Calibration GX- 8000 - with Riken Keiki Service Engineer

Поделиться
HTML-код
  • Опубликовано: 8 ноя 2024

Комментарии • 3

  • @MerchantNavyKnowledge
    @MerchantNavyKnowledge Год назад

    How to rectify flow fail alarm, I opened and cleaned it still showing flow fail .

    • @akosiapat1140
      @akosiapat1140  Год назад +1

      check the expiry of the sensor, try to replace new one.
      if the flow fail alarm still exist, check the pump if it still working properly, no harm noise.

    • @MerchantNavyKnowledge
      @MerchantNavyKnowledge Год назад

      @@akosiapat1140 can u advise how to check expiry, it has one long number written in it and nothing else