Honeywell XCD - Setting the Sensor Range
HTML-код
- Опубликовано: 18 сен 2024
- Learn more about the Honeywell Analytics Sensepoint XCD Explosion-Proof Transmitter for Hazardous Gas Detection: www.lesman.com...
Contact: www.lesman.com/
(800) 953-7626
Follow Lesman on Twitter: @Lesman_Inst
Follow Lesman on LinkedIn:
/ lesman-instrument-company
Good explanation, im doing doing o2 calibration ,i could not get 20.9 percentage . im used surrounding air for calibration o2 as per manual. Could you advice on why i couldn't get 20.9 percentage..thank you
Late reply, but better than never: Atmospheric oxygen levels vary with altitude. If your location is significantly above sea level, your actual O2 level will be less than 20.9%. For example, at 2000 feet above sea level the actual gas concentration might only be 19.2%, at 6000 feet it would be down to around 16.7%.