Continuous Positive Airway Pressure CPAP IN Newborn: PHYSIOLOGICAL EFFECTS, TYPES & COMPONENTS- 1

Поделиться
HTML-код
  • Опубликовано: 15 окт 2024
  • nasal NCPAP # NICU# Bubble NCPAP # Flow drive CPAP # is important mode of noninvasive ventilation in NICU.
    This video discussing :
    What is the difference between CPAP and PEEP
    What is the physiological effects of CPAP? How is CPAP work?
    What are the 3 components of the CPAP
    What are the types of the CPAP

Комментарии • 12

  • @mohamedwafa
    @mohamedwafa 4 года назад +2

    Thank you very much
    Please continue for more videos

  • @preterm28
    @preterm28 4 года назад +1

    Great presentation , full of knowledge and experience

  • @abdelmotelbelshemy1833
    @abdelmotelbelshemy1833 4 года назад +1

    Thanks Dr. Mostafa

  • @khalidibrahim6134
    @khalidibrahim6134 4 года назад +1

    رائع يادكتور مصطفي 🌹

  • @osamamohamedy6385
    @osamamohamedy6385 4 года назад +1

    Highly recommended 👍

  • @emantaha5522
    @emantaha5522 4 года назад +1

    Thanks a lot Dr Mostafa, very interesting, waiting for more neonatology lectures.

  • @mohammedameen2020
    @mohammedameen2020 4 года назад +1

    جزاك الله خيرا دكتور مصطفي 🌹

  • @AAA-kt4ei
    @AAA-kt4ei 3 года назад +1

    بارك الله فيك دكتورنا ليه حضرتك مافيش فيديوهات تانى على التنفس الصناعي ؟ فيديوهات حضرتك جميله بس قليله

    • @nicu2222ify
      @nicu2222ify  3 года назад

      thank you for your kind comment, soon Inshallah

  • @adham45736
    @adham45736 4 года назад +1

    لو سمحت د مصطفي ال flowmeter
    حضرتك قولت من ٥ ل ١٠ ، بنحدده ع أساس ايه ؟ وهل بيأثر علي ال fio2 ؟ وبنبدأ من كام ؟

    • @nicu2222ify
      @nicu2222ify  4 года назад

      Thanks Dr. Adham,
      We adjust the Fio2 by the o2 blender,
      The flow-meter usually start by 8 liter/ minute. The flow is controlling the pressure in flow driver CPAP( Variable flow CPAP).
      However in bubble CPAP ( continuous flow) the pressure controlled by the depth of the end expiatory tube under water, and changing the flow minimally affect the pressure.
      Start sitting with : Fio2 40%, flow 8 L/ minute, and pressure 6 cm H2o