Wonderful presentation 👏 Sir I want to know ANOVA (analysis of variance) how to calculate this and how to interpretation of this statistical data, Sometimes ANVISA asking for this also And put some light on t-95 upper and t 95 lower. I have some confusion regarding this. Thank you.
As per FDA regulations if intercept more than 2 then linearity shall not be accepted. Please make clarity, if is there situation like this How to resolve
You may narrow down the range, in case of not meeting acceptance criteria. High background noise (for instrument methods) and residuals may lead to high bias.
Hello Jay. Please find a video on the calculation of the residual sum of squares for your kind information. Link: ruclips.net/video/vhVBolLIC5Y/видео.html
@@pharmagrowthhub3083 Thank you for reply. I am talking about ANVISA calculation parameter such as homoscedasticity, angular coefficient, and 5% significance level in full calculation.
According to ANVISA, "The solutions used for the evaluation of linearity must be prepared independently and diluted solutions from the SAME MOTHER SOLUTION of the CRS can be used." Where they says three different stock solutions must be prepare?
Sir One class is needed for excel for this calculationa.In maximum cases regulatory body excels are locked.So calculation formula for excel highly needed
Most of the people are not aware about ANVISA guideline about Linearity.Thank you for teaching us beautifully.
Glad you think so, Rahul! thank you!
What about recovery 3 preparation or single
3 preparation
Wonderful presentation 👏
Sir I want to know ANOVA (analysis of variance) how to calculate this and how to interpretation of this statistical data, Sometimes ANVISA asking for this also
And put some light on t-95 upper and t 95 lower. I have some confusion regarding this.
Thank you.
Any limit for residuals and bias at 100% and homoscedticity
As per FDA regulations if intercept more than 2 then linearity shall not be accepted.
Please make clarity, if is there situation like this
How to resolve
You may narrow down the range, in case of not meeting acceptance criteria. High background noise (for instrument methods) and residuals may lead to high bias.
Can you say the guidelines reference no for intercept nmt 2
Please prepare a video on all calculation parameters like the residual sum of squares.
Hello Jay. Please find a video on the calculation of the residual sum of squares for your kind information. Link: ruclips.net/video/vhVBolLIC5Y/видео.html
@@pharmagrowthhub3083 Thank you for reply. I am talking about ANVISA calculation parameter such as homoscedasticity, angular coefficient, and 5% significance level in full calculation.
According to ANVISA, "The solutions used for the evaluation of linearity must be prepared independently and diluted solutions from the SAME MOTHER SOLUTION of the CRS can be used."
Where they says three different stock solutions must be prepare?
Sir One class is needed for excel for this calculationa.In maximum cases regulatory body excels are locked.So calculation formula for excel highly needed
Sensitivity should be high any criteria
Hi Sir please tell me limit of %Y intercept why is +-10%
Make video for 5 % significance level