New lot QC (Parallel testing)

Поделиться
HTML-код
  • Опубликовано: 26 окт 2024

Комментарии • 9

  • @bahaaseada8158
    @bahaaseada8158 Год назад

    thank you for your kind and generous science

  • @drksbabu8728
    @drksbabu8728 7 месяцев назад

    Madam, can we use manufacturer SD insted of our SD, as it is very narrow compared to manufacturer SD?

  • @bahaaseada8158
    @bahaaseada8158 11 месяцев назад

    VERY RICH INFORMATION DR ANU

  • @jaffersufi5665
    @jaffersufi5665 7 месяцев назад

    Thank you for nice teaching

  • @Ooo-u2l6t
    @Ooo-u2l6t 11 месяцев назад

    Very useful information

  • @HaroonRashid-d3u
    @HaroonRashid-d3u 11 месяцев назад

    Very informative..

  • @drpallavijain5057
    @drpallavijain5057 3 года назад

    Maam ,after finding the new lot SD, howto use them for lot to lot verification ?

    • @gordonchristophertubo3164
      @gordonchristophertubo3164 Год назад

      Use new SD to make sure assay is stable while doing parallel run for new lot, to establish again a new SD. Repeat. However, pooling SD now is a thing, and you don't have to get new SDs everytime you change a reagent lot. There's an SD pooling formula in the 24th edition of Henry's Clinical Diagnosis, Eq. 11.1

  • @unliuniverse2999
    @unliuniverse2999 4 года назад

    Thank you for this!