HPLC Maintenance - Replacing the Pump Seals on an Agilent 1100/1200/1260 HPLC Pump
HTML-код
- Опубликовано: 9 сен 2024
- HPLC maintenance tutorial on how to replace the seals on the pump head on an 1100/1200/1260 Agilent HPLC.
For more information, please visit www.Agilent.com.
You don't mention replacement of the film washers. Are these no longer being used? Recommended? Thank you
Do you have a similar video for a pump with active seal wash?
Thank you
What Agilent part no. are these seals? Thank you for the info.
At least for the 1260, it's 5063-6589.
How often does this need to be done?
Replacing the pump seals is a part of routine maintenance or a Preventative Maintenance(PM). Pump seal wear depends on usage, how often the pump is used and what solvents are being used. Agilent recommends replacing the pump seals once a year as a general rule.
Brilliant - thanks so much! I've taken over care of our Agilent GPC system from a former postdoc, and his answer to my question "when did you replace the seals in the last 2 years" was "never". So I guess I'll be buying some new seals!
Please share in india supplier