how to check load cell !! load cell test !! load cell repair !!

Поделиться
HTML-код
  • Опубликовано: 22 мар 2021
  • How to Check the Load Cell with a Multimeter
    Set the multimeter in DC millivolts and connect the output wires of the load cell to the multimeter. Supply a voltage of 5V or 9V DC at the excitation leads and place a test weight on the load cell. The multimeter will register a change in voltage measured across the load cell’s output.
    How to Test a Load Cell
    To test a load cell before putting it to use, you’ll need a multimeter and a voltage source. Measure the resistance of the input and output leads of the load cell by setting up the multimeter in Ohms. Compare the measurement values with the calibration certificate from the manufacturer to see if they closely match each other. Similarly, check the load cell for accuracy by measuring the millivolts signal from the input leads. With no force applied to the load cell, the value should be zero. Apply a calibrated dead weight as specified in the calibration certificate and compare the values again.
    How to Measure Load Cell Output
    Load cell output is measured using a digital meter. The digital meter connects to the output of the load cell. It converts the digital signals produced by the load cell into readable digital values. You can also measure the output of a load cell using a multimeter. However, a multimeter will delete the output voltage in millivolts, and will not convert it into force or weight.
    How to Check Load Cell Resistance
    A load cell test is performed using a digital multimeter. The digital multimeter is connected between the positive signal wire and the negative signal wire of the load cell. The output between them should be equal or to a value specified in the datasheet. This is the test for load cell output resistance. Now check the signal between the positive excitation wire and the negative excitation wire. They should be equal. This is the test for input resistance.
    Load Cell Value Fluctuation
    Load cell values can fluctuate due to several reasons. From physical damage such as shock loading and overloading for a longer duration to environmental conditions such as temperature, moisture, water ingress, or corrosion, the load cell is likely to produce erroneous output. Measurement values will also fluctuate if the cables break or if there’s a short circuit. To check what is causing fluctuations in the load cell value, perform a visual check to identify the fault location. Perform a zero balance check to identify if the strain gauge has undergone permanent deformation. An insulation resistance check can further help you identify if moisture is getting inside the load cell. Additionally, a bridge resistance check will determine if there’s a short circuit within the load cell.
    Load Cell Zero Drift
    Zero drift is the condition where zero measurements of the load cell change randomly under no-load conditions. It can also happen when the apparatus is loaded, and this phenomenon is called Drift. Several reasons such as mechanical errors, fluctuation in excitation voltage, and temperature variations could cause drift. To troubleshoot the load cell for zero drift, it is important to inspect the entire system.
    Load Cell Negative Reading
    Load cell negative reading occurs when the load cell is in an incorrect orientation. If the load cell is upside down, it would produce negative readings under loading. There is usually an arrow on the load cell that shows the direction of loading. A Load cell used to measure tension will not reflect negative reading if installed upside down and will result in an erroneous reading. However, if the load cell is installed correctly and the readings are still negative, verify the wire connection according to the color code specified by the manufacturer.
    Load Cell Overload
    Every load cell comes with rated capacity. Loading the load cells beyond the rated value overloads the load cell. The telltale signs of load cell overload are inconsistent display reading, reading not coming back to zero even after the load is removed, the dramatic change of zero balance, etc. Shock overload is one of the most damaging among overloads. Here the weight on the load changes to a significant degree in a very short period. Most load cells endure some overload and this value is called Safe overload. Anything beyond that can lead to permanent dam

Комментарии • 105

  • @shivpatel6017
    @shivpatel6017 3 года назад +10

    Red & Black in put
    Green & white out put he

  • @kanahiyakumar7304
    @kanahiyakumar7304 3 года назад +1

    Bahut sundar sir 🙏🙏

  • @yadveshsvs7474
    @yadveshsvs7474 Год назад +1

    आपने इस समस्या को सही ढंग से समझाया इसके लिए धन्यवाद

  • @_BHeEm_831
    @_BHeEm_831 2 года назад +1

    Well explained

  • @engineeringthought2524
    @engineeringthought2524 3 года назад +1

    very usefull information

  • @digitalelectronics6798
    @digitalelectronics6798 2 года назад +2

    Apki waja se nera problm solve huwa thanx

  • @ChandanKumar-gd8nj
    @ChandanKumar-gd8nj 2 года назад +1

    Very good 👍

  • @atharhashmi5762
    @atharhashmi5762 Год назад +1

    Good work

  • @arjunborude5340
    @arjunborude5340 3 года назад +1

    Nice concepts

  • @Funny_TV1992
    @Funny_TV1992 2 года назад +4

    Pahle pura knowladge lelo wrong video updates mt karo

  • @TRCVJU
    @TRCVJU Год назад +1

    Very useful project, nicely explained. Great efforts. Thanks for sharing.

  • @sagarsupekar8402
    @sagarsupekar8402 3 года назад

    good concept

  • @schenck4021
    @schenck4021 2 года назад +8

    Boss ....Colour code and resistance depends on manufacturer and type..😊😊

  • @vijayiparkar9193
    @vijayiparkar9193 3 года назад

    Well knowledge

  • @himalpariyar04
    @himalpariyar04 Год назад +4

    There is a little bit of confusion in input-output terminals.
    Red +input
    Black -input
    Green +output
    White -output

    • @kellaprasad1914
      @kellaprasad1914 11 месяцев назад

      Iam explain..
      Green..black...input terminal s ok
      Red ..white.....output terminals ok

    • @gangland6772
      @gangland6772 5 месяцев назад

      You are l wrong @@kellaprasad1914

    • @elite_kshatriya823
      @elite_kshatriya823 5 месяцев назад

      Red positve and black negative exciation voltage, green positive output and white negative output,loadcell giving output 2mv per volt.

  • @niteshgriyam7195
    @niteshgriyam7195 4 месяца назад +2

    Read and black input hai,or white and green output hai

  • @sarat305
    @sarat305 Месяц назад +1

    red & black humesha input me reheta he.

  • @chintanmehta4695
    @chintanmehta4695 Год назад +1

    Nice explanation sir.

  • @Funny_TV1992
    @Funny_TV1992 2 года назад +3

    Input aur output colour code pr depend nhi karta but aap regesistance check karke confirm kar sakte hai

  • @ashishgautam9189
    @ashishgautam9189 2 года назад +1

    Superb

  • @wasimahmed3586
    @wasimahmed3586 Год назад +1

    Nice .... good luck

  • @saruaralam440v
    @saruaralam440v 6 месяцев назад

    Input and output wire kesy samaj ayega.r+b input pirbe tom bhatarahy w+r input

  • @Little-Engineer
    @Little-Engineer Год назад +1

    Nice boss

  • @rohits.s1857
    @rohits.s1857 3 года назад

    Nice

  • @nikhilgawali5053
    @nikhilgawali5053 3 года назад

    Nice explains

  • @worldinstrumentationengineer
    @worldinstrumentationengineer 2 года назад +1

    Veryy Good

  • @ajaythapliyal4490
    @ajaythapliyal4490 Год назад +2

    Glat colour code select

  • @PradipranjanBiswas
    @PradipranjanBiswas 4 месяца назад +1

    Red +black- -exitation voltage green+white -signal output mili volt

  • @sambalpurisongsulhabaysi.8343
    @sambalpurisongsulhabaysi.8343 Год назад +1

    last me jo aap bole cornering bole kya agar cornering bole hain to ye kese kiya jata he....kya he cornering?

  • @saikatbaidya4999
    @saikatbaidya4999 2 года назад +1

    Need More videos on loadcell nd pcb

  • @MdJillurRohoman-xk2qb
    @MdJillurRohoman-xk2qb Год назад +1

    60/100 to 15000 Lode cell cek Batari It working OK

  • @proud_engg
    @proud_engg 2 месяца назад

    Input/Output resistance same kaise hoga??
    Input resistance high hota hai, output resistance Kam hota hai

  • @pascoscales1030
    @pascoscales1030 9 месяцев назад

    GOOD VIDEO

  • @VidarrKerr
    @VidarrKerr 3 месяца назад

    What to do when the wires are removed from the load cell? It seems no one knows how to fix this type of issue.

  • @prabhakaranthekkadavan5910
    @prabhakaranthekkadavan5910 2 года назад +2

    I see one video about loadcell, he is telling that red and black wire is input. which is correct?

    • @PappuSingh-vz1ud
      @PappuSingh-vz1ud 2 года назад +1

      Red 🔴 and black input u right this man worng information

  • @kuldeeppanchal538
    @kuldeeppanchal538 12 дней назад

    Loadcell me resistense okay hai but not ok working please send other type testing loadcell

  • @pubg_lovegame7974
    @pubg_lovegame7974 2 года назад

    Aap ne Metter glt lgaya tha is liya _ me tha output but aap ka video acha tha ji

  • @sohailalvi5816
    @sohailalvi5816 Год назад +1

    Sir agar load cell k exitation voltage 10v likhy hon 0r weighing indicator voltage load cell ko 5v de rha ho to kia ye thk hai kam chal jay ga ya nh

  • @ankitkumar-yy1zt
    @ankitkumar-yy1zt 2 года назад +2

    Bhai aadha adhura gayan leke video upload mat kia karo yaar

  • @parasgamer70
    @parasgamer70 Год назад +1

    Load shel ko e.o.t crane me connection kaise krte h

  • @ranjeetchauhan5031
    @ranjeetchauhan5031 7 месяцев назад +1

    SLI me hierr load ke jagha aarha hai kay problem hai

  • @rahulpatelgujratahmedabad3497
    @rahulpatelgujratahmedabad3497 5 месяцев назад +1

    🎉😂❤😂🎉😂❤

  • @vinodpal3457
    @vinodpal3457 3 года назад +1

    1 ton ke weigh me circuit kaisek ur callibration kaise kare

  • @thakornj9493
    @thakornj9493 2 года назад +1

    Sab loadcell me ye do kalar ke input hoga?

  • @ratiranjanbhoi2396
    @ratiranjanbhoi2396 5 месяцев назад +1

    Signal mv pe check hota hai DC V pe nahi .

  • @komalrandive1258
    @komalrandive1258 Год назад +1

    Kitne k g ka hai

  • @shai460
    @shai460 3 года назад +1

    Sir 3o kg wala kata hai light jane ke bad 2o gram - me so keta eskké bare me batae

  • @InnovateYourselfashu
    @InnovateYourselfashu 2 года назад +1

    Hey there, thank you so much your valuable content. We are working on a project so we need your help regarding that and it's really urgent. Can we please connect over a call for discussion.

  • @LarryP248
    @LarryP248 8 месяцев назад +1

    This is an extraordinary piece. There was a book I read with related content that shifted my perspective. "A Life Unplugged: Reclaiming Reality in a Digital Age" by Theodore Blaze

  • @bhurgulanaresh7094
    @bhurgulanaresh7094 7 месяцев назад

    100kg machine ko 200 kg capacity load cell fix kar sakth hai?

  • @preeturmaliya7077
    @preeturmaliya7077 3 года назад

    Usefull information

  • @vickrampupathi904
    @vickrampupathi904 2 года назад +1

    Brother please English subtitles tq

  • @KrishnaGupta-mn4td
    @KrishnaGupta-mn4td Год назад +1

    Wet raning kar raha hai kaise thik kare

  • @UmeshSharma-lx8ei
    @UmeshSharma-lx8ei 2 года назад +1

    Hydrolic load cell how to work btaye

  • @amit.negi01
    @amit.negi01 2 года назад +1

    How to calibrate load cell by 4 to 20 mA

  • @manjunathhosamani2059
    @manjunathhosamani2059 2 года назад +1

    Load cell ka healthiness kaise check karna hye

  • @HIROND
    @HIROND 3 года назад +1

    Thank you, but how to repair it when load cell is measuring continually? For example you put 2kg weight, the load cell measure 2000.00 .01 .02 03 .5 .... No stop. Thank you sir

  • @priyankayadav1531
    @priyankayadav1531 2 года назад +1

    Agar Koi load nhi hai load cell per aur input me 12v injecte kiya aur output me negative voltage dikhta hoga to kya karna hai sir

    • @mohit777100
      @mohit777100 2 года назад +1

      Madam ji multimeter probes ki polarity change kro positive me aa jayega voltage

    • @PradipranjanBiswas
      @PradipranjanBiswas 4 месяца назад

      Change polarity

  • @LokeshKumar-by4dq
    @LokeshKumar-by4dq Год назад +1

    Hii

  • @zishanytgamers5641
    @zishanytgamers5641 Год назад +1

    Hi

  • @neerajkumar9981
    @neerajkumar9981 3 года назад +2

    It's wrong information about wiring

  • @rajkishorepradhan463
    @rajkishorepradhan463 10 месяцев назад +1

    What is the common bridge resistance value of load cell

  • @biswanathkundu8600
    @biswanathkundu8600 8 месяцев назад

    You provided wrong information about connection details, not matching with the drawing you shown and as a result you did not get desire result when connectied with battery.

  • @toshanvishwakarma7720
    @toshanvishwakarma7720 4 месяца назад

    9volt input me 1 volt DC kyun

  • @kanahiyakumar7304
    @kanahiyakumar7304 3 года назад +1

    Please sir

  • @devendrasharma6404
    @devendrasharma6404 2 года назад +2

    Wrong information please read or study then upload clips

  • @kanahiyakumar7304
    @kanahiyakumar7304 3 года назад +1

    Aaap na no degea na sir

  • @shanil67
    @shanil67 Год назад +1

    Wrong information

  • @bapankundu5452
    @bapankundu5452 Год назад +1

    Black and red input and green and white output ok

  • @faizankafi
    @faizankafi Год назад +1

    Please speak english

  • @seifabdaoui4270
    @seifabdaoui4270 Год назад +1

    Speak in English please , why you don't speak in English in order to all people benefice of the knowledge .

  • @viseshsolution74
    @viseshsolution74 3 месяца назад

    go and study then explain

  • @PradipranjanBiswas
    @PradipranjanBiswas 4 месяца назад +1

    You are sumthing wrong.

  • @UnlogicalLogical
    @UnlogicalLogical 5 месяцев назад +1

    Wrong knowledge

  • @md.iqbalalam8966
    @md.iqbalalam8966 10 месяцев назад +1

    Short of knowledge, shouldn’t make any video without knowing correctly.

  • @MdKalam-ju7ph
    @MdKalam-ju7ph Год назад +1

    Nice