Repair mindray BC-3200 Baground abnormal and WBC clog

Поделиться
HTML-код
  • Опубликовано: 9 ноя 2024

Комментарии • 37

  • @badis_dz5808
    @badis_dz5808 Год назад +2

    Thanks bro 🙆

  • @aamery7198
    @aamery7198 Год назад +1

    Great video brother 👍

  • @joelbaguma8551
    @joelbaguma8551 Год назад +2

    Great video sir thank you for sharing. Am happy about the English subtitles 👍👍 because I can understand the procedures

  • @TechnoMedBME
    @TechnoMedBME Месяц назад +1

    Thank you

  • @jlmonteblack
    @jlmonteblack Год назад +1

    Great vídeo, cómo puedo contactarlo?

  • @zaheermaniyar8806
    @zaheermaniyar8806 10 месяцев назад

    Supper 🎉🎉🎉🎉

  • @irzaharun4683
    @irzaharun4683 6 месяцев назад +1

    Izin bertanya pak, kalau hasil running sampel pasien WBC, RBC, dan PLT smuanya **** itu diagnosanya apakah terdapat sumbatan di jalur probe nya ya pak? 🙏

    • @d.i.yrepairnmaintenance9538
      @d.i.yrepairnmaintenance9538  6 месяцев назад

      @IrzaHarun4683 jika disertai eror clog / re-count di akhir pembacaan sample..berarti terjadi sumbatan bisa di sebabkan oleh darah sample nya (kurang homogen)..coba masuk maintenance..zap aperture 3x.setelah itu flush aperture 3x..coba baca lagi sample nya

  • @andisomeng1208
    @andisomeng1208 5 месяцев назад +1

    Mas mau tnya klau gak kluar gelembung dan sensor merahnya cuma berkedip cepat apanya y mas..mksih

    • @d.i.yrepairnmaintenance9538
      @d.i.yrepairnmaintenance9538  5 месяцев назад

      Bisa karena ada valve yang agak terbuka atau kurang menutup nya...jadi ada gelembung Udara .... bisa juga bocor di sambungan² selang di chamber nya..bisa juga sumbatan di ruby aperture nya..itu yang paling sering terjadi

  • @MohamedHassan-pw6xc
    @MohamedHassan-pw6xc 5 месяцев назад +1

    i have wbcs good but no diff error can u help me plz
    and if there a way i contact u
    thanks in advance

    • @d.i.yrepairnmaintenance9538
      @d.i.yrepairnmaintenance9538  5 месяцев назад

      Are all the samples... the diffcount results are not there? try cleaning the ruby ​​stone...and check the exhaust valve from the WBC Chamber

  • @Muslim.25ad
    @Muslim.25ad 10 месяцев назад +1

    Background abnormal with wbc&rbc clog in mindrey 3000 plus.how we remove the error

  • @muzaffarathar1841
    @muzaffarathar1841 10 месяцев назад

    Same problem my machine wbc clog back ground abnormal ***********

    • @d.i.yrepairnmaintenance9538
      @d.i.yrepairnmaintenance9538  10 месяцев назад

      Try ..zap aperture 3x..after that..flush aperture 3x..and then..pres aspirate without material sample

  • @akhmadfirzon383
    @akhmadfirzon383 Год назад

    Klo background abnormal terus di plt gimana bang ?

  • @DermawanFaaghnini-mc7pv
    @DermawanFaaghnini-mc7pv Год назад

    Hasil wbc 0 trus knapa ya mas?