Ultrasonic testing calibration-part 3

Поделиться
HTML-код
  • Опубликовано: 25 ноя 2024

Комментарии • 9

  • @dudeabc100
    @dudeabc100 8 лет назад +1

    Dude as per AWS D1.1 u need to set the height of the reference reflector between 40% to 60% screen height for an angle beam testing and that for a Zero degree inspection for the lamination chech should be between 50% to 70% screen height..Bcz based on the procedure you will be adding a very high gains above the reference gain and hence for an angle beam testing with AWS procedure its better to peak the reflector signal between 40% to 60% screen height in order to avoid signal saturation due to a very high scanning gain

    • @subashvelu660
      @subashvelu660 4 года назад

      Not exact some clients ask to set 80 % SH for angle beam inspection then should be add minimum 14 dB of the yours reference dB level and it's dependence up one the matal path value..

  • @medjallouli9244
    @medjallouli9244 7 лет назад

    please the position of hole in the calibration v1??? , merci

  • @meriemtazi7957
    @meriemtazi7957 6 лет назад

    Hello, I have a few questions regarding this device. Would you be able to assist? Thanks

  • @deanbunn753
    @deanbunn753 8 лет назад

    Do you do this for a living?

  • @deanbunn753
    @deanbunn753 8 лет назад

    2.25 transducer.....