Autoclave / HPHV Steam Sterilizer in pharmaceutical industry l Interview Questions

Поделиться
HTML-код
  • Опубликовано: 11 сен 2024

Комментарии • 66

  • @PharmaHunt
    @PharmaHunt 8 месяцев назад +8

    Because drainpoint is coolest point of autoclave and sterilization temprature should achieve till coolest point ,for ensuring that one temprature sensor is placed at drain point.thanks

  • @sanchalaldeshmukh2673
    @sanchalaldeshmukh2673 Год назад +6

    Please share article on F0 value and log reduction

  • @msd6178
    @msd6178 Год назад +3

    It is very important information... Thank you sir

  • @AJ-rp9vg
    @AJ-rp9vg Год назад +2

    Many chances to contamination from drain point ,that's why require temperature sensor in drain for assure steam penetration achieve sterilization

  • @sanapathisekhar9780
    @sanapathisekhar9780 Месяц назад +1

    Ans : to ensure Cold spot sterilization.
    Kudo to your piece of work 🤌🏾👌🏿👏🏿👏🏿. #valuable information.

    • @PharmGrow
      @PharmGrow  Месяц назад

      Thank you so much 👍

  • @vmahendran269
    @vmahendran269 Год назад +2

    In drain point all water condensate drain in the point only so if temperature achieved in drain point all locations will be achieved normally

  • @pulavarthijanakiram9701
    @pulavarthijanakiram9701 Месяц назад +1

    Drain temperature point cold to
    Condensate temprature to forward stream line n cool water

  • @rajamoulicherala2199
    @rajamoulicherala2199 Год назад +3

    Drain point is cold point in autoclave chamber during sterilization

  • @Kumar-sb3kd
    @Kumar-sb3kd 2 месяца назад +2

    What is difference between standard cycle and HPHV cycle?

    • @PharmGrow
      @PharmGrow  2 месяца назад +1

      There is no as such concept as standard cycle. Can you please explain for what you claim as standard cycle, so i can answer.

  • @madanprakash7445
    @madanprakash7445 2 дня назад +1

    Some answers are incomplete or partly correct.
    In Autoclave BI used are spores of Geobacillus sterothermophilus because it is most temperature resistance organism in spore form

    • @PharmGrow
      @PharmGrow  День назад

      Thanks for update 👍

  • @arun_108
    @arun_108 Год назад +5

    What is the reason for considering 6 log? Why not 5 log or 7 log? What is the significance of 6 log?

    • @PharmGrow
      @PharmGrow  Год назад +11

      Very good and expected question
      As we know that it is practically impossible to get 100 % assured on the sterilization process that it will Inactivate or kill all microorganisms.
      6 log reduction means reducing microbial population of 1000000 (1 million) to 01 i.e. percent reduction of 99.9999. This is considered as very good % reduction for microbial population and recommended as a sterility assurance level (SAL) for sterilization processes.
      Hence, considering good percent reduction, 6 log reduction is accepted for sterility assurance level. With an overkill approach we can go beyond 6 log reduction i.e. 7, 8 and so on to have additional assurance.
      One more reason to go with 6 log reduction is to meet guideline requirements.
      Guideline reference:
      PDA’s technical report No.01 (revised 2007) recommends for 6 log reduction
      EPA performance standard of Hospital / Healthcare disinfection recommends for 6 log reduction for test organism.
      Hope this will help you.

  • @Heliumvlogs
    @Heliumvlogs Год назад +1

    In a worst condition is drain point to maintain temperature.. So it taken a main control feedback...

  • @ahmadrasel6936
    @ahmadrasel6936 Год назад +3

    Is it not required to perform I-Cold test (steam penetration) test everyday, sir??

    • @PharmGrow
      @PharmGrow  Год назад +1

      Yes. According to AAMI ( Association for the medical instrumentation ) bowie - dick test shall be conducted every day.

  • @Manicstreams
    @Manicstreams Месяц назад

    What is the impact on pre and post loads if B& D test fails how you will assess it?

  • @rajeshpattnaik9480
    @rajeshpattnaik9480 Год назад +1

    Drain sensor is placed due to which all others sensor temperature is being maintained

  • @RakeshKumar-sx8zm
    @RakeshKumar-sx8zm Месяц назад +1

    If post vaccum phase is in running stage, during post vaccum phase gets low due to power interruption.
    Does it consider? Or
    Does it is under deviation?
    Cycle will be considered complete

    • @PharmGrow
      @PharmGrow  Месяц назад +1

      It is deviation and cycle will be considered complete. Deviation has to be raised for root cause analysis and impact assessment likely drying of load.

    • @RakeshKumar-sx8zm
      @RakeshKumar-sx8zm Месяц назад +1

      @@PharmGrow thanks for your information

    • @RakeshKumar-sx8zm
      @RakeshKumar-sx8zm Месяц назад +1

      What is maximum length of silicone tube/ braided hose pipe in autoclave? As per FDA norms/ EN285 or other reference?

    • @PharmGrow
      @PharmGrow  Месяц назад

      I need to check this. I will reply once i confirm

  • @shodhaharshad3043
    @shodhaharshad3043 Год назад +2

    Sir in chemical indicater bacteria present or not ?

    • @PharmGrow
      @PharmGrow  Год назад +2

      Chemical indicators does not have bacteria present. It has chemical which undergoes physical or chemical change after sterilization process.

    • @shodhaharshad3043
      @shodhaharshad3043 Год назад

      @@PharmGrow thanks sir

  • @rajarao6563
    @rajarao6563 Год назад +1

    Very nice

  • @msd6178
    @msd6178 Год назад +1

    Can please make a video about visual inspection qualification kit (Knap kit )

    • @PharmGrow
      @PharmGrow  Год назад +1

      Thanks for suggetion. This topic is already in my upcoming video list. You will get it very soon.

    • @msd6178
      @msd6178 Год назад +1

      Thank you sir... Please make it as early as possible....

  • @SudharaniReddycherla-cy4yj
    @SudharaniReddycherla-cy4yj Год назад +1

    Sir 60 temperature daggara apali hot watertho appudu hot water off cheyyali reacterki

    • @PharmGrow
      @PharmGrow  Год назад

      Please translate in english.

  • @sureshbabuyadla4049
    @sureshbabuyadla4049 Год назад +1

    Why exposed biological indicators should incubate with in 4 hours. What is the rationale?

    • @PharmGrow
      @PharmGrow  Год назад

      To get accurate and reliable results exposed biological indicators should be incubate within 4 hours.
      Exposed biological indicators should be incubated within 4 hours to detect any surviving spores, assess the effectiveness of the sterilization process, and ensure accurate and reliable results in compliance with industry standards.
      Delay in incubation may allow allow other factors to act like temperature, oxygen etc.
      Hope you get answer. I searched for guideline reference but it was not found. We can follow this as standard industrial practice.

  • @pritikagamit6714
    @pritikagamit6714 8 месяцев назад +1

    Sir, how can we know that we have achieved 6 log reduction.

    • @PharmGrow
      @PharmGrow  8 месяцев назад +2

      The concept of confirming 6 log reduction is very simple. We need not to perform testing for 6 log reduction after each routine / commercial cycle.
      We just need to perform validation of parameters (like temp, time and pressure) for sterilization cycle that it will achieve 6 log reduction. During cycle validation / revalidation only we need to prove with testing that set parameters are adequate to achieve 06 log reduction.
      For more detail info you can see my dedicated video on SAL i.e. sterility assurance level. You can get it in playlist of Injectable processing. Link given below,
      ruclips.net/video/pI4VO-psZPE/видео.htmlsi=LUTXBirId53AyC9I
      Hope this info will help you.

  • @jamalpurnareshjamalpurpark3532
    @jamalpurnareshjamalpurpark3532 Месяц назад +1

    Cold sopt

  • @satishthakur5742
    @satishthakur5742 Год назад +1

    Autoclave ki validation video bnao

    • @PharmGrow
      @PharmGrow  Год назад +2

      Ok. I will prepare video on Autoclave Validation

    • @narasimhajogu3532
      @narasimhajogu3532 2 месяца назад

      Autoclave validation video send me

  • @jeshalsinhchavda575
    @jeshalsinhchavda575 Год назад +1

    What is audit trail of KTA In QC...??

    • @PharmGrow
      @PharmGrow  Год назад

      It is as commonly referred i.e. sequence of recorded computer events that involves any activity around the operating system,applications or user actions.

  • @user-eg2ee8np2u
    @user-eg2ee8np2u 5 месяцев назад +1

    How many valves in Autoclave machine?

    • @PharmGrow
      @PharmGrow  2 месяца назад

      It is based on make, model qnd design. It is not fix.

  • @darshansinhvaghela8938
    @darshansinhvaghela8938 Год назад +2

    Sir hindi me video banavo na plz

  • @pulavarthijanakiram9701
    @pulavarthijanakiram9701 Месяц назад +1

    Wt is kelvin

    • @PharmGrow
      @PharmGrow  Месяц назад

      The kelvin (abbreviation K), less commonly called the degree Kelvin (symbol, o K), is the Standard International ( SI ) unit of thermodynamic temperature.

  • @anilshewale9649
    @anilshewale9649 6 месяцев назад +1

    Drain point is worst location in autoclave

  • @ashokbharvad6374
    @ashokbharvad6374 Год назад +1

    Pl answer hindi

  • @PardeepKumar-vd6lq
    @PardeepKumar-vd6lq Год назад +1

    Hindi mai bi nolage da do

  • @ashokbharvad6374
    @ashokbharvad6374 Год назад +1

    Please slow speak