  • @JoeWong81
    @JoeWong81 10 лет назад

    Why did you say not to disinfect the vials that touch both the saline and eluate needles?

    • @sudo_nym
      @sudo_nym 9 лет назад

      Joe Wong
      IPA is like kryptonite for TC-99

    • @JoeWong81
      @JoeWong81 9 лет назад +1

      Sudo Nym
      Really? How so? I am a tech for a company called Lantheus and they tell me to swab the bacterial stats, saline vials, and collect evac vials right before eluting... is that incorrect?
      Thank you very much

    • @sudo_nym
      @sudo_nym 9 лет назад +2

      T99m can act like a catalyst. A reaction can occur with the dehydrogenation of IPA
      This can affect T99m 'grow back' and therefore yield.
      Not really a problem until it enters the fluid path, so it's generally safe to sanitise the tops of the 0.9% w/v saline vials, as well as the vacuum vials just prior to elution.
      So, for sanitising it's fine, but don't introduce IPA to the fluid path!

    • @sudo_nym
      @sudo_nym 9 лет назад

      Just allow it to dry first.

    • @sudo_nym
      @sudo_nym 9 лет назад

      From Lantheus' ''TechneLite' yield chart (online)
      ''Caution: Excessive use of bactericides containing alcohol may adversely affect Technetium Tc99m yield.''

  • @luckychipeya5024
    @luckychipeya5024 8 лет назад

    why double needle for the saline port

    • @sherrydepp176
      @sherrydepp176 3 года назад

      Second needle is for air vent which helps flow of saline to vacuum vail via column