How Do I Calibrate the Crowcon T4 Multi Gas Detector

Поделиться
HTML-код
  • Опубликовано: 14 окт 2024

Комментарии • 13

  • @idrisfirdaus1467
    @idrisfirdaus1467 7 месяцев назад

    do you have key license for sotware portable Pro?

    • @Frontline-safetyCoUk
      @Frontline-safetyCoUk  Месяц назад

      This must be provided via Crowcon and requires official training.

  • @ezwan9449
    @ezwan9449 Год назад

    A bit disappointed with my T4. There's no calibration mode. I have been lock by supplier and they won't allow us to do our own monthly calibration.

    • @Frontline-safetyCoUk
      @Frontline-safetyCoUk  Год назад

      Hi there.
      The T4 does have calibration mode, it must be activated via the compatible Crowcon software to allow manual calibrations to take place. Calibrations must be performed by a trained individual. It is recommended to have the monitor calibrated every 6 months and bump test the monitor before each day's use.
      You can send your monitor to your local service centre and request that manual calibration be switched on.
      Send your monitor to us if you do not have a service centre near you. We are fully trained in the servicing off portable and fixed gas detection. See more about returning to us here - www.frontline-safety.co.uk/rma
      Thanks,
      Callum

    • @idrisfirdaus1467
      @idrisfirdaus1467 7 месяцев назад

      sudah bisa kalibrasi crowcon sendiri?

    • @ezwan9449
      @ezwan9449 6 месяцев назад

      Sudah om. tapi kamu perlukan licence key for software@@idrisfirdaus1467

  • @loayal-obaidi6893
    @loayal-obaidi6893 3 года назад

    I can’t find calibration icon

    • @Frontline-safetyCoUk
      @Frontline-safetyCoUk  3 года назад

      Hi there,
      It may be that manual calibration has been disabled via portables pro software. This can be switched back on via the software, which can be found here: www.crowcon.com/help-and-advice/technical-support/software-downloads/
      If you do not have the licensing to do so, we recommend sending your T4 unit to a local service center where they will be able to calibrate the unit and switch this setting back on.
      Hope this helps!

    • @natanwidodo5597
      @natanwidodo5597 2 года назад

      @@Frontline-safetyCoUk your local service do not want to do that so my users can not calibrate the device by themself, so we must go to your local service and pay whatever price they charge to us which is very expensive just to calibrate the device not to mention the service takes too much time

    • @impactorltdgasdetectionsou2263
      @impactorltdgasdetectionsou2263 2 года назад

      @@Frontline-safetyCoUk Hi I have same issue that it is disabled in the menu - So I did connect it to the software - but it cannot be turned on in the software - the enabling mark gray and cannot be marked to enable.....Appreciate for your advise

    • @Frontline-safetyCoUk
      @Frontline-safetyCoUk  2 года назад

      @@impactorltdgasdetectionsou2263 Apologies for the delay. You would need to be service qualified to be able to do this. If you have not been fully service trained by Crowcon, this will not be an available option. You can contact Crowcon to complete your service training or send your monitor to your local service centre.
      Thanks!

    • @ezwan9449
      @ezwan9449 Год назад

      ​@@Frontline-safetyCoUkI have been trying to get a training certify but no crowcon dealer in Malaysia and Singapore are willing to do training for me. Any online training for this?