【Inspection and Maintenance】Gutta-percha Obturation System Fi-G

Поделиться
HTML-код
  • Опубликовано: 22 окт 2024

Комментарии • 8

  • @abelmounir9526
    @abelmounir9526 2 года назад +1

    Dear Sirs
    Please please 🙏🏻
    Can you confirm the below parameters for sterilisation are acceptable:
    Machine: surgic touch
    Thermal disinfection in an automated washer at 90 degrees with a holding time of 1 min
    Sterilisation - 134 -137 with a holding time of 3-3.5 mins
    Kind regards
    Abel

    • @Woodpecker-Dental
      @Woodpecker-Dental  2 года назад

      Hello Dr. Abel,
      According to the IFU 8.4.1 Automated disinfection-Washer-disinfector,
      1. The washer-disinfector is proved to be valid by CE certification in accordance with EN ISO 15883.
      2. Use high temperature disinfection function. The temperature does not exceed 134℃, and the disinfection under the temperature cannot exceed 20 minutes.
      3. The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883.

  • @ThePioneer2011
    @ThePioneer2011 Год назад

    I need a guide bushing for my device, how yo order one?

  • @trickjacko8482
    @trickjacko8482 4 года назад

    What is a nut?

  • @DrYurievich
    @DrYurievich Год назад

    0:40 sec video inspection and maintenance differs from the original Gun-Fi cleaning instructions "4) Cleaning of heating chamberWhen removing the residue inside the heating chamber and theloading slot, set the temperature of the device to 200 ° C, and then poweroff the device after squeezing all the residual materials out. After pullingthe pushing ram out from back part, insert the supplied cleaning brushfrom the back of the device and pull it out from the front part of the Melting and Filling Gun. Please take care not to add any cleaning agentor chemicals to the cleaning brush." Woodpecker Fi-G Instruction Manual page 17. What is the correct cleaning sequence?

    • @Woodpecker-Dental
      @Woodpecker-Dental  Год назад

      Hello dr, sorry for late reply.
      The way of operation in these two is different, and there is no problem in either.
      The video is the cleaning method after taking out the needle, which only needs to be heated to 150 ° C to clean

    • @Woodpecker-Dental
      @Woodpecker-Dental  Год назад

      The instructions say to heat to 200℃, after finishing the remaining gutta-percha, turn off the power and remove the push rod and needle