Powerful yet Efficient: Mindray BS-600M Chemistry Analyzer Global Launch

Поделиться
HTML-код
  • Опубликовано: 11 май 2022
  • Mindray has launched the BS-600M, a powerful yet efficient chemistry analyzer that is designed to empower medium-volume laboratories with increased productivity, efficiency and reliability, while cutting down costs. This breakthrough analyzer provides the whole package that helps your lab to grow.
    Learn more about BS-600M at:www.mindray.com/en/products/l...
  • НаукаНаука

Комментарии • 9

  • @AyazKhan-ns8mo
    @AyazKhan-ns8mo 4 месяца назад

    I also want to be a part of Mindray team

    • @AyazKhan-ns8mo
      @AyazKhan-ns8mo 4 месяца назад

      I also want to be part of Mindray team from Pakistan

  • @Luis-bo2uj
    @Luis-bo2uj Год назад +1

    got a brandnew BS-360E a month ago. Having si many problems with it...

    • @ThanhTran-dp6yh
      @ThanhTran-dp6yh 2 месяца назад

      i am using a BS-240E. what problems do you have with your analyser?

    • @Luis-bo2uj
      @Luis-bo2uj 2 месяца назад +2

      @@ThanhTran-dp6yh First I had a problem in which I had very scattered absorbance readings. The problem was that one of the power cables for the lamp was loose. Now I have problems with the probe, it detects clots and fibrin in samples that clearly do not have them. It seems that the probe sensor is failing.

    • @ThanhTran-dp6yh
      @ThanhTran-dp6yh 2 месяца назад

      ​@@Luis-bo2uj "scattered absorbance readings" and "detects clots", i had these very issues too, but it happened bc i didn't use CD80 as concentrated wash solution in potion 99 (before that i used water, i thought it would saved me some reagent). the probe then clogged
      i fixed it by using Javel to unclog the probe and fill CD80. it works fine since then.
      you should check your centrifuge too. bad centrifuge could lead to probe clogged. see if your serum/plasma samples have small blood particals within

    • @ThanhTran-dp6yh
      @ThanhTran-dp6yh 2 месяца назад

      "scattered absorbance readings" and "detects clots", i had these very issues too, but it happened bc i didn't use CD80 as concentrated wash solution in potion 99 (before that i used water, i thought it would saved me some reagent). the probe then clogged
      i fixed it by using Javel to unclog the probe and fill wash bottle with CD80. it works fine since then.
      you should check your centrifuge too. bad centrifuge could lead to probe clogged. see if your serum/plasma samples still contain small blood particals within

    • @ThanhTran-dp6yh
      @ThanhTran-dp6yh 2 месяца назад

      the problem was, the probe was not ENTIRELY clogged. it was HALF-CLOGGED. i didn't think of it until a local technician helped me