  • Digital LED Gas Alert Alarm Specification MAP Brand for Hospital
    1. General Requirements:
    Purpose: To provide continuous digital monitoring and alerting of medical gas levels and system status in hospital environments.
    Location: To be installed in critical areas such as ICU, surgical suites, and recovery rooms.
    2. Dimensions and Design:
    Dimensions:
    Height: 275mm
    Width: 415mm
    Depth: 80mm
    Design: Wall-mounted with a digital LED display.
    Material: High-impact, flame-retardant MS Housing.
    Finish: Smooth finish with antimicrobial Powder coating.
    3. Gas Monitoring Capabilities:
    Types of Gases Monitored:
    Oxygen (O2)
    Nitrous Oxide (N2O)
    Medical Air
    Vacuum (suction)
    Carbon Dioxide (CO2)
    Sensors: High-accuracy, long-life digital sensors specific to each gas type.
    4. Alarm Specifications:
    Visual Alarms:
    Color-coded LED indicators for each gas (Green for normal, Yellow for warning, Red for alarm).
    Audible Alarms:
    Adjustable alarm volume with a minimum of 85 dB at 1 meter.
    Distinct tones for different alarm conditions (low gas pressure, high gas pressure, system fault).
    Alarm Conditions:
    High/low gas pressure.
    Sensor failure.
    Power failure.
    5. Display:
    Type: High-resolution LED display.
    Information Displayed:
    Real-time gas pressure levels.
    Alarm status for each gas.
    System status and fault indications.
    User Interface:
    Intuitive interface with easy navigation.
    Test & Mute display settings.
    6. Electrical Specifications:
    Power Supply:
    Input Voltage: 110-240V AC, 50/60Hz.
    Power Consumption: ≤15W
    Safety Standards: Compliant with IEC 60601-1 for medical electrical equipment.
    7. Environmental Specifications:
    Operating Temperature: 0°C to 50°C
    Humidity: 0% to 95% RH, non-condensing
    Ingress Protection: IP54
    8. Installation and Mounting:
    Mounting: Wall-mounted with included hardware.
    Connections: Quick-connect fittings for gas lines and secure electrical connections.
    Installation: To be installed by certified technicians.
    9. Compliance and Safety:
    Certifications:
    CE marked.
    Compliance with ISO 13485:2016. (Medical Gas Pipeline Systems).
    Compliance with NFPA 99 (Health Care Facilities Code).
    Safety Features:
    Overpressure protection.
    Short-circuit protection.
    Surge protection.
    10. Maintenance and Calibration:
    Maintenance:
    User-friendly access for maintenance.
    Regular maintenance checks recommended every 6 months.
    Calibration:
    Factory calibrated with a recommended recalibration interval of 12 months.
    On-site calibration capability with provided software and tools.
    12. Warranty:
    Standard Warranty: 1 years for all components.
    13. Documentation:
    Compliance Certificates: Provided for all relevant safety and performance standards.