Apera Instruments PH20 Pocket pH Tester Quick Start Tutorial

Поделиться
HTML-код
  • Опубликовано: 28 июл 2024
  • The all-new tutorial for the Value Series PH20 Pocket Tester Kit
    00:06 What's in the kit
    00:12 pH calibration
    02:53 pH measurement
    03:33 Parameter settings
    04:21 Probe cleaning
    05:43 Probe storage
    06:45 Battery replacement
  • НаукаНаука

Комментарии • 57

  • @PlantManCheeseHead
    @PlantManCheeseHead 6 месяцев назад +2

    I appreciate the video tutorial. Im all set up!

  • @coachmike3266
    @coachmike3266 Год назад

    Need a video like this for the Apera 910 series please!

  • @sugargraham6868
    @sugargraham6868 6 месяцев назад

    Can you please tell me how to restart mine???😢

  • @diabetekAFA
    @diabetekAFA Год назад +1

    Hello, I just got the PH20 PH tester, but I read in the instruction manuel that: if you find any air bubble in the glass bulb of the PH sensor, simply shake the probe for a few times to remove it. But isn't it normal that their is an air buble? Because I tried and get rid of the air buble, but with no succes. Thanks.

    • @aperainst
      @aperainst  10 месяцев назад

      Try shaking the meter off like whipping.

  • @jimmyhartman573
    @jimmyhartman573 Год назад +4

    i dip it into PH7 fluid and it reads 7.6 and when i short click it it reads er1 and i cannot do anything with it

    • @hmmy92
      @hmmy92 Год назад

      do you find solution?

    • @TheGhostrider999
      @TheGhostrider999 3 месяца назад

      Same

    • @gerardsitja
      @gerardsitja Месяц назад +1

      Welcome to the club. This prove seems as rubbish as the 20$ ones... They don't last more than a year even looking very well after it.

  • @markoxendine9748
    @markoxendine9748 10 месяцев назад +1

    When you say pure water are you talking about distilled or RO water or just tap water ????

    • @aperainst
      @aperainst  10 месяцев назад

      Preferably distilled water. RO water is also fine. Tap water is not recommended.

  • @drewhankey4078
    @drewhankey4078 Год назад +2

    How good is this one need to buy one

    • @cmmnsenserules4339
      @cmmnsenserules4339 11 месяцев назад

      I used to own Blue Labs ($95), I broke it, bought this, $50, THESE hold calibration better. I test daily. Every 2 weeks calibrate her. But again, I use daily 5 different tests.

  • @Goblinterps
    @Goblinterps 4 месяца назад

    When I go to do the 2nd in 4.0. It's just keeps flashing between 4 and 10 and the reading is stuck on 7

  • @diabetekAFA
    @diabetekAFA 7 месяцев назад

    Hello, I have the Storage solution 3M-KCl and I have filled the cap of my PH tester with the solution to the fill line. But do you need to put the probe of the PH meter always in the storage solution or just overnight? Thanks

    • @aperainst
      @aperainst  6 месяцев назад

      Always storing the ph probe when it's not in use is the best practice.

    • @diabetekAFA
      @diabetekAFA 6 месяцев назад

      @@aperainst thanks for the info, I have stored the ph probe in the Storage solution 3M-KCl when I'm not using the PH tester.

    • @jameswilliford5478
      @jameswilliford5478 3 месяца назад

      I wouldn't do that as above and in their tutorial they say only store in the KCL solution for long term storage and not if you use it daily or weekly. They also say store it in the ph 4 buffer solution or with just a few drops of pure (distilled water) in the cap to maintain humidity. Now I am confused too as I've heard not to store in the buffer solution and not to store in distilled water. But they're saying the KCL is just for long term storage or restoring a dried out probe.

    • @diabetekAFA
      @diabetekAFA 3 месяца назад +1

      @@jameswilliford5478 Damn all those different stuff. I just store it in the KCL solution. I guess if it's good for long term storage or restoring a dried out probe, it's good for short term storage too. Better storing it in KCL solution then nothing att all. So how do you store your PH tester?

  • @shadykitty7493
    @shadykitty7493 Год назад

    what s the dif between the blue and green 1?

    • @aperainst
      @aperainst  21 день назад

      Blue screen is for measurement mode. Green screen is for calibration mode.

  • @michelm.6704
    @michelm.6704 9 месяцев назад

    Hello. Can I rinse the probe of this Ph meter in demineralized water instead of distilled water ? Thanks ;)

    • @aperainst
      @aperainst  9 месяцев назад +1

      Yes demineralized or deionized water is perfect.

    • @michelm.6704
      @michelm.6704 9 месяцев назад

      @@aperainst thank you ;)

  • @thombaz
    @thombaz 9 месяцев назад

    Can I or should I store it in kcl solution long term? I don't really understand this as it seeems like all the ph meter use the same techology. Still some say store it in kcl solution all the time, some say it need to be humid, or in water and some do not say anything at all. It's a pricey stuff and I want to have it the longest lifespan as it can.

    • @aperainst
      @aperainst  9 месяцев назад +2

      For long term storage (>1 month) or restoring a dried-out pH probe, we'd recommend storing in 3M KCl soaking solution. But if you use the meter frequently (like daily or weekly), keep the storage cap wet with pure water or soak in pH 4 buffer solution will do the job.

    • @thombaz
      @thombaz 9 месяцев назад

      @@aperainst Thank you!

    • @sendoholic420
      @sendoholic420 2 месяца назад +1

      @@aperainstok thank you, but what do you mean with „pure water“

    • @aperainst
      @aperainst  2 месяца назад +1

      @@sendoholic420 pure water refers to distilled water/RO water/deionized water

  • @saint85
    @saint85 8 месяцев назад

    I have this pen can I replace the prob my pen is not working

    • @aperainst
      @aperainst  8 месяцев назад +1

      You can send the unit back to our facility at 6656 Busch Blvd, Columbus, Ohio 43229. One of our technicans can help you replace the probe for PH20.

  • @davidl7134
    @davidl7134 10 месяцев назад

    How to get into tds mode i am stuck on ph mode

    • @NoneYaBiz5
      @NoneYaBiz5 9 месяцев назад

      The PH20 ONLY reads pH.

  • @jpthsd
    @jpthsd 7 месяцев назад

    Do I need to calibrate before any pH test?

    • @aperainst
      @aperainst  6 месяцев назад

      No. Typically one calibration can hold the accuracy for about a month. But you can always check the accuracy by testing the fresh buffer, if there is discrepancy, you can calibrate the meter to get it right.

  • @gerardsitja
    @gerardsitja Месяц назад

    After a year of use looking really well after it I can no longer calibrate, as the reading of the 7 buffer reads as Ph 8 and wont calibrate, it throw the error er1.. Any solutions or I need to get a new prove/device!? If so I am very disappointed with the product! The only reason I choose you it was cos I though yours will be better, but it did not last more than any regular 20$.. Not worth the extra 40$ by any means.

    • @aperainst
      @aperainst  Месяц назад

      We are sorry to hear this happening to you. It could be due to contaminated buffers, aged probe, or malfunctioning instrument...we need to test on the product to identify the root cause. You can go to support.aperainst.com to submit a RMA ticket. One of our technical support will be in touch to help you resolve the problem.

  • @FatherNatureCNY
    @FatherNatureCNY 25 дней назад

    So, if water isn't for storage, how are the water drops that maybe in there when it's brand new any good for it??? I've always kept drips in my ph pen caps to keep diode from drying up. Wtf do I putting it then? Drops of ph 4 buffer???? Cuz it sure didn't come with storage solution and idk why tf ot wouldn't

    • @aperainst
      @aperainst  25 дней назад

      Hello! We are sorry about the confusion there. Leaving a few water droplets is only for maintaining a humid atmosphere in the probe cap to prevent the pH membrane from drying out. It’s different from soaking/immersing the pH probe in water, which is going to damage the probe gradually. Using 3M KCl soaking solution to store the pH probe is the best practice that can keep the pH sensor at its optimal condition. As people might not always have 3M KCl on hand, the alternative is soaking the probe in pH 4 buffer. When you don’t have pH 4 buffer on hand either, leaving a few water droplets in the cap is going to be much better than dry storage. To put it simply, there is a ranking for pH probe storage method, from worst to best: soaking in pure water like tap or drinking water (Very bad) --> dry storage (Bad) --> leaving a few water droplets in the probe cap (OK) --> soaking in pH 4 buffer (good) --> soaking in 3M KCl soaking solution (best). Hope you can find this info helpful.

    • @FatherNatureCNY
      @FatherNatureCNY 25 дней назад

      @aperainst definitely thanks! So some drops of well water is OK to keep moisture if I use the meter every day?

    • @aperainst
      @aperainst  24 дня назад

      @@FatherNatureCNY yes absolutely.

    • @FatherNatureCNY
      @FatherNatureCNY 23 дня назад

      @aperainst hey I have this new meter that I got when we were talking about storage...hlf the time I open the case,take it out and hit the button, and it doesn't turn on. That sucks, because what am I supposed to do the whole time this is in the mail going back to u guys AND the time it takes for a replacement to reach me? This sucks. And it won't turn on right now at all. I finally after 3½ yrs of using cheap junk throw away meters, went for it and bought an apera instrument. And here I am lol what am I supposed to do about this? Email?

    • @FatherNatureCNY
      @FatherNatureCNY 23 дня назад

      @aperainst even the day I got it, I calibrated it. There was a couple times it shut off before I could dip into the 7 solution, to water,and to the 4ph. Couple times I had to hit the power button a few times cuz it wouldn't turn on. Now it's still not turning on with 50 gallons of plant food mixed up for use this evening. It's an expensive mix every Friday believe me. This is about to go to waste I have a feeling🤬

  • @Gandalf0281
    @Gandalf0281 10 месяцев назад +1

    Hi guys, have a question. When Im trying to calibrate in the first 7ph solution, my ph meter show 6.9ph and 6 on the botton of the screen insted of 7. Seems like its not calibrating. Could you provide me any solution?

    • @aperainst
      @aperainst  10 месяцев назад +1

      You have changed the calibration standard series to NIST instead of USA by accident. It's an easy fix. Just go ahead and change it back to USA standard at settings P1. Please refer to 03:33 in the video on how to change the settings.😄

    • @Gandalf0281
      @Gandalf0281 10 месяцев назад

      @@aperainst Thans for a quick reply, fixed now :).

  • @krustyphish6249
    @krustyphish6249 Год назад

    Are the solutions bottles a 1 time use

    • @infidel66687
      @infidel66687 Год назад +2

      No, just rinse the probe off when going between the different bottles of calibration fluid to avoid cross contamination.

    • @jdawg219dawson8
      @jdawg219dawson8 Год назад +1

      No you can use many times

  • @NicaraguaPhoto
    @NicaraguaPhoto 9 месяцев назад

    Just received Apera PH20 pH Tester.....did calibration....7.00 worked fine but 4.00 came out as 3.9, not 4.0.

    • @sir6591
      @sir6591 7 месяцев назад

      So jusr know its .1 off

  • @dirtykobra5575
    @dirtykobra5575 2 месяца назад

    Are the solutions a one time use? Also can I use calibration solutions from other brands or does it have to be from apera only?I haven’t found any answers to this one yet. Thank!

    • @aperainst
      @aperainst  2 месяца назад

      pH 4 and 7 buffer solutions are good for up to 10 times of calibration as long as they are not expired or contaminated.

  • @davidl7134
    @davidl7134 10 месяцев назад +1

    I swear this vidoe was worthless how do you get from ph to tds mode?????

    • @aperainst
      @aperainst  10 месяцев назад +1

      The PH20 Tester only measures pH. It doesn't support tds measurement. Here are some pf our models that support pH and TDS dual measurement: PC60, PC60-Z, GroStar GS4.