Introduction to pipe sweating and tools you'll need for the job.
reposted due to sideweays view.
Intro to pipe sweating. I'll show you the tools and a technique
Просмотров: 191

Видео

Whats Inside; brushless rotary vane electric motor from a medical vacuum pump system
Просмотров 566День назад
Whats Inside; Vacuum Pump
Mindray Datascope DPM6 module not recognized
Просмотров 18День назад
Mindray Datascope DPM6 Module is not recognized. I'll show you how to fix this and open the unit
Mindray Datascope DPM6 AA/CO2 calibration
Просмотров 31День назад
AA/CO2 calibration on a Mindray DPM6 patient monitor
Whats Inside; an Electrical Contactor
Просмотров 12621 день назад
What's inside; pulling apart an electrical contactor that burned up
Steris Safety Pressure Valve Disassembly
Просмотров 4221 день назад
Another Episode of What's Inside; Disassembly of a Steris Safety Pressure Valve, 50psi
Zoll AED Plus Inspection
Просмотров 2721 день назад
Zoll AED Plus testing, setting up, brief explanation of function/use
Philips Pagewriter TC70, printout is too light
Просмотров 4128 дней назад
now and then we get the complaint about the printout being too light, but it's because they are using the wrong paper
GE Corometrics 170 series Fetal monitor
Просмотров 98Месяц назад
Fetal Monitor Overview
A very brief intro to plumbing fittings
Просмотров 1,4 тыс.2 месяца назад
very brief introduction to types of plumbing fittings
Steris Amsco 444 Instrument Washer, rack adjustments
Просмотров 982 месяца назад
Amsco Steris 444 Instrument Washer adjusting the rack
Steris Amsco Century V120 water manifold rebuild
Просмотров 693 месяца назад
water manifold rebuild for a steris v120 century. also applies to the v116, and 400 series. part numbers may differ however for the 400 series. this also solved a insufficient vac level for leak test issue. but on that topic, still have to. heck venturi, heat exchanger, and exhaust manifold(already rebuilt that one)
Drucker Centrifuge Testing
Просмотров 444 месяца назад
Drucker Centrifuge Testing
Tuttnauer EZ10 In/Out test
Просмотров 584 месяца назад
Tuttnauer EZ10 In/Out test
Steris Irrigation pump troubleshooting and explanation of how devices works
Просмотров 874 месяца назад
Steris Irrigation pump troubleshooting and explanation of how devices works
Steris/Amsco century V120 steam trap replacement
Просмотров 924 месяца назад
Steris/Amsco century V120 steam trap replacement
Steris Harmony LED585 surgical light, 1 head not working
Просмотров 1204 месяца назад
Steris Harmony LED585 surgical light, 1 head not working
Capacitors 101
Просмотров 444 месяца назад
Capacitors 101
Steris Century - Exhaust manifold pull apart for service
Просмотров 274 месяца назад
Steris Century - Exhaust manifold pull apart for service
Valleylab/Covidien Force FX(C) calibration step 7 - checking and adjusting the reactance gain
Просмотров 775 месяцев назад
Valleylab/Covidien Force FX(C) calibration step 7 - checking and adjusting the reactance gain
Valleylab/Covidien Force FX(C) calibration step 8 - checking and adjusting the ECON factor
Просмотров 1485 месяцев назад
Valleylab/Covidien Force FX(C) calibration step 8 - checking and adjusting the ECON factor
Valleylab/Covidien Force FX(C) calibration steps 1 thru 4
Просмотров 1035 месяцев назад
Valleylab/Covidien Force FX(C) calibration steps 1 thru 4
Mindray/Datascope Passport 2 CO2 calibration
Просмотров 1025 месяцев назад
Mindray/Datascope Passport 2 CO2 calibration
Stryker 737 Stretcher not locking/brief description of issues to look for
Просмотров 305 месяцев назад
Stryker 737 Stretcher not locking/brief description of issues to look for
Tuttnauer EZ11Plus how to access menu and change measurement settings
Просмотров 885 месяцев назад
Tuttnauer EZ11Plus how to access menu and change measurement settings
Tuttnauer - Door Gasket Replacement on a Tuttnauer EZ11 Plus Sterilizer
Просмотров 1435 месяцев назад
Tuttnauer - Door Gasket Replacement on a Tuttnauer EZ11 Plus Sterilizer
Philips Pagewriter TC70 how to get into service mode and run tests
Просмотров 1786 месяцев назад
Philips Pagewriter TC70 how to get into service mode and run tests
Tuttnauer - as requested, installation of the water level sensor on a Tuttnauer autoclave
Просмотров 2336 месяцев назад
Tuttnauer - as requested, installation of the water level sensor on a Tuttnauer autoclave
GE B40, how to remove the measurement module
Просмотров 2597 месяцев назад
GE B40, how to remove the measurement module
Datex Ohmeda/GE Cardiocap 5 - Update to gas/agent calibration on a Cardiocap5
Просмотров 767 месяцев назад
Datex Ohmeda/GE Cardiocap 5 - Update to gas/agent calibration on a Cardiocap5

Комментарии

  • @samuelsereda7929
    @samuelsereda7929 7 дней назад

    Just wanted to say thank you from one biomed to another very helpful information. Please continue making more content 😅

  • @aonomus
    @aonomus 11 дней назад

    Maybe I missed it, did you mention what this pump came out of? Any chemical contamination/hygeine issues that you had to plan for?

    • @The_Biomed
      @The_Biomed 11 дней назад

      @aonomus this is a medical gas vacuum pump. They live-in equipment rooms, dirty, hot, but free of patients, general bacteria.

    • @aonomus
      @aonomus 11 дней назад

      @@The_Biomed Ah so these are for recovering inhaled gasses?

    • @The_Biomed
      @The_Biomed 11 дней назад

      @aonomus no, patient suction.

  • @Mario456SS
    @Mario456SS 11 дней назад

    Very cool!

  • @KevinFrancis-p3u
    @KevinFrancis-p3u Месяц назад

    Happy holidays bro, need to ask you about a steris high low valve assembly.

    • @The_Biomed
      @The_Biomed Месяц назад

      Sure thing, what's the question and what device are we talking about

    • @KevinFrancis-p3u
      @KevinFrancis-p3u Месяц назад

      @The_Biomed it's an old machine with a model marking R88S-1. It's an old amsco machine that I can't find any door gasket for. Measures 16x16, but when I install the 16x16 gasket it is way too much slack remaining. The question I have though is replacing the high low valve kit, how to set the two temperatures on the dial.

  • @msramya446
    @msramya446 Месяц назад

    What is the technical name and specifications of metal ring, any pressure regulator is there?. From cylinder how much flow we can admit to calibrate the machine?

  • @msramya446
    @msramya446 Месяц назад

    Explain details of connectors used

    • @The_Biomed
      @The_Biomed Месяц назад

      @msramya446 its a luer connection. Comes in multiple colors and has a metal ring on the end. If you google it,www.medtronic.com/content/dam/medtronic-wide/public/united-states/products/patient-monitoring/microstream-advance-luer-ordering-information-sheet.pdf

    • @msramya446
      @msramya446 Месяц назад

      Thank you for your reply Brother, I am from India,I am asking about the T price connector and End of T piece connection one metal ring valve is there ,what is the pressure setting .? How much pressure we have to give from cylinder?

  • @AbelMasonic
    @AbelMasonic Месяц назад

    Thank you sir, you are a gentleman and a colleague.

  • @eliefrancois7084
    @eliefrancois7084 Месяц назад

    Yes!!! I try that with no accessories but it keep saying the same thing… Do you have that can be a board issue

    • @diyhomeimprove
      @diyhomeimprove Месяц назад

      Yes, there could be some debris. You could pull 5he cover and move the front section off the board. Examine the contacts. If it is still giving you those codes, look for damage, remove and reseat all connections. After that, try a calibration. If you're still hitting those errors, you'll have to find a repair company

  • @eliefrancois7084
    @eliefrancois7084 Месяц назад

    Hello there!!!! I have a valleylab cautery machine that’s show an error code 197 and 81 every time I open it. I mean it’s doesn’t work, what can I donto fix it

    • @The_Biomed
      @The_Biomed Месяц назад

      Ok, well, 197 is stuck mono 2 coag. So try starting the unit with no pedals nor accessories plugged In. Hopefully 81 is just piggybacking on the 197 error.

  • @joes2085
    @joes2085 2 месяца назад

    Looks like a sterilizer with the wall separating clean and dirty

    • @The_Biomed
      @The_Biomed 2 месяца назад

      @@joes2085 steelco instrument washer

  • @beyondsafe9958
    @beyondsafe9958 2 месяца назад

    thank you sir, well done

  • @otisbigsby
    @otisbigsby 2 месяца назад

    That nipple is brass not copper. The hose you mentioned is a 3/4 iP (iron pipe) threads. Not garden hose Adapter is called a bushing. The nipple is called a shoulder

  • @otisbigsby
    @otisbigsby 2 месяца назад

    That nipple is brass not copper. The hose you mentioned is a 3/4 iP (iron pipe) threads. Not garden hose

  • @CatnamedMittens
    @CatnamedMittens 2 месяца назад

    Id love to see soldering.

    • @The_Biomed
      @The_Biomed 2 месяца назад

      one day, i'll work on that one...can't hold the camera and do the work, but i'll see what I can do

  • @leonelsanchez8589
    @leonelsanchez8589 2 месяца назад

    Hello friend, I have the gas and a Cardiocap 5, but when I inject the gas, the menu to adjust the deviations doesn't appear.

    • @The_Biomed
      @The_Biomed 2 месяца назад

      How long has the monitor been on? Takes about 10 minutes before it's ready

    • @leonelsanchez8589
      @leonelsanchez8589 2 месяца назад

      I've kept this on for more than 30 minutes. On my channel, I uploaded the calibration of the Escaio module. Thank you.

    • @The_Biomed
      @The_Biomed 2 месяца назад

      well the menu isn't going to just show up. have you checked your configuration to make sure you have AA and CO2? and selected FMG with the functions you have?

    • @leonelsanchez8589
      @leonelsanchez8589 2 месяца назад

      @@The_Biomed I'll check it tomorrow and let you know. I hope you can keep helping me. Thank you.

  • @EmmanuelOnigbo-m8p
    @EmmanuelOnigbo-m8p 2 месяца назад

    Hello my datex cardicap 5 monitor does not open . The screen is blank How do I fix it thanks

    • @The_Biomed
      @The_Biomed 2 месяца назад

      @EmmanuelOnigbo-m8p oh. When you say blank, do you see the table for the functions but no data, or a totally black display? Are the backlights working?

  • @Samlol23_drrich
    @Samlol23_drrich 2 месяца назад

    There’s no way I’m buying an $800 kit to calibrate my old machine. That price kills me. Wish I could rent it. I’ve got my 2540ea apart on my center island in my,kitchen as I type this. Took apart all 4 solenoids and ultrasonic cleaned the plungers, hoping for the best. I’ve got a high temp fault. The machine doesn’t get to 30psi of pressure. Gets up to about 24 or 25 psi when it hits 273 degrees and then faults at either 274 or 275 or a few minutes with no more pressure increase. I was hoping it was the air jet. I replaced it, and the rpi part I got was spitting water worse than my original. So I’m waiting for a second one. I’m afraid it’s the temperature sensor, but the service manual didn’t include any ohm readings to test the sensor. Any ideas? I’ve fixed just about every part on this 20 plus year old machine and hate to trash it. It’s got the Adjunct 3 board. Thanks for making this video. Not too many out there. 😊

    • @The_Biomed
      @The_Biomed 2 месяца назад

      1st step would be to determine if the readings are accurate bc that temp and pressure are not inline with each other. If the temp is accurate then you either have a pressure leak or need a calibration. That's where I would go 1st. Verify and calibrate

    • @Samlol23_drrich
      @Samlol23_drrich 2 месяца назад

      @ thanks very much for getting back to me. I was thinking the air jet was leaking, not allowing it to build pressure. My problem is that I don’t have the tools to properly test the pressure and temperature. (And that means I can’t properly diagnose the machine). if the air jet isn’t the issue I will have to trash the machine. It gets to the point of not being worth the cost of repair. What a shame. Thanks again.

    • @The_Biomed
      @The_Biomed 2 месяца назад

      @Samlol23_drrich the air jet is supposed to leak. It's how the steam circulates in the chamber. But the safety release valve shouldnt...if you overfill the rez so the safety release is half covered, you'll be able to see if it's leaking, same for the air outlet valve. There's probably a biomed company near you with a test kit than can trouble shoot it for you

    • @Samlol23_drrich
      @Samlol23_drrich 2 месяца назад

      @ truly appreciate all of your expertise. The reason plumbers and electricians are required to do apprenticeships is for exactly what I learned from your reply. Would have never thought to test the safety release for leaks by overfilling the reservoir. (Maybe that’s what my staff is doing every friggin time they fill it). Something I did fixed it. Even though I bought a new air jet, after I installed it the machine had the same symptoms, and the new jet was leaking much more than I thought it should. I know the machine like the back of my hand. How much steam comes out of that jet and the new one just didn’t look right. I ordered a second new one but figured I would replace the entire relief valve assembly, so I bought the entire thing. As I said in the prior post, while waiting for the 2nd day delivery, I brought the machine home and took it apart just in case there was a leak from one of the solenoids. I took all four apart, cleaned the copper seats and ran a carb cleaner brush through all the orafices, then pulled the temp sensor and cleaned it (gently), pulled every wiring harness and sprayed contact cleaner on both ends then put it all back together this morning with the new part that was delivered late yesterday. Just ran a cycle and it worked perfectly. So something I did fixed it. Once again, thanks for getting back to me and sharing your expertise. It’s something most won’t do (see my video on the lack of support from Hayward when my pool heater had an issue. Once they found out I was a consumer they wouldn’t help me). Thanks again! Rich in NY

  • @johnm4962
    @johnm4962 3 месяца назад

    What component is bad on the board to cause it to jump around.

    • @The_Biomed
      @The_Biomed 3 месяца назад

      It was suggested there may be a cold solder joint on either the pots, or the larger connector on the back. I have yet to look into that.

    • @johnm4962
      @johnm4962 3 месяца назад

      @@The_Biomed thanks I do a lot of board repair so I was curious thanks.

    • @The_Biomed
      @The_Biomed 3 месяца назад

      @johnm4962 if you come across an adjunc 3 pcb with a similar issue, let me know about the cold solder joint scenario

    • @johnm4962
      @johnm4962 3 месяца назад

      @@The_Biomed yeah I don’t come across these boards often and my machine doesn’t have this fault.

    • @johnm4962
      @johnm4962 3 месяца назад

      You could send me a bad board and I could go over it ?

  • @POETICJUSTICE15
    @POETICJUSTICE15 4 месяца назад

    Thank you for this video. VERY HELPFUL!

  • @Burger-s8z
    @Burger-s8z 4 месяца назад

    What if the voltage is wrong?

    • @The_Biomed
      @The_Biomed 4 месяца назад

      @@Burger-s8z what do you mean

    • @The_Biomed
      @The_Biomed 2 месяца назад

      there is no voltage to a steam trap, it's a purely mechanical, temperature controlled device

  • @indianvlogchannel1665
    @indianvlogchannel1665 4 месяца назад

    Pls make video ,vpro how can callibrate I am from India

  • @hassan_ALTia
    @hassan_ALTia 4 месяца назад

    What is the solution to the E008 problem?

    • @The_Biomed
      @The_Biomed 4 месяца назад

      from what i've seen, just power cycle. if it continues, you may have a pcb issue

    • @hassan_ALTia
      @hassan_ALTia 4 месяца назад

      ​@The_Biomed Yes there is a short circuit in the circuit but I could not detect what the harmonic element is

  • @FundsClinic
    @FundsClinic 4 месяца назад

    Do you know anyone in the Illinois area that works on Tuttnauer autoclaves.

    • @The_Biomed
      @The_Biomed 4 месяца назад

      @FundsClinic afraid not. Could probably find someone on the bmet/biomed reddit groups, or thru dotmed. Actually tuttnauer keeps a listing of every company they have trained. So tuttnauer would be the best resource to finding a company that is knowledgeable and authorized to work on them

  • @kevinfrancis4513
    @kevinfrancis4513 5 месяцев назад

    Is this the eagle 2000 or 3000? Man you're the best. Definitely good content. Need a longer video though....keep it up.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @kevinfrancis4513 this is the century v116. The 3000 series has individual valves, no manifolds

    • @kevinfrancis4513
      @kevinfrancis4513 5 месяцев назад

      I got one of these coming up to work on..

    • @kevinfrancis4513
      @kevinfrancis4513 5 месяцев назад

      Ok, think I saw individual valves but somehow they tore off the serial plate and name plate to identify the model. Let's see how it goes, thanks bro.

  • @hassan_ALTia
    @hassan_ALTia 5 месяцев назад

    Hello sir, can you explain steps 5 and 6 in calibrating this device?

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      no, i skipped those steps on purpose. I need a better set up to complete those steps. Once I have it I can share it.

    • @hassan_ALTia
      @hassan_ALTia 4 месяца назад

      Can you help me calibrate the devices? Can I have your WhatsApp?

  • @daisizhou
    @daisizhou 5 месяцев назад

    Hi sir.Have you ever used the DNI 454A ESU analyzer?

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @daisizhou been awhile but yes. Had that unit at several biomed shops I've worked in

    • @daisizhou
      @daisizhou 5 месяцев назад

      @@The_Biomed Can you get the service manual for 454a? I have been looking for it for a long time but have not been able to find it.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      I have been unable to locate a service manual for that unit. If I ever come across one I'll let you know and try to share it on #medwrench

  • @jasonberube3186
    @jasonberube3186 5 месяцев назад

    Can you list your test equipment that was used for the pressure calibration? Make/model

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @jasonberube3186 I'm using an old bio-tek nibp monitor tester(that's the only model I can find written on it) it doesn't go to the upper values so I'm still trying to get this employer to buy the right one. Then it's just some hoses and connectors.

    • @jasonberube3186
      @jasonberube3186 5 месяцев назад

      Oh ok. What’s the new make and model you trying to get. I just need to get something for the air source.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      ​@@jasonberube3186​ I was looking at the contec ms200, but after double checking, it only goes up to 400mmhg. I can't find one that goes up high enough. Just have to do a manual pump

    • @jasonberube3186
      @jasonberube3186 5 месяцев назад

      @@The_BiomedI found something from fluke it’s the 3130 portable pressure generator it will definitely create the pressure required to cal the ATS

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @jasonberube3186 wow, that's like an infusion tester and at $8-10k geez

  • @christinamoffett7684
    @christinamoffett7684 5 месяцев назад

    So, the hose set up for two ports connecting to 1 pressure meter and the proper pressure meter is all that is needed for the calibration. Is this all that you do for the pm procedure, or do you also perform a leak test? I heard you speak to the battery check.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      so the entire pm procedure starts with the normal visual inspection, 1: look for any damage, 2: do electrical safety tests(leakage is really a waste of time), 3: check the calibration, calibrate if necessary. 4: leak test( attach a large cuff, set the pressure to 475mmHg, place it into standby; Press the ON/STANDBY button until the alarm message “CUFF” “NOT” “DEFL” appears. Release the ON/STANDBY button and within 5 seconds of the alarm discontinuing, press and hold the ON/STANDBY button again. The button must be held in for an additional 10 seconds before the unit will be set to STANDBY. wait 10 minutes, then turn it back on then press alarm silence and next press Pressure. the display should read 400mmHg minimum. I don't think you can do 2 at once, but i could be wrong. I'll test this out next time an ATS 3000 comes by my bench. 5: pull the battery, pull battery cables, check battery voltage(12v minimum) plug the unit in and check battery voltage while attached, should be higher than 12 volts. that's it

  • @Ronnock
    @Ronnock 5 месяцев назад

    "If you don't know how to do electrical safety, you're in the wrong field." Gee, way to inspire future generations who would otherwise might want to learn about electrical safety. Is it a shame to be ignorant if one is willing? Did you come out of the womb knowing electrical safety?

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @@Ronnock actually I invented it when I was just an unfertilized egg. Also, this is for people already in the field, pretending to do the job and just slapping stickers. There are a lot of them

    • @Ronnock
      @Ronnock 3 месяца назад

      Yeah, I get that. I don't like those who just "lick and stick," but I maintain one should extend an olive branch to those who are too scared to admit they don't know.​@@The_Biomed

    • @The_Biomed
      @The_Biomed 3 месяца назад

      @@Ronnock agreed

  • @TDrTruong
    @TDrTruong 5 месяцев назад

    I HAVE THE SAME 2540M MY STERALIZER, IT IS BLOWING OUT STEAM FROM THE RESEVOUR TANK. WHY IS THAT? SOUNDS LIKE ITS GOING TO EXPLODE.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      Most likely, your safety release valve is compromised. I have a video showing the internals. Replace it and remember to test it during heatup once a week

    • @TDrTruong
      @TDrTruong 5 месяцев назад

      @@The_Biomed Thank you I will check it out.

  • @abdulla716
    @abdulla716 5 месяцев назад

    How many devices one kit can calibrate?

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      @@abdulla716 I don't know off hand how much has it uses per cal. There's no gauge on the regulator.

  • @achmadindra6848
    @achmadindra6848 5 месяцев назад

    how to fix ecg mac 2000 failed to boot

  • @ejp123md6
    @ejp123md6 5 месяцев назад

    what is the typical service charge? I am trying to decide to fix or buy another one.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      depends, labor rate could be anywhere from 100-150/hr depending on what your service company charges, contract or none. then parts used come with a markup based on mfr pricing. . if it has to be picked up/dropped off then a travel charge. I've got some open repair estimates running $500 up to $3000. the higher ones usually just result in a replacement quote. Lack of proper maintenance can result in a hefty charge. just following the user manual instructions saves a lot of money. having an annual service contract also means we find your issues before you suddenly find out you are without a working device right when you need it = cancelled cases = $$???

  • @Blueshandle
    @Blueshandle 5 месяцев назад

    Does this place not have a janitor? I feel like I’m gonna contract fourteen diseases just by looking at the floor.

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      it''s an industrial warehouse, not a patient. no sterility

    • @Blueshandle
      @Blueshandle 5 месяцев назад

      @@The_Biomed update : I now have malaria, tuberculosis, cancer, aids, diabetes, alzheimer’s, chickenpox, arthritis, asthma, cystic fibrosis, conjunctivitis, chronic fatigue syndrome, every alphavirus, and crabs. Bad looking floors cause bad feeling people. See something, say something. Generic government mandated health PSA poster saying.

  • @kevinfrancis4513
    @kevinfrancis4513 6 месяцев назад

    Think i am gonna need a full calibration for Analong input error with that test13 unit. Youre the best bro

    • @The_Biomed
      @The_Biomed 6 месяцев назад

      @kevinfrancis4513 working on it right now. Ran into a heatup error on an elara 11 and had to stop filming

    • @kevinfrancis4513
      @kevinfrancis4513 6 месяцев назад

      @@The_Biomed Have a lab line 3870EA with that analog error and trying to acquire that calibration tool to fix it. (Bacsoft)

  • @kevinfrancis4513
    @kevinfrancis4513 6 месяцев назад

    Got a tuttnauer 3870EAD giving me "Analog input error", changed the water electrode and the temperature sensor works. The pressure transducer I don't know how to check if its outside of normal range, can you shed any light on this please.

    • @The_Biomed
      @The_Biomed 6 месяцев назад

      @kevinfrancis4513 When are you getting this message, at startup, during cycle(which phase?)? the adjunc3 pcb is your analog board, check to see if all of your connectors are on the right pins. For checking your pressure transducer, easiest way is to try and pressurized the chamber with an external pressure gauge attached.

    • @kevinfrancis4513
      @kevinfrancis4513 6 месяцев назад

      @@The_Biomed It's a Bacsoft, and I get that error on startup. I had to reload the OS but it seems its a calibration issue i may be having. I realise Tuttnauer dictates each part must be within a value range.

    • @The_Biomed
      @The_Biomed 6 месяцев назад

      @kevinfrancis4513 yes, there are cal directions in the tech manuals. There's a kit to use your multimeter with to verify values and a pressure gauge

    • @kevinfrancis4513
      @kevinfrancis4513 6 месяцев назад

      @@The_Biomed i need to get that kit for the other machine. This machine can be programmed from the GUI and all seems to work well except that analog error. I found a EZ11 manual that is similar, am gonna check the instructions. Thanns

  • @The_Biomed
    @The_Biomed 6 месяцев назад

    Does it print when you got rhythm? When you hit ekg, Does it ask you to enter patient info? Are you getting a blank page or no paper even comes out?

  • @sean7854
    @sean7854 6 месяцев назад

    thank you for the video. I have one and when i hit print it says printing but it wont print the ekg . it does print other reports though . what do you think can be my issue ?

    • @The_Biomed
      @The_Biomed 6 месяцев назад

      what other reports?

  • @beyondsafe9958
    @beyondsafe9958 7 месяцев назад

    I have some questions

    • @diyhomeimprove
      @diyhomeimprove 7 месяцев назад

      It's a rough intro video. What questions do you have?

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      just realized i replied from a diff account. what questions do you have?

  • @jhonjairopelaez8459
    @jhonjairopelaez8459 7 месяцев назад

    Show us the instalation. Thanks

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      asked and answered. I've uploaded an installation video for you

  • @The_Biomed
    @The_Biomed 7 месяцев назад

    Update on this, I recently learned the calibrate ID is not really used on the cardiocap 5. They want you to calibrate thru parameters-airway gases. See my more recent video

  • @The_Biomed
    @The_Biomed 7 месяцев назад

    Update on this, I recently learned the calibrate ID is not really used on the cardiocap 5. They want you to calibrate thru parameters-airway gases. See my more recent video

  • @cameronwinship2555
    @cameronwinship2555 7 месяцев назад

    Thank you for this video!

  • @ratanaalibaba3405
    @ratanaalibaba3405 8 месяцев назад

    Hello dear my machine sample line block please help

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      Are you getting. Message saying that? You might need some new naifon tubing. Most co2 monitors have a pm kit with it included

  • @basilsaju6944
    @basilsaju6944 8 месяцев назад

    Very useful video but next time kindly add the refixing of the part back together also

    • @The_Biomed
      @The_Biomed 5 месяцев назад

      I made one, I just have to find time to sit down, edit and piece it together. Unfortunately the OR I was in had music playing and I couldn't get them to turn it off, that would result in copywrite failure and a video youtube won't let me post

  • @Pitmaster_GoldenShellback
    @Pitmaster_GoldenShellback 8 месяцев назад

    How do you actually remove the sensor from the autoclave? I have the nut loose, but it only loosens so much and wants to stay on the part where the nut threads onto. I can turn the sensor inside the chamber, but it won’t come out. EDIT - I got it!

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      once you remove the compression nut, it can be pulled out from inside of the chamber. If you'e is stuck, give it some more force, perhaps a pair of needle nose in a prying method, or channel locks

    • @Pitmaster_GoldenShellback
      @Pitmaster_GoldenShellback 8 месяцев назад

      @@The_Biomed thanks! Having never done this before, I wasn't sure how much force I needed to get the compression nut off the nylon tube so I was being overly cautious at first.

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      @Pitmaster_GoldenShellback no problem, it's just brass so being weary was the right move. I keep stock of multiple sizes and ferrels just in case. They do strip and warp easily.

  • @alexrousedds
    @alexrousedds 8 месяцев назад

    Is there any advice at all you have for a chamber that floods when the autoclave door opens and then the panel says "water low"? It keeps gushing water out the front when we open the autoclave door. Thank you!

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      without seeing the unit, I'd have to surmise that your water level sensor is dirty, or malfunctioning. the midmark sterilizers use a continuity circuit to detect when there is the right amount of water in the chamber. water low is because all of your water is in the chamber, (and floor?)

  • @nicholasochiengokoth1596
    @nicholasochiengokoth1596 8 месяцев назад

    Hi. What is the regulator that can be used? I'm over shooting the values,?

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      I don't have a good one yet. I requested the contec ms200 Sim. The listed capabilities will meet our needs.

    • @nicholasochiengokoth1596
      @nicholasochiengokoth1596 8 месяцев назад

      @@The_Biomed I meant the knob contraption that can allow steady flow from the compressor.

    • @The_Biomed
      @The_Biomed 8 месяцев назад

      oh, i usually default to a Parker Watts R364. then you just have to get caps and fitting for your application

  • @ALTONBENALLY
    @ALTONBENALLY 10 месяцев назад

    LOW HEAT Error? I'm working on an EZ10 that keeps failing with LOW HEAT. The Air Outlet Valve seems fine, but the Safety Relief Valve has a noticeable leak at the ports. So, I took it apart with two 3/4" wrenches. It doesn't look as bad as yours in the video, but certainly not in top condition. I'm replacing it tomorrow and hopefully that resolves my issue. I cleaned after disassembly and ran a test cycle for experimental purposes, and it actually performed a little better. Reached pressure and temp much quicker, but still failed with a LOW HEAT Error. Next will be the Air Outlet Valve which has extremely tiny damages on seat of the valve base.

    • @The_Biomed
      @The_Biomed 10 месяцев назад

      That will do. All your steam is escaping thru the safety valve. Did you drop a wore brush Into it to clean out the seat as well? You could just plug it and run a cycle. Just have to keep an eye on the pressure readings

    • @ALTONBENALLY
      @ALTONBENALLY 10 месяцев назад

      @@The_Biomed Yes, I cleaned the seat on the Air Valve but didn't test it after cleaning it. I'm going to replace the Safety Valve first, run a test. If it fails again then the Air Valve/Plunger Assembly is next.

  • @pwest3732
    @pwest3732 10 месяцев назад

    Thanks for the vid. Taking 2 of these apart today for flickering displays. And the service manual is garbage. Thanks Philips Engineers and tech writers!

    • @The_Biomed
      @The_Biomed 10 месяцев назад

      Soooooo many screws, lol