Honeywell XCD - Setting the Sensor Range

Поделиться
HTML-код
  • Опубликовано: 18 сен 2024
  • Learn more about the Honeywell Analytics Sensepoint XCD Explosion-Proof Transmitter for Hazardous Gas Detection: www.lesman.com...
    Contact: www.lesman.com/
    (800) 953-7626
    Follow Lesman on Twitter: @Lesman_Inst
    Follow Lesman on LinkedIn:
    / lesman-instrument-company

Комментарии • 2

  • @dayarao1555
    @dayarao1555 Год назад

    Good explanation, im doing doing o2 calibration ,i could not get 20.9 percentage . im used surrounding air for calibration o2 as per manual. Could you advice on why i couldn't get 20.9 percentage..thank you

    • @supermarioisacat
      @supermarioisacat 10 месяцев назад

      Late reply, but better than never: Atmospheric oxygen levels vary with altitude. If your location is significantly above sea level, your actual O2 level will be less than 20.9%. For example, at 2000 feet above sea level the actual gas concentration might only be 19.2%, at 6000 feet it would be down to around 16.7%.