9100c NXT Anesthesia Machine: Breathing System Assembly (4/4)

Поделиться
HTML-код
  • Опубликовано: 31 мар 2022
  • Use this video to familiarize yourself with steps to assemble the breathing system of 9100c NXT anesthesia machine after disassembly and/or cleaning/sterilization, and the necessary precautions to take. Improper assembly of the breathing system post-disassembly cleaning/ sterilization can lead to checkout failure, O2 sensor calibration failure, flow sensor calibration failure, incorrect volume delivery during ventilation and issues related to the APL valve not working. If you face any of these issues, please check the breathing system assembly as demonstrated in this video and retry the checkout. Please refer to the user manuals for complete information.