- Видео 28
- Просмотров 56 798
Learning by Doing with Peter Tran
Сингапур
Добавлен 28 авг 2020
SECRETS REVEALED
SUBSCRIBE to my channel for a myriad of SECRETS to be revealed as no one has done this before.
Hi there, you are receiving this message from a Service Engineer who has had eight years experience with Agilent customized Gas Chromatographs.
My channel’s idea is become a reference where you can refer to anytime when you face a challenge with the following:
🌿Openlab Chemstation Intelligent Reporting: Boost your productivity by 1000 percents with the five-steps proven approach.
🌿GC Troubleshooting Series: Learn from my real hands-on experience that I face everyday.
🌿GC Operation: Understand instrument parameter, take the advantage of my experience to acquire, integrate and report each method properly.
I will be publishing at least once a week. For your own benefits I strongly urge you to SUBSCRIBE NOW and switch on the NOTIFICATION BELL.
You can reach me anytime by commenting in my videos or via triplettt1719@gmail
💪💪💪Learning By Doing💪💪💪
Peter.
SUBSCRIBE to my channel for a myriad of SECRETS to be revealed as no one has done this before.
Hi there, you are receiving this message from a Service Engineer who has had eight years experience with Agilent customized Gas Chromatographs.
My channel’s idea is become a reference where you can refer to anytime when you face a challenge with the following:
🌿Openlab Chemstation Intelligent Reporting: Boost your productivity by 1000 percents with the five-steps proven approach.
🌿GC Troubleshooting Series: Learn from my real hands-on experience that I face everyday.
🌿GC Operation: Understand instrument parameter, take the advantage of my experience to acquire, integrate and report each method properly.
I will be publishing at least once a week. For your own benefits I strongly urge you to SUBSCRIBE NOW and switch on the NOTIFICATION BELL.
You can reach me anytime by commenting in my videos or via triplettt1719@gmail
💪💪💪Learning By Doing💪💪💪
Peter.
Agilent GC Operation Tip | What is backflash and how to avoid it with a calculator ? Part 2
Part of Agilent GC Operation Tip and Troubleshooting Series this video explains the phenomenon called backflash when performing injection, what factors to look for and how to avoid it using a free tool
Part 1: What is Backflash ?
ruclips.net/video/Xr547rwBblk/видео.html
Part 2: How to avoid Backflash with the calculator ?
ruclips.net/video/AdIWKDQ71g8/видео.html
----------------------------------------------------------------------------
🌈Learn more about Openlab Chemstation Intelligent Reporting here
🌈ruclips.net/user/playlist?list...
☀️Useful skills in GC Troubleshooting Series
☀️ruclips.net/user/playlist?list...
☄️Tips and Tricks for GC Operation
☄️ruclips.net/user/playlist?list...
-...
Part 1: What is Backflash ?
ruclips.net/video/Xr547rwBblk/видео.html
Part 2: How to avoid Backflash with the calculator ?
ruclips.net/video/AdIWKDQ71g8/видео.html
----------------------------------------------------------------------------
🌈Learn more about Openlab Chemstation Intelligent Reporting here
🌈ruclips.net/user/playlist?list...
☀️Useful skills in GC Troubleshooting Series
☀️ruclips.net/user/playlist?list...
☄️Tips and Tricks for GC Operation
☄️ruclips.net/user/playlist?list...
-...
Просмотров: 563
Видео
Agilent GC Operation Tip | What is backflash and how to avoid it with a calculator ? Part 1
Просмотров 9583 года назад
Part of Agilent GC Operation Tip and Troubleshooting Series this video explains the phenomenon called backflash when performing injection, what factors to look for and how to avoid it using a free tool Part 1: What is Backflash ? ruclips.net/video/Xr547rwBblk/видео.html Part 2: How to avoid Backflash with the calculator ? ruclips.net/video/AdIWKDQ71g8/видео.html 🌈Learn more about Openlab Chemst...
Agilent Gas Saver Mode | Have you made use of it ? Chemstation Software Tutorial | Part 4
Просмотров 6083 года назад
This video provides a key understanding about Agilent Gas Saver Mode. Five specific questions will be answered: 1. What is Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 2. How to activate Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 3. Should Agilent Gas Saver Mode always be turned ON ? ruclips.net/video/gaM-itVkQEc/видео.html 4. What time should...
Agilent Gas Saver Mode | Have you made use of it ? Chemstation Software Tutorial | Part 3
Просмотров 8583 года назад
This video provides a key understanding about Agilent Gas Saver Mode. Five specific questions will be answered: 1. What is Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 2. How to activate Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 3. Should Agilent Gas Saver Mode always be turned ON ? ruclips.net/video/gaM-itVkQEc/видео.html 4. What time should Ag...
Agilent Gas Saver Mode | Have you made use of it ? Chemstation Software Tutorial | Part 2
Просмотров 2,7 тыс.3 года назад
This video provides a key understanding about Agilent Gas Saver Mode. Five specific questions will be answered: 1. What is Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 2. How to activate Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 3. Should Agilent Gas Saver Mode always be turned ON ? ruclips.net/video/gaM-itVkQEc/видео.html 4. What time should Agile...
Agilent Gas Saver Mode | Have you made use of it ? Chemstation Software Tutorial | Part 1
Просмотров 2,1 тыс.3 года назад
This video provides a key understanding about Agilent Gas Saver Mode. Five specific questions will be answered: 1. What is Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 2. How to activate Agilent Gas Saver Mode ? ruclips.net/video/iSm8rVGklu4/видео.html 3. Should Agilent Gas Saver Mode always be turned ON ? ruclips.net/video/gaM-itVkQEc/видео.html 4. What time should Agilent...
Relative Response Factor | How to apply on FID with Weight Basis
Просмотров 1,5 тыс.3 года назад
This video provides a key understanding about Relative Response Factor, how to calculate Relative Response Factor, Relative Response Factor calculation example and what can Relative Response Factor be used for when it comes to FID and Weight basis as concentration unit. 🌈Learn more about Openlab Chemstation Intelligent Reporting here 🌈ruclips.net/user/playlist?list... ☀️Useful skills in GC Trou...
Relative Response Factor | How to apply on FID with Mole Basis | Part 3
Просмотров 6723 года назад
This video provides a key understanding about Relative Response Factor, how to calculate Relative Response Factor, Relative Response Factor calculation example and what can Relative Response Factor be used for when it comes to FID and Mole basis as concentration unit. Part 1: How to calculate Relative Response Factor (Experimental vs Theoretical) ruclips.net/video/BOjP5H7Prb8/видео.html Part 2:...
Relative Response Factor | How to apply on FID with Mole Basis | Part 2
Просмотров 7663 года назад
This video provides a key understanding about Relative Response Factor, how to calculate Relative Response Factor, Relative Response Factor calculation example and what can Relative Response Factor be used for when it comes to FID and Mole basis as concentration unit. Part 1: How to calculate Relative Response Factor (Experimental vs Theoretical) ruclips.net/video/BOjP5H7Prb8/видео.html Part 2:...
Relative Response Factor | How to apply on FID with Mole Basis | Part 1
Просмотров 1,6 тыс.3 года назад
This video provides a key understanding about Relative Response Factor, how to calculate Relative Response Factor, Relative Response Factor calculation example and what can Relative Response Factor be used for when it comes to FID and Mole basis as concentration unit. Part 1: How to calculate Relative Response Factor (Experimental vs Theoretical) ruclips.net/video/BOjP5H7Prb8/видео.html Part 2:...
D2887 Simulated Distillation | Detailed Explanation | Part 3| SIMDIS Tutorial
Просмотров 9733 года назад
Part of the Simulated Distillation (SIMDIS) Tutorial, this series provides a detailed explanation about D2887 Simulated Distillation. Part 1: Retention Time Mixture ruclips.net/video/4oWuqNFsxZY/видео.html Part 2: Reference Gas Oil ruclips.net/video/dUnm2oS-Y-w/видео.html Part 3: Sample analysis with cut points ruclips.net/video/aGV6ANNBL6M/видео.html Part 4: Commercial software SDE10 Demo Stay...
D2887 Simulated Distillation | Detailed Explanation | Part 2 | SIMDIS Tutorial
Просмотров 1,6 тыс.3 года назад
Part of the Simulated Distillation (SIMDIS) Tutorial, this series provides a detailed explanation about D2887 Simulated Distillation. Part 1: Retention Time Mixture ruclips.net/video/4oWuqNFsxZY/видео.html Part 2: Reference Gas Oil ruclips.net/video/dUnm2oS-Y-w/видео.html Part 3: Sample analysis with cut points ruclips.net/video/aGV6ANNBL6M/видео.html Part 4: Commercial software SDE10 Demo Stay...
D2887 Simulated Distillation | Detailed Explanation | Part 1 | SIMDIS Tutorial
Просмотров 3,6 тыс.3 года назад
Part of the Simulated Distillation (SIMDIS) Tutorial, this series provides a detailed explanation about D2887 Simulated Distillation. Part 1: Retention Time Mixture ruclips.net/video/4oWuqNFsxZY/видео.html Part 2: Reference Gas Oil ruclips.net/video/dUnm2oS-Y-w/видео.html Part 3: Sample analysis with cut points ruclips.net/video/aGV6ANNBL6M/видео.html Part 4: Commercial software SDE10 Demo Stay...
Agilent Chemstation Tutorial | Master Different Chemstation Calculation Modes
Просмотров 3,4 тыс.3 года назад
In this Agilent Chemstation tutorial video I am going to share the math how Chemstation perform the calculations with Area%, ESTD, ESTD%, NORM%, ISTD and ISTD%. 00:00 Introduction 00:33 Area% Report 03:40 ESTD Report 06:02 NORM% Report 08:50 ESTD% Report 11:42 ISTD Report 15:05 ISTD% Report 🌈Learn more about Openlab Chemstation Intelligent Reporting here 🌈ruclips.net/user/playlist?list... ☀️Use...
How to Calculate Limit of Detection | Peak Width | Openlab Chemstation Intelligent Reporting
Просмотров 1,1 тыс.3 года назад
This video provides a guideline on How To Calculate Limit Of Detection based on Noise | Peak Width | Area, which is similar to Signal to Noise approach, using Openlab Chemstation Intelligent Report. Signal to Noise Approach: ruclips.net/video/hyEDyslYcoU/видео.html 3 Standard Deviation Approach (With Valid Blank): ruclips.net/video/qf_LRxM8ql0/видео.html 3 Standard Deviation Approach (Without V...
3 Tips to Operate Chemstation like an Expert | Agilent Chemstation Tutorial
Просмотров 4,4 тыс.3 года назад
3 Tips to Operate Chemstation like an Expert | Agilent Chemstation Tutorial
How to Calculate Limit of Detection | 3 SD without Blank | Openlab Chemstation Intelligent Reporting
Просмотров 1,7 тыс.4 года назад
How to Calculate Limit of Detection | 3 SD without Blank | Openlab Chemstation Intelligent Reporting
How to Calculate Limit of Detection | 3 SD with Blank | Openlab Chemstation Intelligent Reporting
Просмотров 2,8 тыс.4 года назад
How to Calculate Limit of Detection | 3 SD with Blank | Openlab Chemstation Intelligent Reporting
Agilent GC Troubleshooting Series | Autostart Macro failed
Просмотров 3,8 тыс.4 года назад
Agilent GC Troubleshooting Series | Autostart Macro failed
Agilent GC Troubleshooting Series | MethodFile$ key in section [PCS,1] of Chemstation.ini not valid
Просмотров 2,5 тыс.4 года назад
Agilent GC Troubleshooting Series | MethodFile$ key in section [PCS,1] of Chemstation.ini not valid
Repeatability Calculation | Case 3 | Openlab Chemstation Intelligent Reporting
Просмотров 7464 года назад
Repeatability Calculation | Case 3 | Openlab Chemstation Intelligent Reporting
Repeatability Calculation | Case 2 | Openlab Chemstation Intelligent Reporting
Просмотров 8624 года назад
Repeatability Calculation | Case 2 | Openlab Chemstation Intelligent Reporting
Repeatability Calculation | Case 1 | Openlab Chemstation Intelligent Reporting
Просмотров 1,3 тыс.4 года назад
Repeatability Calculation | Case 1 | Openlab Chemstation Intelligent Reporting
How to Calculate Standard Deviation | Agilent Chemstation Tutorial
Просмотров 1,2 тыс.4 года назад
How to Calculate Standard Deviation | Agilent Chemstation Tutorial
How to Calculate Heating Value of Natural Gas | Openlab Chemstation Intelligent Reporting
Просмотров 4,4 тыс.4 года назад
How to Calculate Heating Value of Natural Gas | Openlab Chemstation Intelligent Reporting
Agilent GC Troubleshooting Series | Heater Sensor of an Inlet or Detector
Просмотров 2,6 тыс.4 года назад
Agilent GC Troubleshooting Series | Heater Sensor of an Inlet or Detector
How to Calculate Limit of Detection | Signal to Noise | Openlab Chemstation Intelligent Reporting
Просмотров 6 тыс.4 года назад
How to Calculate Limit of Detection | Signal to Noise | Openlab Chemstation Intelligent Reporting
How to Convert Mole Percent to Mass Percent | Openlab Chemstation Intelligent Reporting
Просмотров 1,7 тыс.4 года назад
How to Convert Mole Percent to Mass Percent | Openlab Chemstation Intelligent Reporting
Thank you so much from Korea.
Hello, thank you for your videos, they are very helpful in clarifying some doubts. Query: is the data processing similar for ASTM D7169?
Yes, D7169 works exactly the same way, except an additional step of calculating total theoretical area so when converting accumulated area to weight percentage, you gotta divide it to total theoretical area.
@@LearningbyDoingwithPeterTran Thank you so much!!😊
I learned that RF is Peak Area/Compound Amount, but why is it the opposite here?
Thank you for the great video. However, I don't quite understand the process of deriving Theoretical RRF. Even after looking into Scanlon's approach, I'm still having trouble grasping it. Would it be possible to provide further explanation?
Thankyou,This is vedio is very useful,but i have small doubt any disadvantage of gas saver mode.
Hello, the only drawback I have ever encountered was the Fid baseline was upset for a few seconds when activating the gas saver mode from more than 500 mL/mjn down to 20.
Hi I am QC. Which version used? Couse I do step by step but don't have?
I configured the GC when trying to configure the headspace autosampler, it says validation failed after entering the correct IP address of the headspace
How to fix the thermal shutdown issue
Hi Peter, Thank you for this great video. We can calculate the LOD that equals Cblank + 3*SD. Then, how can we calculate the LOQ from this approach? Please help. Thank you very much!
Hello, LOQ is typically calculated as Blank + 10SD. Statically speaking, if you calculate SD from 8 runs just like LOD, 10 SD is equivalent to 99.999% probability that the signal is not noise. In short, with 8 runs, 3 SD (or LOD) is 99% while 10 SD (LOQ) is 99.99%.
which column you suggest to setup for helium carrier and nitrogen carrier?
Hello, there are many vendors providing the suitable column for both He and N2. Restek is one of those.
Thanks Peter. I find it helpful knowing that 3 x SD means that 99% of the noise values will be captured in the calculated noise. This means anything above the 3 x SD value is likely to be real signal and not noise.
Thanks, glad you found it helpful.
Thank you for your informative video about RF and RRF. Myself Dr. Raj, and I am using TCD and FID for CO, CO2 quantifications and FID for hydrocarbons and oxygenates (GC 7890A). Can you please do a video for oxygenated products RF along with HCs in FID. When you are using two detectors for total product analysis how to get correct quantifications to get a good carbon balance? Thanks for your time and consideration. Looking forward your reply.
Hello, unfortunately oxygenates do not follow the relative response factors as of the normal hydrocarbons. You will need to determine their response factors experimentally.
Man, I'm sorry to tell you, but that is not how you calculate repeatability at all :/
Hello, would you mind sharing more ?
@@LearningbyDoingwithPeterTran Maybe you explained it in your other videos, I didn't watch them, but what you calculated is if second value is in repeatability range. Your math to check if value is in proper % range is good, but methodology is not. One cannot calculate repeatability using only 2 values. Assuming that your instrument has random distribution you have to have more average values from run, then calculate relative standard deviation (RSD) and if the RSD is in the range you can say that the process is repeatable or not.
I struggled with this for too long. Thanks for the quick and easy instructions!
You are welcome
Can you make a tutorial on integration events using chemstation ?
That would be too general to talk about. However using timed peak width and slope sensitivity would be more than enough for most if not all chromatograms.
Can this formula applied on TCD .?
Hi, no it does not apply to TCD
Thank you for the detailed explanation.
You are welcome.
which one do you trust more this one or D86? is it usually more expensive
Hello, D2887 reports weight% while D86 reports volume%. Unless you are converting weight% to volume%, I don’t see any points of comparison.
Thank you very much Mr Peter. well explained
You are welcome. Make sure to check out other parts to get a full understanding of this method.
Thank you so much!
You are welcome !
감사합니다! 문제해결 했습니다
물론이죠
Hello! Your videos are a blessing! Thank you for taking the time to teach. I have one question. How does one calculate threshold? We’re using an Agilent GCMS with chemstation. Trying to learn more before we start running EPA 524.2 Our baseline right now is 10000 abundance and wish to decrease this to see our low concentration standard of .5ppb Our threshold is set at 100 now. How is threshold determined?
Hello, it seems like you are referring to the threshold in MSD Chemstation’s integration. If so, there is no formula to correlate its value to concentration as far as I know. You would probably want to run your 0.5 ppb first then slowly adjust the threshold until your target peaks start integrated.
Thank you Sir!
I thought I broke chemstation today. Tomorrow I will fix it. Thanks for the help!
You are welcome. Hope the video points you to the right direction.
Update: I was able to update the initialization sequence file name to an existing file in the chemstation.ini file. I'm working with v2. something... I would never have figured this out on my own. Thank you again! Now if only I could fix the pump on my prep HPLC so the flow doesn't randomly cut out me.
Great to hear that. Cheers.
Thanks for the videos Peter! Hope you 're well!
Thanks for viewing, have a good day
I am using gas saver mode on , my gc is not getting ready and showing waiting for prep run
Hello, that’s normal. When you start the data acquisition, the GC will clear the gas saver mode automatically to get it ready for injection (by default). Alternatively you can always press the prep run button before your analysis.
You have to press configuration button on GC keypad and select instrument option there, you will find autoprep run On and off you can choose as per your need
From where gas saver flow come and In which part it is flowing , is it coming from inlet side or where? Please answer urgently
It’s coming from your external gas supply of course. The flow rate is controlled by the EPC. When gas saver mode is active, the saver flow is the split flow. Please refer to Part 1 for more details.
Hi. How do you get the menus in the report editor to show the ‘code’ menu option. My software only shows the options above it
Hello, the option is visible by default as long as the version you are using is C.01.09 or later. I’m pretty sure C.01.07 and earlier do not have the custom code option but not quite sure if it would be available in C.01.08
Thank you so much!! I am new to SimDis and need to understand this for my job! Just this one video has been very helpful. I will watch your other videos too. Cheers!
Glad you find it useful. Feel free to drop any questions you might have here.
Thanks so mucho, from Mexico. very useful information, i have to the 7890B CG and you help to resolve the problem
Glad you solved the problem
Thankyou sir. Our HPLC is 1220 infinity lc and we have done configured as per the instructions. There is no error now. But we couldn't find pump, DAD, sampler options in instrument control. Please help with that sir
I'm not very familiar with LC configuration so I guess you would need to reconfigure the modules
Hi Peter, I followed your instructions, but I still got the same error. The interesting part is I can launch the instrument offline no issue. Any other fix for this? Thanks
Hi, the easiest option would be to configure a new instrument and copy the data, methods, sequences, templates over. If that still does not work high chance the software has corrupted that you might need to reinstall it.
hello brother, can you help me? I don't get the chemical structures when checking the peaks. that started since I reinstalled the chemstation. Thank you
Did you mean the software you are using is MSD Chemstation ? If so you would have to configure the library to the software in order to show the chemical structure of the hint during MS search. Let me know if it’s the case then I can walk you through the process.
@@LearningbyDoingwithPeterTran Hello, I am writing to you from Colombia. what you say is correct. When I want to review a peak, the usual information appears, only the chemical structure no longer appears. If you can guide me, I would appreciate it. I searched the web but didn't find any help. I remind you that my MSD-GC (5975-6850) works with Chemstation and that software is already old. this is what i tell you. minute 5:30. the information appears minus the structure. ruclips.net/video/Wef1yjrjU1Y/видео.html Thank you
Have you linked the NIST library (disc 2) to the software ? You could possibly switch to the Demo.L first and load a demo data which should show the chemical structure to ensure the software is working properly.
I want to use the standard deviation and mean to calculate the detection limit by running blanks. Can this work? I tried on a sequence I ran but it says "no (or not enough) sample runs with cal. compounds available"
Yes both mean and SD are available in the statistical report for your LOD calculation. For the message you are getting it’s probably your sample type defined in the sequence does not match sequence output number 8 as it is strictly for “sample” type only. You may want to try output 7 or 9 (I can’t remember) but it’s said for calibration type. Alternatively reprocess your sequence to change all data to “sample”. If you happen to use Openlab Chemstation head over to my other video in the limit of detection series in which i am showing how to generate a LOD report directly from the software (using mean and SD of course).
Thank you very much Peter for the excellent work about method ASTM D2887.
Glad you find it useful
Excellent one. Clear and precise explanation on how the software does the calculation. Thumbs up (y)
🤙🤙🤙
Thanks Peter, very helpful. Just solved the same issue like this. Thanks Buddy!
Glad to hear that pal
tq so much! it works. i was so scared when this happened. tq again
Glad that it helps
Thank you
You are welcome
nice explanation with one mistake. for Volume Basis as per presentation your equation = CV mix * P / R * T. as your excel sheet formula you made it CV mix * P / T / R :). @ 6:51.
Thanks for your comment. In excel when you type CV * P / R / T it’s the same as CV * P / (R*T) so the one I did was actually correct.
Check out other useful operation tips at ruclips.net/p/PLaYUUU64uRTlMPElafU9KA08ciJ8qZQDj
All you need to know about gas saver mode is covered in this series ruclips.net/p/PLaYUUU64uRTlLGaHHAIbHhc4a-QcnXB2Q
All you need to know about gas saver mode is covered in this series ruclips.net/p/PLaYUUU64uRTlLGaHHAIbHhc4a-QcnXB2Q
All you need to know about gas saver mode is covered in this series ruclips.net/p/PLaYUUU64uRTlLGaHHAIbHhc4a-QcnXB2Q
All you need to know about gas saver mode is covered in this series ruclips.net/p/PLaYUUU64uRTlLGaHHAIbHhc4a-QcnXB2Q
Incredibly practical and explanatory. There are things I didn't know that could be done. Thank you so much for sharing your knowledge!
Glad it was helpful ! Let me know if you have any problems when trying on your own.
Concise👍
Glad you find it useful.
GS Flow to replace Split flow, thanks for the pointer!
Thanks for watching buddy
That was a great sharing! Thanks for sharing Peter!
My pleasure ! Let me know if you have any questions
Are you currently using Gas Saver Mode and how fast are you changing the carrier gas cylinders ?
Check out the other videos about Relative Response Factor at ruclips.net/p/PLaYUUU64uRTmWpuzym_3R8K2hnAYfCZ1F
Hi sir, can you explain me what does it mean level in calibration curve? Thank you.
Hi, a simple way to think of “level” is a storage of a number of compounds, the concentration of each compound and the response (area or height) of each compound. So if you plan to build a calibration table with only one level, all of your compounds of interest should be included in this level. If you build a multiple level calibration table, you can have some or all compounds in level 1, some or all compounds in level 2...It’s a bit confusing to describe so i think it’s best if you can share exactly how many compounds you want to add to calibration table and the list of your calibration standards then i will be able to give you a more precise way to set up the calibration table.
@@LearningbyDoingwithPeterTran thank you sir 🙏🙏🙏
Can you explain how to add A control sample in to a already method
@@TherealSunshyne Do you mean to have the method flagged out a control sample that fails your predefined acceptance criteria ? If so that gotta be something to do with Intelligent Reporting. Also let me know which software version you are using.