- Видео 104
- Просмотров 87 741
Biomedical Engineer Abid
Пакистан
Добавлен 9 фев 2011
Welcome to Biomedical Engineer Abid-your go-to destination for all things Biomedical Engineering and Ophthalmoscopy Equipment! 🌐🔬
👩🔬 Dive deep into the fascinating world where technology meets healthcare as we explore the groundbreaking field of biomedical engineering. Join us on a journey where innovation and precision converge to revolutionize medical diagnostics and treatment. Whether you're a seasoned professional in the field or a curious enthusiast, our channel is tailored to cater to all levels of interest and expertise.
At our channel, we're not just exploring the present of biomedical engineering and ophthalmoscopy equipment; we're shaping the future. Subscribe now, hit the notification bell, and join us on this exciting journey of discovery and innovation! Together, let's illuminate the world of healthcare through the lens of technology. 🌈💡 #BiomedicalEngineering, #Ophthalmoscopy, #HealthTechInnovation
#biomedicalengineerabid
👩🔬 Dive deep into the fascinating world where technology meets healthcare as we explore the groundbreaking field of biomedical engineering. Join us on a journey where innovation and precision converge to revolutionize medical diagnostics and treatment. Whether you're a seasoned professional in the field or a curious enthusiast, our channel is tailored to cater to all levels of interest and expertise.
At our channel, we're not just exploring the present of biomedical engineering and ophthalmoscopy equipment; we're shaping the future. Subscribe now, hit the notification bell, and join us on this exciting journey of discovery and innovation! Together, let's illuminate the world of healthcare through the lens of technology. 🌈💡 #BiomedicalEngineering, #Ophthalmoscopy, #HealthTechInnovation
#biomedicalengineerabid
Видео
Service mode basic test detail of perimeter (PTS 2000 Optopol)
Просмотров 31Месяц назад
How to check the hardware of the Visual field (Perimeter)?
Booting sequence of Visual field Perimeter PTS 2000, Optopol.
Просмотров 16Месяц назад
How Visual field Perimeter normally starts after pressing ON button? PTS 2000, Optopol.
V7: Multimeter & components testing Part 7 (MOSFET checking)
Просмотров 293 месяца назад
How to check MOSFET?
How to use Optical Biometry software of Optopol?
Просмотров 973 месяца назад
Optical biometry, Optopol.
Multimeter & components testing Part 4 Variable resistances
Просмотров 293 месяца назад
Types of variable resistances & it's testing.
Multimeter & component testing Part 3 (Resistances)
Просмотров 613 месяца назад
How to measure resistance?
Multimeter & components testing Part 2 (Fuses)
Просмотров 293 месяца назад
Types of fuses & testing/checking.
How to make OCT Scanning Protocols? #biomedicalengineer #ophthalmology #Opticalcoherencetomography
Просмотров 2487 месяцев назад
Welcome to our channel! In this video, we delve into the intricate world of Optopol OCT protocols, unravelling their significance in assessing various aspects of eye health. Whether you're a healthcare professional or an individual keen on understanding eye diagnostics, this video offers valuable insights into the protocols essential for comprehensive eye examinations. Optopol OCT, renowned for...
Indian Aapa YAG LASER parts detailing, Model 307. Part 1/2.
Просмотров 5911 месяцев назад
Troubleshooting: The YAG Laser was not firing. After replacing the YAG LASER PCB, the problem has been fixed.
How to download OCT software from Onedrive?
Просмотров 70Год назад
How to download OCT software from Onedrive?
Surah Anfal (Ghulba e Deen miss understanding)
Просмотров 5Год назад
Surah Anfal (Ghulba e Deen miss understanding)
Topography Part 21/21, Last step for Excimer patient selection.
Просмотров 43Год назад
Topography Part 21/21, Last step for Excimer patient selection.
How do I select an excimer patient for surgery? Topography Part 20/21, #biomedicalengineer
Просмотров 26Год назад
How do I select an excimer patient for surgery? Topography Part 20/21, #biomedicalengineer
Topography Part 19/21, shapes & evaluation.
Просмотров 51Год назад
Topography Part 19/21, shapes & evaluation.
How to read oct report please make a video on this topic
Please check my playlist having name OCT INTERPRETATION.
Very Informative 💯
Yes
Can we measure iris thickness and curvature using this oct?
Yes
Great job sir G
Thank you
You're welcome
thank you for the video! I had everything set on default and the colors looked really dull in some illustrations. now I'm surprised as to how close it it to what's on my 95% sRGB screen. personally, it made a big difference changing the gamma from 2.2 to 1.8 when setting it up to Adobe RGB. cheers!
Can you make a details video about OCT for beginners?
Yes but would you please elaborate that what kind of information you want to have in that video?
Very help video , Thank you
You are welcome
Sir print seting photopepor grosiy deneki bad black aarahahi kiyakare
Share the pic
very nice and informative video, Thank you so much for great information
Glad it was helpful!
Very informative video
Glad you think so!
Thanks so much for your precious help
Sir can u please explain interpretation of optopol visual field Also in comparison with humphrey
I have in my plan to spare some time and upload a video on interpretation as well
MASHALLAH
Abid Bhai W.salam Kaise Ho Bhai Print Cliear Nahi Arahi 😇😇😇😇😇😇
We have an OCT machine just like the one in the video, but the fundus camera is not activated, can you tell me how to activate it please. Thanks in advance.
Which model do u have? Is it Revo NX or Revo FC?
Very informative! What does that yellow triangle means on right eye?
Thanks... Which triangle? Will you pl pin point?
sir a data shit kaha se milega jo aapke paas hai
It is available in operating manual
ماشاء الله Excellent
Thanks dear. How are you? I don't have your Whats App number.
Funds mosaic is possible in Revo Fc model...? Funds update is possible in new Revo 60...?
Not yet.
@@MrAbidchaudhary ok OCT optical biometer possible in NS-3 or NS-4....?
To be sincerely honest... in my humble opinion, articulating succinctly and without a hidden agenda, visions of grandeur, ulterior motives, sentimentalism, sarcasm, becoming nostalgic or disingenuous and of course, without offending anyone who holds views contrary to my own, and without burying any impure thoughts within my subconscious and without lies or deception, to the actual truth with my clear open mind and loyal loving heart, expressing whatever has been embedded inside me for a long time which I didn't share with others because I was petrified of doing so, is about to be revealed without myself becoming a slave to fear and/or a co-signee of preposterous political correctness. Today, by gathering all the courage and motivation I could muster, I just want to state the fact that I actually feel and think that I have absolutely nothing prophetic, inspirational or even remotely interesting to say.
What the
Hey boss.. do you have the measurement of voltage/wattage of slit lamp bulb for illumination? Thanks
12 Volt/30 Watt
@@MrAbidchaudhary what type of 12v 30w? Pin bulb or Osram 64260 30w 12v PG22 M 185 Halogen Lamp Medical Bulb
HALOGEN, 12V 30W, CSO
Sir, what do you think the problem of system error code E06-Y SHUTTER SENSOR? Is it repairable or replaceable? Or how to do? Or maybe if you have the instruction, we will buy the parts and let our enggineer install the parts
Try to clean the shutter first.
Why does the inferior keratometry not appear?
We have to select the option to activate it.
@@MrAbidchaudhary I am thinking of buying one. But I still have many doubts about the topography, really the keratometric map is more like that of an AKeratometer than a topographer. Why can't you see the upper and lower keratometris, right and left, at 3mm, at 5mm central cornea.
As per my opinion this is the best option if we compare the price vs features ( available in the market so far)
Very informative 👍
Glad it was helpful!
Thx Sir Abid for provided these details its very informative for me.
Dose the spot size shown on the screen the actual size on the retina? Do i need to multiply with the lens factor?
You will need to multiply it with lens factor.
Hi doctor So what to do to cure the red parts ?? Is it reversible??
Please send your reports at +923219445989
Sir Its Very Informative 👍 Thx for share ur knowledge with us 😊
i hawe have 1 ellex integre. e604 gives an error. the equipment does not work. can you send me a service manual or can you help?
Please Whats App the Error picture at +923219445989
Hello? Do you know where can i find this battery? Best regards
Hello, Do you know where can i find this battery? Best regards
Sorry , I couldn't read your message. Have you arranged the battery or still finding?
Very Informative. 👍🏽
Thanks for the upload, keep up the good work!!! Good luck with growing your channel! Did you ever look into using promosm!!? You should use it to promote your videos!!!
Thanks dear. Thanks for the tip. I will check it out. Your comments are valuable for me.
Can you give us information on tomey oa-2000 optical biometer
Ok. I will make a separate video in it
Thank you sir
Welcome
Thank you sir
Welcome
Easy to understand
Sound quality should be good.
Ok
Its Really Informative but sound not clear soo i camt understand some points
Thanks. Trying to improve the sound quality
Very helpful Sir
Thanks and welcome
Wonderful Information God bless you Abid Bhai
It's my pleasure
Means k hamny kuch patient oct ki hai to unka fundas clear nh aata because reason ya hai k vision nh hai patient ki ap btao k ham usey ksy clear kry
Use external target. Option available in acquire mode.
Sir agar koi patient coprative na hu to ham kasy usy oct kry
Try to use Alkaine drops if Patient is blinking too much.
Sir OCT SCANS ki videos kese generate krni hai us pe video bnayen
Ok
I have uploaded the video. Hope it will help you out.
ماشاء الله. Brother Excellent Farooq
Nice
Very informative and detailed. 👍
Great sir
This video is informative. I've been trying to research for a vid that breaks down the ideas in this RUclips video! 🙌 👍Your explanation totally is similar to the content of this new medical student Dr. Ethan. Ethan's explanations are knowledgable and he really helped me a lot on my exams. Go watch his page out and give the medical student a like here! ➡️ #DrEthanEducation
Thanks. I am trying to open your given link but nothing there.
Hello. If you use patter 6 spots and power 240. It’s mean this 240 separate on 6 spots? Or every spots will have power 240mw?
Every spot will have 240 mw Power. If you are using Laser lens then depending upon the magnification of lens you will adjust spot size and power will react accordingly. That will be explained in my second video which will be uploaded later on.
@@MrAbidchaudhary ok. Thank you. I have a case when I need power 320mv if I use 180 I can’t do laser coagulat is it normal or not. I use volk g3 for laser.
@@ibragimov1987 What is the spot size of Laser?
@@MrAbidchaudhary 200
@@ibragimov1987 Ok. At this spot size you will have to increase the Power gradually. If your Laser is of dual wavelength and media is not clear then yoy may try Red Laser for deeper penetration. Depends case to case. But dont worry you may increase the power at this spot size but be careful near macular area. Thanks