Gary Mabbott
Gary Mabbott
  • Видео 68
  • Просмотров 1 005 206

Видео

FluorescenceLab
Просмотров 4927 лет назад
This video describes some of the special considerations and artifacts that are common to recording fluorescence spectra. Although the instrument is a much older model than in current use, the issues are still of concern with newer model instruments.
Introduction to Statistics for Chemists
Просмотров 4,1 тыс.7 лет назад
This is the first video in a series of lessons about the fundamentals of statistics for analytical chemists prepared for a university course in quantitative chemical analysis.
ArduinoTitrator
Просмотров 4,1 тыс.8 лет назад
This video demonstrates an auto titration system built around an Arduino microcontroller board and a digital balance. For more information see “An Automated Titrator Based on a Microcontroller: an Interfacing Exercise for Advanced Analytical Chemistry Students,” The Chemical Educator, 2014, 19, pp 096-101. DOI 10.1007/s00897132539a (posted by permission)
Introduction to Statistics in Chemistry Lesson 3
Просмотров 3,2 тыс.8 лет назад
This lesson introduces the concept of a confidence interval. This video is part of a series of lessons created for a university course in analytical chemistry.
Finding the molecular formula from a mass spectrum
Просмотров 464 тыс.8 лет назад
This is the first in a series of 3 lessons about the interpretation of electron impact mass spectra. This video was created for a university course in instrumental analysis in chemistry. Spectra were taken from webbook.nist.gov/chemistry/ and used with permission. The isotope calculator mentioned in the video can be found at www.sisweb.com/mstools/isotope.htm
Mass Spectral Fragmentation Pathways
Просмотров 10 тыс.8 лет назад
This lesson continues the discussion about fragmentation processes that are useful in interpreting electron impact mass spectra. This video is the second in a series on the subject of mass spectral interpretation.
Mass spectral interpretation practice problem
Просмотров 8 тыс.8 лет назад
This is the third lesson in interpreting mass spectra.
Pipetting
Просмотров 1,2 тыс.8 лет назад
This lesson demonstrates proper use of volumetric pipets for accurate transfer of solution volumes in chemical analysis.
Proper use of a buret for a titration procedure
Просмотров 3,3 тыс.8 лет назад
This lesson describes important points for the proper use of a buret for quantitative analysis.
Use of cleaning solution for volumetric glassware
Просмотров 1,3 тыс.8 лет назад
This lesson demonstrates the proper handling of an acid persulfate solution for cleaning volumeteric glassware.
The proper use of volumetric flasks
Просмотров 11 тыс.8 лет назад
This lesson demonstrates the use of volumetric flasks for chemical analysis.
Standard operating procedure for HP 5890-5972 GC MS
Просмотров 13 тыс.8 лет назад
This video covers the standard operating procedure for a HP 5890-5972 GC-MS instrument.
An indirect determination using EDTA--Example problem
Просмотров 1,1 тыс.8 лет назад
This lesson demonstrates calculating the residual amount of an organic monomer determined indirectly through a titration with EDTA. This video was created for a university course in analytical chemistry.
Practice problem using EDTA
Просмотров 8 тыс.8 лет назад
This lesson demonstrates the determination of a metal based on a titration with EDTA. This video was created for a university course in analytical chemistry.
The Proton Balance Equation
Просмотров 16 тыс.8 лет назад
The Proton Balance Equation
Proton Balance Equation Example Calculation of pH
Просмотров 4,7 тыс.8 лет назад
Proton Balance Equation Example Calculation of pH
chemical activity
Просмотров 4,9 тыс.8 лет назад
chemical activity
Internal Standards
Просмотров 10 тыс.8 лет назад
Internal Standards
Significant Figures
Просмотров 908 лет назад
Significant Figures
F test for differences in precision
Просмотров 3,7 тыс.8 лет назад
F test for differences in precision
Grubbs' Test
Просмотров 25 тыс.8 лет назад
Grubbs' Test
Example calculation of the popagation of error
Просмотров 948 лет назад
Example calculation of the popagation of error
Error propagation
Просмотров 34 тыс.8 лет назад
Error propagation
t test for comparing individual differences
Просмотров 1,2 тыс.8 лет назад
t test for comparing individual differences
t-test for comparing two average values
Просмотров 7 тыс.8 лет назад
t-test for comparing two average values
t-test comparing an averge with an accepted value
Просмотров 2,4 тыс.8 лет назад
t-test comparing an averge with an accepted value
Normal Curve Class Problem 1
Просмотров 868 лет назад
Normal Curve Class Problem 1
Normal Curve Class Problem 2
Просмотров 738 лет назад
Normal Curve Class Problem 2
Introduction to Statistics for Chemists--lesson 2
Просмотров 2,2 тыс.8 лет назад
Introduction to Statistics for Chemists lesson 2

Комментарии

  • @RuhmaRehman-t9t
    @RuhmaRehman-t9t 14 дней назад

    why we didn't pick silicon instead of Sulphur as second isotope?

  • @aliahmadi7235
    @aliahmadi7235 2 месяца назад

    magnificent

  • @hemrajrimal6284
    @hemrajrimal6284 2 месяца назад

    very nice and clear video, can you please make video for Kinetics too?

  • @baigasafisaidi8141
    @baigasafisaidi8141 3 месяца назад

    Zoom the text

  • @johnlecavalier6199
    @johnlecavalier6199 3 месяца назад

    poor guy's hand probably looks like a pin cushion by now

  • @Abumustard6364
    @Abumustard6364 3 месяца назад

    Thank you for this excellent video.

  • @heeralalsoni9307
    @heeralalsoni9307 4 месяца назад

    Dear Gary, which reference electrode should be used for non aqueous system?

  • @borghiire
    @borghiire 6 месяцев назад

    Thank You so much Doctor Mabbott! These videos are very helpful and so well spoken. I'm using them to better understand these concept as my Analytical Chemestry exam is in 4 days and I want a good grade! Immense gratitude from the University of Pisa, Italy!

  • @AnyiEspana-vn3dv
    @AnyiEspana-vn3dv 7 месяцев назад

    Wow, gracias me ayudaste a entender este tema.

  • @hadassachata4270
    @hadassachata4270 9 месяцев назад

    Please come back sir, the world need you

  • @oliviawoods9239
    @oliviawoods9239 10 месяцев назад

    Thank you for explaining this so thoroughly it helped me so much!!!

  • @bfybfy438
    @bfybfy438 10 месяцев назад

    You saved me. I wish I found your video a long time ago. Thank you.

  • @Astromagine
    @Astromagine 11 месяцев назад

    Really good

  • @kumodgurav
    @kumodgurav Год назад

    I am not understanding the difference between m/z value and mass of following compound please guide Compound Name Retation Time(RT) m/z Mass Score Formula Quadrigemine A 11.715 735.4126 690.4143 90.09 C44 H50 N8

    • @garymabbott4064
      @garymabbott4064 Год назад

      Hi Kumodgurav, I am going to make some assumptions about the information that you gave me. You are reporting a retention time that suggests that your data is from an LC/MS run. Is it possible that you used an electrospray ionization (ESI) detector in the negative ion mode? ESI frequently produces adduct ions that are a combination of the analyte molecule plus some an ion present in the buffer. The mass difference of 45 amu fits nicely with the idea that the observed parent ion is the neutral drug molecule plus one formate ion [M + HCO2]- See: www.acdlabs.com/blog/common-adduct-and-fragment-ions-in-mass-spectrometry/

  • @alexivanov7653
    @alexivanov7653 Год назад

    Those assholes from Horiba didn't even bother to create a normal calibration mode

  • @nurasalah5263
    @nurasalah5263 Год назад

    VERY HELPFUL

  • @pingüinosagaz
    @pingüinosagaz Год назад

    gracias gary!! just one doubt, wich condictions are not controlled at E standard than E formal?

    • @garymabbott4064
      @garymabbott4064 Год назад

      That's a good question and you need to ask it whenever someone reports a value for a formal potential, because people do different things. However, most often people work at conditions other than unit activity for all of the reactants and products for the half reaction of interest. Rarely do people work at pH 0 (unit H+ ion activity), for example. The pH is probably one of the most influential parameters that varies among experiments. Temperature is also frequently varied. Excellent experimentalists are careful to report temperature, ionic strength, pH, reactant concentrations for any measurements that are not at standard state.

  • @Desipi
    @Desipi Год назад

    So thankful I found this lesson

  • @ZawiChristine-ju8hq
    @ZawiChristine-ju8hq Год назад

    Help me with this question.the mass spectrum of C2H5CL shows peaks calculate the mass number of most abundant molecular ion

    • @garymabbott4064
      @garymabbott4064 Год назад

      Think of a molecule with the formula that you have given containing atoms of only the most abundant isotope of each type. (If you don't have a table of isotopes for elements common to organic molecules, you can find one on-line.) For most non-metallic elements, the most abundant isotope tends to be the lightest. Do this for carbon, hydrogen and chlorine separately. Multiply those isotope masses by the number of atoms of that element that appears in each molecule and sum these numbers for all the elements in the molecule.

  • @theoverlord3839
    @theoverlord3839 Год назад

    Best lecture ive seen yet on this subject thanks man

  • @kenzadrive1359
    @kenzadrive1359 Год назад

    😍😍😍😍

  • @jonasmortier9056
    @jonasmortier9056 Год назад

    Thanks for this very informative video Gary! I have a question about the slide at 9:42 How can a working electrode that is far from the reference electrode create a potential V1+Vref at the reference electrode? Are they connected in some way?

    • @jonasmortier9056
      @jonasmortier9056 Год назад

      My question comes down to: how can you bring a solution to a certain potential? Is it because the inert electrolyte will create the potential through the solution?

    • @jonasmortier9056
      @jonasmortier9056 Год назад

      And if the potential is indeed created through the electrolytes, does the distance to the working electrode matter, or is every point in the water at the same potential?

    • @garymabbott4064
      @garymabbott4064 Год назад

      Jonas, Good questions. Your reasoning is also correct on both counts. It is a bit of a challenge to follow the operation of a three-electrode system, but basically the reference and counter electrodes work together to pump charge into or out of the solution when the power source (potentiostat) applies a voltage to the cell. (It takes very tiny amounts of charge to do this.) This excess charge migrates very quickly to the interface between the solution and the working electrode. The fact that the solution conducts charge (in the form of ions) keeps the potential throughout the solution uniform in a manner similar to the way a metal conductor would if you where charging up a pair of plates in a capacitor. We say that the reference electrode senses the voltage on the solution side. Really what the potentiostat monitors is the voltage between the reference electrode and the working electrode (which is being held at ground or zero volts). The surface of the working electrode is like the second plate in the capacitor and it is in contact with ground (that is, it is held at zero volts). So, the applied voltage is a measure of the voltage between the capacitor plates. It is the electrochemical potential energy drop across this interface between the solution and the working electrode surface and it is that energy that is available to drive a chemical reaction. You pointed out an important idea about the placement of the reference electrode. That can be important if either the electrical resistance of the solution or the current going through the system is big. It takes energy for the ions to move the charge through the solution. That energy is lost as heat. The potentiostat reports the total voltage applied to the cell. If that energy spent moving ions is a significant fraction of the energy applied to the cell, then the actual energy applied across the working electrode/solution interface will be less than the applied cell voltage. That means we will have an error in the applied voltage (the x-axis) on our voltammogram. The error is equal to the product of the current times the solution resistance. So, if we keep the supporting electrolyte concentration high (0.1 M KCl has a resistance on the order of 100 Ω per cm of solution) and the current below 10 µA, then errors are on the order of 1 mV or less. That's usually acceptable. When either the solution resistance is high or the currents are greater than 10 µA, placing the reference electrode close to the surface of the working electrode helps to reduce this type of error (also called "iR-drop" loss).

    • @jonasmortier9056
      @jonasmortier9056 Год назад

      @@garymabbott4064 A big thank you for your answer Gary! This really helps me clarify this concept. I had another question: We are interested in the potential and current flowing from the working electrode for obvious reasons. The counter electrode however is mostly there to get rid of the electrons our working electrode produces. I was wondering how the potential of this counter electrode is regulated? dependent on which reaction happens there, it seems to me that a different voltage would be needed at the working electrode to perform both the half reactions? If a reaction at the counter electrode happens that needs a high driving force voltage wise, it would seem to me that the working electrode would need to work harder (have a higher potential) to make sure the reaction happens? Or is there some sort of mechanism in the technical part that senses how many volts are needed at the counter electrode for that half reaction to happen to make up for it? I have also been struggling a bit with the concept of an electrolyte. To me, it seems that as long a reduction reactions happens at the cathode, and an oxidation reaction happens at the anode, current is flowing. Does the electrolytes facilitate the transition of the electrons from the electrodes to the species? why does the 'current' wants to go through the solvent instead of taking the electrodes as starting and ending point of a circuit?

    • @garymabbott4064
      @garymabbott4064 Год назад

      @@jonasmortier9056 Jonas, In our experiment, we really don’t have a direct measure of what is happening at the counter electrode, but we can infer a few things. We know that if the working electrode is oxidizing some species in solution, then it is pulling electrons out of the cell. In order to maintain the balance (hold the potential that the instrument is designed to control at the working electrode) the counter electrode will pump the same number of electrons into the solution. That means that the potential of the counter electrode surface will be negative with respect to the solution (just the opposite polarity that appears at the working electrode). Now, you were right to think that there must be some reaction at the counter electrode that will allow it to transfer electrons into the solution. Often there is no obvious species in solution that can be easily reduced. Then the instrument will drive the potential difference at the counter electrode more negative trying to achieve the balance (or really to keep the potential across the working electrode interface at the target value). Eventually, the potential difference at the counter electrode becomes negative enough to reduce the solvent or some species in the supporting electrolyte. In aqueous solutions we observe the reduction of hydrogen ions to hydrogen gas. (You may have noticed bubbles forming on the counter electrode.) In the opposite situation (when a reduction is taking place at the working electrode) the counter electrode is likely oxidizing water to form molecular oxygen (bubbles again) and hydrogen ions. So, the counter electrode potential is driven in the opposite direction compared to the working electrode. We don’t have a measure of the actual potential that the counter electrode reaches. It will be whatever value it needs to reach in order to pump electrons in the opposite direction compared to whatever is happening at the working electrode. The supporting electrolyte acts to carry charge through the cell in the form of ions. Our experimental cell must be a part of a complete electrical loop that provides charge to move from one side of our instrument, down a cable to an electrode, through the cell, out another electrode, up a second cable to the other side of the instrument (completing the circuit inside the instrument). The conductivity of the solution goes up (and resistance goes down) with the concentration of ions dissolved in solution and that is good for the experiment. (Recall that we want to minimize errors caused by “iR loss”-the energy wasted in moving ions through solution.) In most experiments, we expect the supporting electrolyte to be inert and not participate directly in solution. An important exception is the oxidation or reduction of organic molecules that usually are accompanied by the incorporation or release of hydrogen ions from solution. There are also very interesting reactions that involve oxidation or reduction of the solvent or supporting electrolyte that turn around and exchange electrons with the analyte species (and are re-generated back to their original state). This, of course, is a catalysis process that can be of enormous interest for theoretical and practical purposes. However, we generally expect that the supporting electrolyte is not involved in electron transfer at the working electrode in simple electroanalytical experiments (until we scan to the limit of our useful voltage range where we see increasing background current even in blank solutions containing only supporting electrolyte.) I am happy to continue answering your questions, but I would encourage you to recommend that your library purchase this book: www.amazon.com/Electroanalytical-Chemistry-Principles-Monographs-Applications-ebook/dp/B084GM6C3R/ref=sr_1_6?qid=1673106623&refinements=p_28%3AElectroanalytical+Chemistry&s=books&sr=1-6 (Yes, it is my book. I apologize for the shameless self promotion, but it could be helpful to you and others interested in this field.)

  • @riveratrackrunner
    @riveratrackrunner 2 года назад

    Why even have school when we can learn from books and internet?

  • @neopets25
    @neopets25 2 года назад

    What would happen if there were no counter electrode?

  • @Errys
    @Errys 2 года назад

    you explained it better than my university professor, thank you

  • @Getgot-en1kv
    @Getgot-en1kv 2 года назад

    How interpret the MS when the molecular peak isn’t the highest peak?

    • @garymabbott4064
      @garymabbott4064 2 года назад

      There are these situations when the molecular ion is not the tallest peak in the high mass cluster: When the sample is contaminated and more than one compound is being ionized at the same time. We can usually rule this out, if our separation and/or clean up steps are rigorous. However, when the signal is weak background peaks (trace level contaminants and electronic noise) can complicate interpretation. When the compound of interest breaks apart so easily that no molecular ion survives to give a signal. This happens with long alkyl chains, for example. That is a tough puzzle, but one can get a lot of information by applying the same strategy to the cluster of ions at the highest mass with the caveat that this cluster is due to a fragment of the molecular ion is quite possibly an even mass ion. (See Fred McLafferty, Interpretation of Mass Spectra for an in-depth discussion.) When the molecule has multiple chlorine and/or bromine atoms. In those cases the probability of having a molecule with at least one heavy halogen isotope is greater than a molecule with all light isotopes. But these cases are easy to spot, since the tall peaks are spaced by two mass units from each other. Remember that everything heavier than the molecular ion with all light isotopes must be the result of ions with a combination heavy isotopes (and the same chemical structure, namely, the molecule minus one electron). Other ionization methods that we were not considering here also give a different signal for the molecular ion. Electrospray ionization (ESI) is very important method in LC/MS and in desorption ESI (or DESI). These methods lead to much less fragmentation, but form “adducts” (combinations of the molecule plus a hydrogen ion or sometimes a sodium ion). Here the signals in the high mass cluster must be explainable by combinations of isotopes for the structure of the whole molecule plus a hydrogen, that is, (M+H).

  • @harshitshukla7208
    @harshitshukla7208 2 года назад

    @Gary Mabbott, Sir why did you consider only sulfur in the last example, why not silicon, its mass was 30 also.

    • @garymabbott4064
      @garymabbott4064 2 года назад

      Hi Harshit, That is a very good question. Perhaps, I should have addressed that in the video. Let's try it now. Going back to the table of isotopes (in the video) we see that Si would also contribute to the M+2 peak (3.4 % compared to 4.4% for S, but close enough to consider it. We want to be flexible when the intensities are weak, since the relative uncertainty in real data is often bigger with weak signals.) Now, notice that the Si also contributes 5.1 % to the M+1 peak. We need to correct the intensity of the M+1 peak for the contribution of 1 Si atom before calculating the number of carbon atoms, but we see it is bigger than the signal at M+1. (5.1% > 4.2%) If the M+1 peak is the result of a silicon atom, then there is no room for carbon. So, a silicon atom does not seem likely. Make sense?

    • @harshitshukla7208
      @harshitshukla7208 2 года назад

      Thanks for the explaination. So, sir in the nutshell, do we have to consider only those elements whose M+1 intensity percentage in the table is less than the M+1 intensity given in question? Because if (5.1% of 75.2 =3.8) < 4.1 is consider here, we might not be able to put Carbon in the formula.

    • @garymabbott4064
      @garymabbott4064 2 года назад

      @@harshitshukla7208 Yes. I think you have it. I want to apologize for sloppiness in my earlier answer. I should have multiplied 5.1% (the contribution to the M+1 peak that a silicon atom would provide) by 75.2 (the intensity of the M peak) to get 3.8 as you have done. That is the Si contribution to M+! in intensity units. That is the number to compare with the signal at M+1. 3.8 is still smaller than 4.1, but it seems to leave little room for carbon. If we go back to correct our calculation for the number of carbon atoms, we must subtract the Si contribution out first. --> [(4.1-3.8)/75.2]x100/1.1 = 0.398 carbon atoms. Much less than 1.0 does not look good. If the data were really noisy we might round up and consider it, one carbon and one silicon, but we now have to go back and account for 75 amu in the molecular ion again. That would also lead to more inconsistencies. Silicon seems unlikely.

    • @harshitshukla7208
      @harshitshukla7208 2 года назад

      Thanks sir.

  • @drmufarreh
    @drmufarreh 2 года назад

    Thanks for the valuable presentation

  • @baldudak1
    @baldudak1 2 года назад

    Thank you so much

  • @juanc3242
    @juanc3242 2 года назад

    00:32 definition

  • @yo6493
    @yo6493 2 года назад

    Thank you sir!

  • @Kaimli
    @Kaimli 2 года назад

    Gary you are a gem! My professor is fanastic in her field but I'm afraid teaching isnt quite for her, and this really helped me out!! Thank you!

  • @trevorlumaii7220
    @trevorlumaii7220 2 года назад

    Thank you very much, this video really helps clear all the confusions about how to determine molecule structures using mass spectroscopy...I will be having test on this topic tomorrow, I'm glad that I understand this concept before it's too late for me...thank you sir

  • @VKA_IITian
    @VKA_IITian 2 года назад

    best video ever to simplify the mass spectra.....................

  • @السيدةرواءعبدالعليم

    Hi, can you help me please How can I use synchronous spectrum in this device?

  • @kashmirvalley1821
    @kashmirvalley1821 2 года назад

    for the supercapacitor application how we use glassy carbon electrode

  • @Ananya_BTSArmy
    @Ananya_BTSArmy 2 года назад

    Great

  • @Wealily
    @Wealily 2 года назад

    Awesome! Thank you!

  • @emaanimtiaz4371
    @emaanimtiaz4371 2 года назад

    Can someone please explain to me...if the working electrode cannot pass current (transfer of electrons into or from the solution) then how does it reduce or oxidize the species??

    • @garymabbott4064
      @garymabbott4064 2 года назад

      Hi Emaan, I am not sure what conditions you are thinking of where the working electrode cannot pass current. You are correct that its main purpose is to reduce or oxidize species in solution. (We prevent the reference electrode from passing current so we don't disturb its potential. Could that be what you were thinking of?) We force the working electrode to different potentials in order to drive an exchange of electrons with something in solution. That potential must be near or more negative than the E˚' for the redox species in order to reduce any of the oxidized form and we monitor the current that passes in the voltammogram. (We see current for the oxidation of any reduced species when the working electrode is near or more positive than the E˚'.)

    • @emaanimtiaz4371
      @emaanimtiaz4371 2 года назад

      @@garymabbott4064 thankyou for replying! And im confused because supposedly, a polarizable electrode cannot exchange electrons with the solution. I dont understand, if its not giving or receiving electrons (as a polarizable working electrode) how is it able to reduce or oxidize species? Im really really new to electrochemistry (my whole career is in bio) so pardon me if im asking something really obvious

    • @garymabbott4064
      @garymabbott4064 2 года назад

      @@emaanimtiaz4371 Electrochemistry can be confusing, in part because of all of the jargon. The term “polarizable electrode” means that voltage (the energy for an electron) between the surface of the electrode and the solution can be changed by imposing energy (from a battery or electrical power source) from outside the cell. We need at least one other electrode in the cell in order to do that experiment. A reference electrode, such as a silver/silver chloride electrode, can be used for that purpose. A good reference electrode will not move from its “rest potential” (its value that can be calculated from the Nernst Equation for the reference reaction, such as AgCl + e  Ag + Cl-). We say that the reference electrode is non-polarizable. By selecting a non-polarizable reference electrode for a voltammetry experiment, virtually all of the energy (voltage) that we apply from the outside circuit ends up polarizing the working electrode. That is, all of the energy applied to the cell from the outside changes the voltage at the interface between the working electrode and the solution. If that voltage is big enough, it can force an electron exchange process there. A platinum wire or carbon electrodes are good working electrodes since they are polarizable and will exchange electrons readily with species in solution. One point of possible confusion with electrochemical measurements is that there are two common types of experiments. Voltammetry is the type I was referring to in the paragraph above. Voltammetry is based on applying an external voltage to the system which forces an electron exchange at the working electrode. The current is measured and its magnitude is proportional to the concentration of the electroactive species in solution. Voltammetry uses a polarizable working electrode. The other common experiment is potentiometry. The rest voltage of the working electrode is determined by the solution conditions and it is this quantity that we measure in a potentiometric experiment. In that case the working electrode or sensor should not be polarizable. We try to prevent significant amount of current going through the cell, since it could distort the voltage that appears at the working electrode. That is, allowing current in the system would push the working electrode away from the voltage that naturally develops at its surface because of the concentration of the species that it senses. I hope that helps.

    • @garymabbott4064
      @garymabbott4064 2 года назад

      So, the bottom line is that a polarizable electrode CAN exchange electrons with something in solution, if we push it to the right voltage.

    • @emaanimtiaz4371
      @emaanimtiaz4371 2 года назад

      @@garymabbott4064 thankyou, thankyou, thankyou! Iv been stuck on this for 2 days straight, watching a dozen youtube videos and googling it in a hundred different ways and NO ONE explained that. You've saved me from another sleepless night. Huge thanks

  • @serenity4457
    @serenity4457 2 года назад

    Who knows how much 200 series AA cost ? Please

  • @sydneywilliams1536
    @sydneywilliams1536 2 года назад

    Thank you for this video sir

  • @fatmaelsayed8933
    @fatmaelsayed8933 2 года назад

    thanks a million. today you kindly enlightened me with those precious tips about how molecular formulae are guessed based on the vaue of m/z ratio

  • @thetoniest
    @thetoniest 2 года назад

    counter connects to reference?

  • @brijeshparmar2836
    @brijeshparmar2836 2 года назад

    Hou can separate same retaintion time peak

    • @garymabbott4064
      @garymabbott4064 2 года назад

      Your question goes to the heart of the field of chromatography. Basically, when two compounds elute a the same time, you need to change the separation conditions (either the stationary phase or the mobile phase or both) in order to retain one more strongly than the other.

  • @mabillycoxhbdias9329
    @mabillycoxhbdias9329 2 года назад

    Thank you soooo much!

  • @wilwilliams4586
    @wilwilliams4586 2 года назад

    Very helpful and well explained! I've been looking for a good video to explain how to interpret M.S spectra.

  • @47salopubg94
    @47salopubg94 2 года назад

    good

  • @Lauren007E
    @Lauren007E 2 года назад

    I almost feel bad for the fact I'm paying my lecturer who couldn't (and also wouldn't) explain this in weeks when you managed to explain it in minutes. Thank you for the video! It was a life saver

  • @ibrahimomoyayi9325
    @ibrahimomoyayi9325 3 года назад

    Thank you!!!

  • @dawsonconlon2482
    @dawsonconlon2482 3 года назад

    what is slit width and what does it do PLEASE SOMONE

    • @garymabbott4064
      @garymabbott4064 3 года назад

      Hi Dawson, The term “slit width” refers to the physical width of the opening used to pass a narrow portion of the spectrum of light to be used in the experiment. The light from the lamp is dispersed by a grating into a rainbow across a wall inside the box that houses the monochromator. The opening in the wall (called the slit) allows only a narrow band (a small range of wavelengths) to pass out of the monochromator and on to the detector. In some contexts, slit width may also refer to the range of wavelengths (usually in nanometers for visible light) that pass through the monochromator (that is, through the physical slit). More often this wavelength range is called the "spectral bandpass". The narrower the slit width the better job the monochromator does in isolating the wavelength for the atomic absorption transition. However, beyond a certain point decreasing the slit width is detrimental because it weakens the signal intensity and, consequently, poorer sensitivity.