Step-by-Step Guide: Operating the DewPort™ by ZEGAZ Instruments

Поделиться
HTML-код
  • Опубликовано: 19 янв 2025

Комментарии • 7

  • @samandolawatta9261
    @samandolawatta9261 2 года назад

    which gas has to use validated instrument

  • @khanmuhammadabdullahalmamu6656
    @khanmuhammadabdullahalmamu6656 3 года назад

    Does this equipment require coolant like CO2 or liquid nitrogen to lower the temperature?

    • @ZEGAZInstruments
      @ZEGAZInstruments  3 года назад

      No. The cooling system is completely electronic. No cooling gas is needed.

  • @khanmuhammadabdullahalmamu6656
    @khanmuhammadabdullahalmamu6656 3 года назад

    Dear concern, does this equipment can measure water dew point of instrument air? Pressure 110-400 psi range.

  • @samandolawatta9261
    @samandolawatta9261 2 года назад

    which gas has to use a validated instrument

    • @ZEGAZInstruments
      @ZEGAZInstruments  2 года назад

      You can use any gas with a known HC or water dew point to validate. We recommend pure Ethane because its phase diagram (dew points at different pressures) is well known. You can also use pure propane, or any other gas with a precisely known dew point.
      Please note that this instrument measures both moisture and HC dew point using the same mechanism; i.e. unlike some other instruments, it is not using two different sensors. So if you validate with HC gas, you are assured that it is also working for moisture.