Dear Sirs Please please 🙏🏻 Can you confirm the below parameters for sterilisation are acceptable: Machine: surgic touch Thermal disinfection in an automated washer at 90 degrees with a holding time of 1 min Sterilisation - 134 -137 with a holding time of 3-3.5 mins Kind regards Abel
Hello Dr. Abel, According to the IFU 8.4.1 Automated disinfection-Washer-disinfector, 1. The washer-disinfector is proved to be valid by CE certification in accordance with EN ISO 15883. 2. Use high temperature disinfection function. The temperature does not exceed 134℃, and the disinfection under the temperature cannot exceed 20 minutes. 3. The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883.
0:40 sec video inspection and maintenance differs from the original Gun-Fi cleaning instructions "4) Cleaning of heating chamberWhen removing the residue inside the heating chamber and theloading slot, set the temperature of the device to 200 ° C, and then poweroff the device after squeezing all the residual materials out. After pullingthe pushing ram out from back part, insert the supplied cleaning brushfrom the back of the device and pull it out from the front part of the Melting and Filling Gun. Please take care not to add any cleaning agentor chemicals to the cleaning brush." Woodpecker Fi-G Instruction Manual page 17. What is the correct cleaning sequence?
Hello dr, sorry for late reply. The way of operation in these two is different, and there is no problem in either. The video is the cleaning method after taking out the needle, which only needs to be heated to 150 ° C to clean
Dear Sirs
Please please 🙏🏻
Can you confirm the below parameters for sterilisation are acceptable:
Machine: surgic touch
Thermal disinfection in an automated washer at 90 degrees with a holding time of 1 min
Sterilisation - 134 -137 with a holding time of 3-3.5 mins
Kind regards
Abel
Hello Dr. Abel,
According to the IFU 8.4.1 Automated disinfection-Washer-disinfector,
1. The washer-disinfector is proved to be valid by CE certification in accordance with EN ISO 15883.
2. Use high temperature disinfection function. The temperature does not exceed 134℃, and the disinfection under the temperature cannot exceed 20 minutes.
3. The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883.
I need a guide bushing for my device, how yo order one?
Hello dr, you can purchase from our local dealer
What is a nut?
0:40 sec video inspection and maintenance differs from the original Gun-Fi cleaning instructions "4) Cleaning of heating chamberWhen removing the residue inside the heating chamber and theloading slot, set the temperature of the device to 200 ° C, and then poweroff the device after squeezing all the residual materials out. After pullingthe pushing ram out from back part, insert the supplied cleaning brushfrom the back of the device and pull it out from the front part of the Melting and Filling Gun. Please take care not to add any cleaning agentor chemicals to the cleaning brush." Woodpecker Fi-G Instruction Manual page 17. What is the correct cleaning sequence?
Hello dr, sorry for late reply.
The way of operation in these two is different, and there is no problem in either.
The video is the cleaning method after taking out the needle, which only needs to be heated to 150 ° C to clean
The instructions say to heat to 200℃, after finishing the remaining gutta-percha, turn off the power and remove the push rod and needle