Selectra Pro Clinical Chemistry Analyzer: Check R Check S [TGIF21]

Поделиться
HTML-код
  • Опубликовано: 22 авг 2024
  • The service procedure to check the accuracy of the Selectra fluidity system is called Check R Check S, a quality control of the Reagent (R) and Sample (S) pipetting. The procedure measures the pipetting precision of the instrument and indicates if pipettes or other parts of the liquid system need to be cleaned or replaced.
    The Check R Check S procedure is normally done by our service technician during the preventive or annual instrument maintenance and certification.
    Please contact our service team to discuss if you need a preventive maintenance, or if you want to verify the pipetting variation of the #Selectra instrument.
    #ClinicalChemistry #PreventiveMaintenance #ClinChemAnalyzer #IVDCalibration #IVD4U #MRKDiagnostics