REPLACE LAMP CHEMISTRY ANALYZER MINDRAY BS-230

Поделиться
HTML-код
  • Опубликовано: 8 ноя 2024

Комментарии • 12

  • @mohamedasim6888
    @mohamedasim6888 3 месяца назад

    Bs 200 show too low light intensity after change new one what should I do to solve it

  • @Mantenimiento-k9s
    @Mantenimiento-k9s 24 дня назад

    Hey, BS-240 error log "dark current yis too high"- how i can adjust?

    • @gowek
      @gowek  24 дня назад

      @@Mantenimiento-k9s only replace lamp

  • @dobzinoentertainment3480
    @dobzinoentertainment3480 Год назад

    BS 200E Showing light intensity too weak.
    We have changed to Mindray lamps and not still working out....the reaction temperature out of range.
    Please what should we do right now?

    • @gowek
      @gowek  Год назад

      Login Debug. Check Dark Current

  • @rideformemories463
    @rideformemories463 3 года назад

    Mindray auto chemistry analyzer kaisa rahega

    • @gowek
      @gowek  2 года назад

      Okay

  • @swimfun1
    @swimfun1 2 года назад

    Password bs 240 we have a problem in optical filtre

    • @gowek
      @gowek  2 года назад

      Gain abnormal?

    • @swimfun1
      @swimfun1 2 года назад

      @@gowek low result glycemie 0.58

  • @sushreekalyanijena5443
    @sushreekalyanijena5443 2 года назад

    Erba h360 service password kay he

    • @gowek
      @gowek  2 года назад

      no password for H360