7 месяцев назад +1

    "Thinks about it's life, decides if it wants to work" That's literally every device in a lab, especially whenever you need that device specifically. xD

  • @vinicius_stelet
    @vinicius_stelet 2 года назад

    Never though it was so easy to check calibration from our pipettes using only water and a balance. Thanks for sharing.

  • @raviuniqueVlogs
    @raviuniqueVlogs 2 месяца назад

    Are you changing the boats after?

  • @wahajaijaz7690
    @wahajaijaz7690 Год назад +2

    what's the acceptable range of deviation in ml of pipette and grams shown in balance.

    • @lacyrmoss-phdmls
      @lacyrmoss-phdmls  8 месяцев назад

      It depends on the manufacturer and the volume of the pipette. Each pipette will come with an insert that states the acceptable volume for the pipette at the high end and low end of the pipette. You can then convert that volume to grams based on the density (1.000 g/mL) and temperature of the water you are using.

  • @beautynaturefaza8052
    @beautynaturefaza8052 4 месяца назад

    Hi Lacy, may i know how do you calculate the density of water during testing? Any simple calculation? Kindly help.