Dear Zedfrey Parco Please perform a vacuum test and let us know how many mbar the furnace reaches and shows after 4:50 min. on the display. Please do not hesitate to contact me directly: serviceline@ivoclarvivadent.com Best regards ServiceLine, Ivoclar Vivadent
Please help! I'm getting error 802,, (No vacuum increase could be measured. Check the following points: Is the firing chamber tight (nocontamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?Is the fuse F1 o.k.?)
Dear drtanya4ut, As soon you start the program you see a timer and the time is running back, if the timer is on zero, the program is finished. Best regards, ServiceLine, Ivoclar Vivadent
Dear Dark Dark, Error 1300 is a hint. It means, that the last calibration was made a long time ago. If you have any further questions please contact us directly. serviceline@ivoclarvivadent.com Best regards ServiceLine, Ivoclar Vivadent
tengo un p300 g2 le hize mantenimiento y ahora solo prende el led verde no sale display y no prende alguna solucion gracias
Estimado Sr. Raul Ochoa
Muchas gracias por su mensaje.
Le rogamos entre en contacto con nuestro distribuidor.
www.ivoclarvivadent.com/en/contact/distributors
Cordiales saludos
ServiceLine, Ivoclar Vivadent
I have a p300 g2 I did maintenance on it and now it only turns on the green led, no display comes out and no solution turns on, thanks
Dear Raul Ochoa
I kindly ask you to contact one of our local distributor directly.
www.ivoclarvivadent.com/en/contact/distributors
Best regards
ServiceLine, Ivoclar Vivadent
how to to trouble shoot the error 800 of programat p510, i already check the vacuum pump and clean it, cunduct dehumidification program
Dear Zedfrey Parco
Please perform a vacuum test and let us know how many mbar the furnace reaches and shows after 4:50 min. on the display.
Please do not hesitate to contact me directly: serviceline@ivoclarvivadent.com
Best regards
ServiceLine, Ivoclar Vivadent
Please help!
I'm getting error 802,,
(No vacuum increase could be measured. Check the following points: Is the firing chamber tight (nocontamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?Is the fuse F1 o.k.?)
Dear Wael Najem
If you checked all of that, then we kindly ask you to contact one of our local distributor.
Best regards
ServiceLine, Ivoclar Vivadent
How do you know if dehumidification is completed?
Dear drtanya4ut,
As soon you start the program you see a timer and the time is running back, if the timer is on zero, the program is finished.
Best regards,
ServiceLine, Ivoclar Vivadent
Can I use ATK2 for a second time?
Dear Modestas Bliudzius,
No, the ATK 2 calibration can only be used once.
Best regards,
ServiceLine, Ivoclar Vivadent
Please help error 1300
Dear Dark Dark,
Error 1300 is a hint. It means, that the last calibration was made a long time ago.
If you have any further questions please contact us directly.
serviceline@ivoclarvivadent.com
Best regards
ServiceLine, Ivoclar Vivadent
Sari aap hindi language mai video dalo na
eror 800,это что за ошибка?